  • HighFDA (Devices)·Z-1066-2025·2025-02-05

    Infinity Central Station speakers may fail to produce critical patient alarms

    Edifier speakers in Draeger's Infinity Central Station may not produce audible alarms when in analog mode, potentially delaying patient treatment. Approximately 1,013 units are affected worldwide.

    Product
    Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of dat
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1039-2025·2025-02-05

    Knee Implant Component Recalled for Incorrect Metal Post Assembly

    Zimmer, Inc. is recalling 155 units of NexGen LCCK Legacy knee implant components because the metal support post was assembled incorrectly in some units. The defective assembly could compromise implant function.

    Product
    NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1022-2025·2025-02-05

    Medtronic Percutaneous Reference Pins May Not Fit Properly

    Medtronic is recalling sterile percutaneous reference pins used in image-guided surgery because the pins may not fit into the patient reference frame or adapter, potentially affecting proper surgical tracking.

    Product
    Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 97733236 The Sterile Percutaneous Reference Pin Set is intended to be used to place and remove a pin percutaneously into bony patient anatomy for rigid attachment of a pat
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1062-2025·2025-02-05

    Karl Storz Ureteroscope Model 27001K Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling its Ureteroscope (Model 27001K) because the Instructions for Use contain reprocessing procedures that have not been reviewed or approved by the FDA. All 185 units distributed nationwide are affected.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27001K, Ureteroscope, 9.5 Fr., 6¿, 34 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·25V056000·2025-02-05

    2025 Honda Pilot third-row seat belt may not restrain occupant

    Honda is recalling certain 2025 Pilot vehicles because the third-row passenger seat belt may not function due to incorrect routing. A non-functioning seat belt cannot properly restrain occupants, increasing injury risk during a crash.

    Product
    HONDA — HONDA PILOT
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1059-2025·2025-02-05

    Karl Storz Uretero-Renoscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz uretero-renoscope devices are being recalled because their instructions for reprocessing and sterilization have not been FDA-reviewed or approved. This poses a potential patient safety risk.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27002L, Uretero-Renoscope, 9.5 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1040-2025·2025-02-05

    Medline Procedure Kits Recalled for Surgical Mask Bioburden Testing Failure

    Medline has recalled 749 procedure kits containing surgical masks that failed to meet bioburden testing standards under EN 14683. The affected kits were distributed nationwide.

    Product
    Medline procedure kits labeled as: 1) Port Access Kit, Reorder number CVI5130; 2) VANTEX 7FR 3L 20CM CVC STOCK+ BUNDLE US. Reorder number STCVC01USGV; general surgical tray
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1057-2025·2025-02-05

    Karl Storz Uretero-Renoscope with Unapproved Reprocessing Instructions Recalled

    Karl Storz Endoscopy is recalling 59 units of its uretero-renoscope due to instructions for use containing reprocessing procedures that the FDA has not reviewed or approved for safety and efficacy.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27001L, Uretero-Renoscope, 8 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1068-2025·2025-02-05

    Implantable Vagus Nerve Stimulation Device May Stop Delivering Therapy

    LivaNova's SenTiva DUO vagus nerve stimulation device may stop delivering therapy due to an internal component issue. The recall affects 84 units distributed worldwide.

    Product
    Brand Name: SenTiva DUO" Product Name: VNS Therapy¿ SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-D Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, le
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1031-2025·2025-02-05

    AXIOS Stent Delivery System recalled due to distal sheath detachment

    Boston Scientific is recalling 38 units of AXIOS Stent and Delivery System (15mmx10mm) because the outer sheath's distal tip may detach, preventing proper stent expansion and requiring device replacement.

    Product
    AXIOS Stent and Delivery System 15mmx10mm UPN: M00553750
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1067-2025·2025-02-05

    Implantable VNS Therapy Generators May Stop Delivering Treatment

    LivaNova is recalling approximately 8,223 SenTiva VNS Therapy generators due to a potential internal component failure that may cause the implanted device to stop delivering seizure therapy. Affected patients should contact their physician.

    Product
    Brand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and exte
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1060-2025·2025-02-05

    Karl Storz Ureteroscope Recall Due to Unapproved Reprocessing Instructions

    Karl Storz ureteroscopes are being recalled because their instructions contain reprocessing procedures that have not been reviewed or approved by the FDA. The recall affects 3,069 units distributed nationwide.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27010K, Ureteroscope, 7 Fr., 34 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1023-2025·2025-02-05

    CyberKnife Treatment System: Internal Snap Ring Detachment Risk

    Snap rings may detach from internal shafts in CyberKnife Treatment Delivery Systems, potentially causing uncontrolled rotation on treatment axes. This affects 212 units distributed worldwide for stereotactic radiosurgery.

    Product
    CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1033-2025·2025-02-05

    HOT AXIOS Stent Delivery System Outer Sheath May Detach, Preventing Stent Expansion

    Boston Scientific's HOT AXIOS stent delivery system has a defect where the outer sheath can detach and prevent proper stent expansion, requiring device replacement.

    Product
    HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 8mmx8mm UPN: M00553530
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1055-2025·2025-02-05

    Karl Storz Uretero-Renoscope recalled for unapproved reprocessing instructions

    Karl Storz Endoscopy has recalled the KARL STORZ ENDOSKOPE Uretero-Renoscope nationwide because its instructions for use contain reprocessing procedures that lack FDA review and approval for safety and efficacy.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27010L, Uretero-Renoscope, 7 Fr., 43cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1028-2025·2025-02-05

    AXIOS Stent Distal Tip Detachment Prevents Proper Expansion

    Boston Scientific is recalling AXIOS Stent devices due to outer sheath distal tip detachment that can prevent proper stent expansion and require device replacement during the procedure.

    Product
    AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553680
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1064-2025·2025-02-05

    Karl Storz Ureteroscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz ureteroscopes contain reprocessing instructions that have not been reviewed and approved by the FDA. Affected devices should not be reprocessed using unapproved methods.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27011K, Ureteroscope, 8 Fr., 6¿, 34cm, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1050-2025·2025-02-05

    BD Pyxis MedStation Door Latch Assembly Failures in Automated Dispensing Cabinets

    The FDA is recalling approximately 33,589 BD Pyxis MedStation automated dispensing cabinets due to door latch assembly failures that may leave doors open, potentially delaying medication access and enabling medication diversion.

    Product
    BD Pyxis MedStation ES AUX Tower REF 343 BD Pyxis " MedStation 4000 AUX Tower REF 314 The BD Pyxis" MedStation ES and Pyxis" MedStation 4000 are automated dispensing cabinets (ADC) that are intended to securely store and dispense medications to a qualified and authorized he
    Category
    Medical Device
    Distribution
    54 states
  • HighFDA (Devices)·Z-1020-2025·2025-02-05

    HemosIL Heparin Calibrators Recalled Over Quality Control Test Failures

    Instrumentation Laboratory recalls HemosIL Heparin Calibrators worldwide due to quality control failures. The diagnostic controls did not meet expected performance standards, potentially affecting accuracy of heparin testing in clinical laboratories.

    Product
    HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0455-2025·2025-02-05

    Alpha Test supplement capsules recalled for unapproved cathine ingredient

    Iovate Health Sciences is recalling Alpha Test 120-count capsules containing cathine, an unapproved ingredient not permitted in dietary supplements. Affected products were distributed across the United States and internationally.

    Product
    Alpha Test 120 count Black labeled plastic bottle, 12 bottles per case of 120 count capsules each per bottle, Store in a cool, dry place (60 F to 80 F), UPC 6-31656-60676-8, SKU0001191, SKU0002075
    Category
    Drug
    Distribution
    30 states
  • HighFDA (Devices)·Z-1061-2025·2025-02-05

    Karl Storz Ureteroscope recalled nationwide for unapproved reprocessing instructions

    Karl Storz Ureteroscope model 27000K is being recalled due to instructions for use containing reprocessing modalities not reviewed or approved by the FDA. The recall affects 82 units distributed nationwide.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27000K, Ureteroscope, 7 Fr., 6¿, 34 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1036-2025·2025-02-05

    Medical stent delivery system outer sheath component detachment prevents proper expansion

    The HOT AXIOS Stent delivery system's outer sheath tip may detach during use, preventing proper stent expansion and requiring device exchange. Boston Scientific is recalling 1,102 units distributed nationwide.

    Product
    HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 20mmx10mm UPN: M00553560
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1053-2025·2025-02-05

    Karl Storz Pediatric Cysto-Urethroscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling 185 Pediatric Cysto-Urethroscope instruments nationwide. The product instructions contain reprocessing methods that have not been reviewed or approved by the FDA for safety and effectiveness.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27030KB, Pediatric Cysto-Urethroscope, NON STERILE, CE
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1030-2025·2025-02-05

    AXIOS Stent and Delivery System Recalled for Improper Expansion Risk

    The AXIOS Stent and Delivery System is being recalled because the outer sheath can detach and prevent proper stent expansion, requiring device replacement during the procedure.

    Product
    AXIOS Stent and Delivery System 10mmx10mm UPN: M00553740
    Category
    Medical Device
    Distribution
    Distributed nationwide