Most US product recalls have nationwide distribution — the manufacturer ships into every state and the recall sweeps the same way. We include those in this listing because they affect Rhode Island consumers even though the source notice doesn’t name the state. State-specific recalls are usually for products with regional distribution (a regional dairy, a local store brand) or where the agency was able to narrow the affected shipments to specific destinations.

If you’re checking whether a recall affects your purchase, the recall detail page lists the exact production lot codes, UPCs, and best-by dates. Match those against your package — distribution scope is a starting filter, not the final answer.

9026–9050 of 20189

  • HighFDA (Devices)·Z-1858-2024·2024-05-29

    Magnetic Resonance System Loose Terminal Connection May Cause Fire

    Philips MR RT imaging systems may have loose electrical connections that could cause smoke or fire in hospital technical rooms. Two units with serial numbers 45 and 47 are affected worldwide.

    Product
    MR RT - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781439
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1842-2024·2024-05-29

    MR Imaging System Recall for Loose Electrical Terminal Connection Fire Risk

    Philips AchievaXR MR systems may have loose terminal connections in the power distribution unit that could create hotspots and cause smoke or fire. The FDA has recalled 35 affected units worldwide.

    Product
    AchievaXR -Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): 781253
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1884-2024·2024-05-29

    Noblus AC Adapter for Ultrasound System Recalled for Smoke Hazard

    FUJIFILM is recalling Noblus AC Adapters nationwide due to fire and smoke hazard. The adapters can overheat and generate smoke when the power cable is repeatedly inserted and removed.

    Product
    Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2126-2024·2024-05-29

    Medline non-sterile syringes recalled for out-of-specification manufacturing

    Jiangsu Shenli Medical is recalling Medline non-sterile syringes (Model 91862, Lot 63722030001) because sizes and configurations exceed FDA-cleared specifications. The recall affects 400 units distributed in California, Florida, Georgia, Illinois, Tennessee, and Virginia.

    Product
    Brand Name: MEDLINE Product Name: SYR 20ML/L WHITE HEP FLUSH Model/Catalog Number: 91862 Product Description: NON-Sterile syringes without needles for single use Component: No
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2101-2024·2024-05-29

    FDA Recalls Wolf Medical Disposable Syringes for Non-Conforming Configurations

    Jiangsu Shenli Medical has recalled over 1 million Wolf Medical WOLF-PAK 1 mL disposable syringes because piston syringe sizes and configurations exceed those cleared under the company's 510(k) approval.

    Product
    Brand Name: Wolf Medical - WOLF-PAK Product Name: DISPOSABLE SYRINGE 1 mL (cc) Luer Lock STERILE LATEX FREE NON PYROGENIC Model/Catalog Number: SYR01L Serial Number: 850007185411 Product Description: NON-Sterile syringes without needles for single use Component: No
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1885-2024·2024-05-29

    Scopis ENT Navigation Software Contains Visual Guidance Alignment Error

    Unreleased software in Scopis ENT surgical navigation systems causes visual guidance to display inaccurately relative to physical reference points, affecting surgical precision. Affects 705 units nationwide.

    Product
    8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. T
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1857-2024·2024-05-29

    Philips MR System 1.5T Marlin MRI devices recalled due to fire hazard

    Philips is recalling 10 MR System 1.5T Marlin MRI devices worldwide due to a fire and smoke hazard. A loose electrical connection in the mains distribution unit may create a hotspot that causes smoke or fire in hospital technical rooms.

    Product
    MR system 1.5T Marlin- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781483
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1861-2024·2024-05-29

    Philips Medical Device Recalled for Fire Hazard in Electrical Connection

    Philips is recalling the SmartPath to dStream for 3.0T medical device due to a loose electrical connection in the mains distribution unit that may cause smoke or fire in hospital technical rooms. Three units are affected.

    Product
    SmartPath to dStream for 3.0T Model Number (REF): 782145
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1859-2024·2024-05-29

    Philips Magnetic Resonance System MR RT Upgrades Recalled for Fire Risk

    Philips is recalling MR RT Upgrades Magnetic Resonance diagnostic systems due to a loose terminal connection in the mains distribution unit that could cause smoke or fire. Two units with serial numbers 70980 and 71569 are affected worldwide.

    Product
    MR RT Upgrades- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781440
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1874-2024·2024-05-29

    FDA Recalls Prelude Guide Sheath Roadster for Component Size Mismatch

    Merit Medical Systems is recalling 18 units of Prelude Guide Sheath Roadster due to sheaths labeled as .018 dilators potentially containing .038 dilators instead. The mismatch poses a risk if the wrong-sized component is used during medical procedures.

    Product
    Prelude Guide Sheath Roadster, REF PG5F45S018, 5F 0.076 (1.9 mm), 45 cm, STERILE EO, RX ONLY
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1848-2024·2024-05-29

    Philips Recalls Ingenia 3.0T MR Systems for Loose Mains Terminal Fire Hazard

    Philips is recalling 1,517 Ingenia 3.0T magnetic resonance (MR) systems due to a potentially loose terminal connection in the mains distribution unit that may create a hotspot and cause fire or smoke.

    Product
    Ingenia 3.0T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782103 (2) 781377 (3) 781342
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2050-2024·2024-05-29

    Non-Sterile Syringes Recalled Due to Manufacturing Specification Violation

    Jiangsu Shenli Medical is recalling 754,600 non-sterile syringes because the piston syringe sizes and configurations exceed those cleared by the FDA. Affected units were distributed nationwide.

    Product
    Brand Name: MEDLINE Product Name: SYR 5ML L/S Model/Catalog Number: 83085 Product Description: NON-Sterile syringes without needles for single use Component: No
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1875-2024·2024-05-29

    Merit Medical Guide Sheath Roadster Recalled for Incorrect Dilator Sizing

    Merit Medical is recalling Prelude Guide Sheath Roadster devices because some sheaths labeled as .018 dilators may actually contain .038 dilators, creating a risk of using incorrectly-sized medical devices.

    Product
    Prelude Guide Sheath Roadster, REF PG6F45S018, 6F 0.087" (2.2 mm), 45 cm, STERILE EO, RX ONLY
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1882-2024·2024-05-29

    Cascade IOMAX Cortical Module recalled for potential electrical shock hazard

    Cadwell Industries is recalling the Cascade IOMAX Cortical Module because an incorrectly installed insulator could cause electrical arcing and potential patient shock.

    Product
    Cascade IOMAX Cortical Module, REF: 190296-200
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1851-2024·2024-05-29

    Philips Ingenia Ambition X MR systems recalled for loose terminal connection fire hazard

    Philips is recalling 509 Ingenia Ambition X MR systems due to loose terminal connections in the g-MDU L3 unit that may create hotspots and cause smoke or fire in hospital technical rooms.

    Product
    Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782138; (2) 782109; (3) 781356
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2123-2024·2024-05-29

    GMAX SYR 60ML/LL Blue Syringes Recalled for Unapproved Specifications

    Jiangsu Shenli Medical Production Co. is recalling GMAX SYR 60ML/LL blue syringes because the manufactured sizes and configurations fall outside the range cleared by the FDA. Approximately 116,160 units were distributed nationwide.

    Product
    Brand Name: GMAX Product Name: SYR 60ML/LL Blue syringe Model/Catalog Number: TS2260LB-M Product Description: NON-Sterile syringes without needles for single use Component: No
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1838-2024·2024-05-29

    Philips MRI System Recalled for Potential Fire in Power Distribution Unit

    Philips is recalling 2 Achieva 1.5T MRI systems worldwide due to a potential loose connection in the power distribution unit that could cause smoke or fire in hospital technical rooms.

    Product
    Achieva 1.5T Initial system Model Number (REF): 781178
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1840-2024·2024-05-29

    Medical MR System Electrical Terminal Connection May Create Fire Hazard

    A Philips Achieva 3.0T for PET MR diagnostic system (Serial 43006) may have a loose mains distribution unit terminal that could overheat and cause smoke or fire.

    Product
    Achieva 3.0T for PET -Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): 781477
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1815-2024·2024-05-29

    Yamaha Surface Mounter Equipment Recalled for Missing Laser Safety Certification Labels

    Yamaha is recalling certain surface mounter equipment models because they lack required certification labels for laser safety compliance under FDA regulations.

    Product
    Yamaha Surface Mounter YR series, YS series, i-Pulse series, Sigma series, S series, M series. Affected models: YSM10, YSM20R(SV)-1, YSM20R(SV)-2, YSM20R-1, YSM20R-2, YSM20-1, YSM20-2, YSM20W-2, YRM20-1, YRM20-2, SIGMA-G5S2, S20, M20 ***Updated as of 6/10/2024*** Surface Mounte
    Category
    Other
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1863-2024·2024-05-29

    Philips MRI System Recall: Loose Electrical Terminal May Cause Fire

    Philips MRI diagnostic systems are recalled because a loose electrical connection in the g-MDU unit may create a hotspot, potentially causing fire and smoke in hospital technical areas.

    Product
    SmartPath to Ingenia Elition X - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 782118
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1881-2024·2024-05-29

    SURGIFOAM Absorbable Gelatin Powder Kit Recalled for Open Applicator Tip Seals

    Ethicon recalls SURGIFOAM Absorbable Gelatin Powder Kit (Product Code 1979) due to open seals on applicator tips in three lots. The recall affects 5,184 units distributed nationwide.

    Product
    SURGIFOAM" Absorbable Gelatin Powder Kit, Product Code 1979.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2048-2024·2024-05-29

    Non-sterile Piston Syringes Distributed Outside FDA-Approved Specifications

    Medline piston syringes (Model 83083) manufactured by Jiangsu Shenli are being recalled because their sizes and configurations fall outside FDA-approved specifications. Approximately 494,000 units are affected across the United States.

    Product
    Brand Name: MEDLINE Product Name: SYR 3ML L/S Model/Catalog Number: 83083 Product Description: NON-Sterile syringes without needles for single use Component: No
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2067-2024·2024-05-29

    Medline 20mL Non-Sterile Syringes Recalled for Design Specification Deviation

    Medline non-sterile 20mL syringes (Model 91835) are being recalled because the piston sizes and configurations do not match FDA-cleared specifications. Approximately 39,200 units are affected.

    Product
    Brand Name: MEDLINE Product Name: SYR 20ML L/L RED Model/Catalog Number: 91835 Product Description: NON-Sterile syringes without needles for single use Component: No
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1855-2024·2024-05-29

    Philips Panorama HFO MR Systems Loose Terminal Connection May Create Fire Hazard

    Philips Panorama HFO MR diagnostic imaging systems (Model 781350) contain a loose terminal connection in the mains distribution unit that may create a hotspot and cause smoke or fire. Approximately 118 units are affected worldwide.

    Product
    Panorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781350
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1887-2024·2024-05-29

    Siemens IMMULITE 2000 Anti-TG Ab Test Kits Recalled for Falsely Elevated Results

    Siemens Healthcare Diagnostics is recalling certain IMMULITE 2000 Anti-TG Ab test kits due to the potential for falsely elevated thyroid antibody results in patient samples.

    Product
    IMMULITE 2000 Anti-TG Ab Catalog # L2KTG6 (600 Tests), SMN 10381655, System, Test, Thyroid Autoantibody.
    Category
    Medical Device
    Distribution
    Distributed nationwide