Hysteroscope Model HYF-1T recalled due to moisture in device channels
Aizu Olympus is recalling Hysteroscope Model HYF-1T devices because the channel air drying process after repair was not validated. Some returned units had residual moisture that could allow contamination and infection.
- Product
- Hysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.
- Category
- Medical Device
- Distribution
- Distributed nationwide