The Recall Desk

State

Nevada product recalls

73,219 recalls have nationwide distribution and so reach Nevada. 5,696 additional recalls listed Nevada specifically in their distribution scope.

What the numbers show

Total
5,696
Critical
951
Severe
1,479
Most recent
2026-09-25

Of the 5,696 recalls naming Nevada we have scored against our rubric, 951 (17%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,479 (26%) are Severe.

These figures cover only the 5,696 recalls whose distribution notice names Nevada specifically. The 73,219 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

16301–16325 of 78915

  • SevereFDA (Devices)·Z-1801-2023·2023-06-28

    Medtronic implantable cardioverter defibrillators recalled for potential reduced-output defect

    Medtronic has recalled certain ICD and CRT-D cardiac devices due to a rare risk of reduced or no energy output during high-voltage therapy. The affected models contain a specific glassed feedthrough component.

    Product
    ICD-VR DVEX2E4 EV ICD OUS EV4, Model Number DVEX2E4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1753-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver High-Voltage Therapy Energy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough defect. Approximately 11,794 units worldwide are affected.

    Product
    CRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1821-2023·2023-06-28

    Implantable cardiac defibrillators recalled for potential therapy failure risk

    Medtronic recalls ICD COBALT VR MRI DF4 defibrillators due to a rare potential for reduced or no energy output during high-voltage therapy. The defect affects devices manufactured with a specific glassed feedthrough component.

    Product
    ICD COBALT VR MRI DF4, Model Number DVPB3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1786-2023·2023-06-28

    Implantable Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic implantable defibrillators with a specific feedthrough may fail to deliver adequate energy during high voltage therapy. Approximately 476 affected units distributed nationwide and worldwide.

    Product
    CRTD CROME HF MRI IS1 DF4, Model Number DTPC2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1791-2023·2023-06-28

    Implantable Defibrillators With Defective Feedthrough May Fail to Deliver Therapy

    Certain Medtronic implantable defibrillators may produce reduced or no energy output during high-voltage therapy due to a defective feedthrough component affecting 71 units.

    Product
    ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1815-2023·2023-06-28

    Implantable Defibrillators With Defective Feedthrough May Fail During Critical Therapy

    Medtronic is recalling 14,270 implantable cardioverter defibrillators with a specific feedthrough that may produce reduced or no energy output during high-voltage therapy. The affected devices may fail to deliver critical treatment when needed.

    Product
    ICD-VR DVMD3D4 PRIMO MRI, Model Number DVMD3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1709-2023·2023-06-28

    Medtronic Implantable Defibrillators May Fail to Deliver Therapy

    Certain Medtronic ICDs manufactured with a specific feedthrough may fail to deliver high-voltage therapy. Patients should contact their healthcare provider immediately.

    Product
    ICD-DR DDBC3D1 EVERA S IS1/DF1 GLOB, Model Number DDBC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1736-2023·2023-06-28

    Medtronic Implantable Defibrillators May Fail to Deliver Therapy

    Medtronic is recalling certain implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators due to a rare risk of reduced or no energy output during high voltage therapy.

    Product
    CRTD DTBA2D1G VIVA XT IS1 DF1 OUS GOLD, Model Number DTBA2D1G; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1716-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Potential Energy Output Failure

    Medtronic is recalling certain implantable cardioverter defibrillators and cardiac resynchronization therapy devices due to a rare potential for reduced or no energy output during high voltage therapy. Approximately 37,547 units are affected.

    Product
    ICD-DR DDMC3D4 EVERA MRI S IS-1/DF4 GLOB, Model Number DDMC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Drugs)·D-0893-2023·2023-06-28

    Prescription Drug Recall: Ziprasidone Cartons Contain Wrong Medication

    Unit Dose cartons labeled as Ziprasidone Hydrochloride 20 mg were found to contain Dronabinol 2.5 mg instead. Lot T04769 (Exp. 2024/12) was distributed nationwide.

    Product
    ZIPRASIDONE — ZIPRASIDONE (ZIPRASIDONE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1705-2023·2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic is recalling 522 implantable cardioverter defibrillators that may deliver reduced or no energy during high-voltage therapy due to a feedthrough defect.

    Product
    ICD-DR DDBB1D4 EVERA XT IS1/DF4 US, Model Number DDBB1D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1729-2023·2023-06-28

    Medtronic CRT-D Implants Recalled for Rare Energy Output Failure Risk

    Medtronic is recalling certain CRT-D (cardiac resynchronization therapy defibrillator) models due to a rare risk of reduced or no energy output during high-voltage therapy. The FDA has classified this as a Class I recall affecting 1,473 units.

    Product
    CRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1734-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Reduced Therapy Output

    Medtronic is recalling 14 units of CRTD VIVA QUAD XT implantable cardioverter defibrillators due to rare potential for reduced or no energy output during high voltage therapy. Affected devices were manufactured with a specific feedthrough defect.

    Product
    CRTD VIVA QUAD XT CUST PARY GOLD, Model Number DTBA1QQPG; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1713-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may experience reduced or no energy output during high-voltage therapy, potentially preventing treatment of life-threatening heart rhythms.

    Product
    ICD-DR DDMB2D1 EVERA MRI XT OUS DF1, Model Number DDMB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1720-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High Voltage Therapy

    Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during high voltage therapy due to a specific feedthrough manufacturing defect. Approximately 8,219 units were affected worldwide.

    Product
    ICD-DR DDME3D4 MIRRO MRI, Model Number DDME3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1751-2023·2023-06-28

    Medtronic Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic is recalling 1,754 CRT-D and ICD implantable defibrillators with a specific feedthrough component that may not deliver therapy when needed. Devices were distributed nationwide and worldwide.

    Product
    CRT-D DTBC2Q1 BRAVA QUAD IS4/DF1 INTL, Model Number DTBC2Q1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1806-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    Medtronic is recalling 15,381 implantable cardioverter defibrillators due to rare potential for reduced or absent energy output during high-voltage therapy caused by a specific feedthrough design defect.

    Product
    ICD-VR VISIA MRI AF XT OUS DF4, Model Number DVFB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1808-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High Voltage Therapy

    Medtronic implantable cardioverter defibrillators may fail to deliver therapy due to a defect in the electrical feedthrough. The devices risk reduced or no energy output during high-voltage treatment.

    Product
    ICD-VR VISIA MRI AF S US/OUS DF4, Model Number DVFC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1807-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillator may lose energy output during therapy

    Medtronic recalls approximately 5,622 implantable cardioverter defibrillators that may produce reduced or no energy output during high-voltage therapy due to a specific feedthrough component defect.

    Product
    ICD-VR DVFC3D1 VISIA AF MRI S OUS/US DF1, Model Number DVFC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1822-2023·2023-06-28

    Implantable Heart Defibrillators May Fail to Deliver Therapy

    Medtronic implantable heart defibrillators may fail to deliver therapy during emergencies. The rare defect affects 510 devices distributed nationwide and worldwide.

    Product
    ICD CROME VR MRI IS1 DF1, Model Number DVPC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1772-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough design issue. This could prevent life-saving electrical therapy from being delivered.

    Product
    CRTD DTMC2D4 COMPIA MRI OUS DF4, Model Number DTMC2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1748-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapy During Critical Events

    Certain Medtronic implantable defibrillators may fail to deliver energy during high-voltage therapy due to a feedthrough manufacturing defect. 523 units distributed worldwide are affected.

    Product
    CRT-D DTBB2QQ VIVA QUAD S IS4/DF4 INTL, Model Number DTBB2QQ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1789-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    Medtronic implantable defibrillators with certain feedthroughs may fail to deliver high-voltage therapy when needed. This rare malfunction could prevent life-saving electrical treatment.

    Product
    ICD-VR DVAB1D1 VISIA AF US IS1/DF1, Model Number DVAB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1749-2023·2023-06-28

    Implantable Cardiac Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Certain Medtronic implantable cardiac defibrillators with specific glassed feedthrough components may deliver reduced energy during high-voltage therapy. This could prevent the device from functioning when needed in cardiac emergencies.

    Product
    CRT-D DTBC2D1 BRAVA IS1/DF1 INTL, Model Number DTBC2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1794-2023·2023-06-28

    FDA Recalls Medtronic Implantable Defibrillators Due to Potential Output Risk

    Medtronic's ICD-VR VISIA AF S implantable defibrillators may fail to deliver energy during high-voltage therapy due to a feedthrough component defect. The FDA Class I recall affects 194 units distributed nationwide and worldwide.

    Product
    ICD-VR VISIA AF S US/OUS DF4, Model Number DVAC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide