The Recall Desk

State

Nevada product recalls

72,992 recalls have nationwide distribution and so reach Nevada. 5,693 additional recalls listed Nevada specifically in their distribution scope.

What the numbers show

Total
5,693
Critical
951
Severe
1,477
Most recent
2026-09-25

Of the 5,693 recalls naming Nevada we have scored against our rubric, 951 (17%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,477 (26%) are Severe.

These figures cover only the 5,693 recalls whose distribution notice names Nevada specifically. The 72,992 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

8176–8200 of 78685

  • HighFDA (Devices)·Z-1021-2025·2025-02-05

    Design Options Combined Spinal/Epidural Tray Recalled for Connector Defect

    B. Braun Medical is recalling 1,090 units of the Design Options Combined Spinal/Epidural Tray due to an incorrect catheter connector that cannot securely connect to the epidural catheter, preventing medication administration.

    Product
    Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. Model/Catalog Number: 530159
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1052-2025·2025-02-05

    Karl Storz Cysto-Urethroscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling 221 cysto-urethroscopes because their reprocessing instructions have not been FDA-approved. Medical facilities should stop using these devices until receiving updated instructions from the manufacturer.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27030KA, Cysto-Urethroscope, NON STERILE, CE
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1040-2025·2025-02-05

    Medline Procedure Kits Recalled for Surgical Mask Bioburden Testing Failure

    Medline has recalled 749 procedure kits containing surgical masks that failed to meet bioburden testing standards under EN 14683. The affected kits were distributed nationwide.

    Product
    Medline procedure kits labeled as: 1) Port Access Kit, Reorder number CVI5130; 2) VANTEX 7FR 3L 20CM CVC STOCK+ BUNDLE US. Reorder number STCVC01USGV; general surgical tray
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1060-2025·2025-02-05

    Karl Storz Ureteroscope Recall Due to Unapproved Reprocessing Instructions

    Karl Storz ureteroscopes are being recalled because their instructions contain reprocessing procedures that have not been reviewed or approved by the FDA. The recall affects 3,069 units distributed nationwide.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27010K, Ureteroscope, 7 Fr., 34 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1023-2025·2025-02-05

    CyberKnife Treatment System: Internal Snap Ring Detachment Risk

    Snap rings may detach from internal shafts in CyberKnife Treatment Delivery Systems, potentially causing uncontrolled rotation on treatment axes. This affects 212 units distributed worldwide for stereotactic radiosurgery.

    Product
    CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0212-2025·2025-02-05

    Duloxetine Delayed-Release Capsules recalled nationwide for nitrosamine impurity

    Duloxetine Delayed-Release Capsules USP 20 mg are being recalled nationwide due to presence of nitrosamine impurity above FDA acceptable limits. No adverse events have been reported.

    Product
    Duloxetine Delayed-Release Capsules USP, 20 mg, packaged in a) 30 Unit Doses (3x10 blister packs) NDC 0904-7043-04 and b) 100 Unit Doses (10x10 blister packs) NDC 0904-7043-61, Rx only, Packaged and Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268 USA.
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1065-2025·2025-02-05

    Alphatec Calibrate CCX Interbody Implants Recalled for Post-Operative Collapse

    Alphatec Spine is recalling Calibrate CCX spinal implants used in fusion procedures after receiving complaints of post-operative implant collapse. Approximately 1,511 implants were distributed nationwide.

    Product
    Calibrate CCX Interbody System for spinal fusion procedures: Implants: (1) Calibrate CCX-O, 7 x 10 x 25mm 5 degrees, Part #325-07102505, nonsterile; (2) Calibrate CCX-O, 7 x 10 x 25mm 10 degrees, Part #325-07102510, nonsterile; (3) Calibrate CCX-O, 7 x 10 x 25mm 15 d
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1058-2025·2025-02-05

    Karl Storz Uretero-Renoscope recalled for unapproved reprocessing instructions

    Karl Storz Endoskope Uretero-Renoscope (Model 27011L) is recalled due to Instructions for Use containing reprocessing procedures that have not been FDA-reviewed or approved. The recall affects 5 units distributed nationwide.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27011L, Uretero-Renoscope, 8 Fr., 6¿, 43cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1062-2025·2025-02-05

    Karl Storz Ureteroscope Model 27001K Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling its Ureteroscope (Model 27001K) because the Instructions for Use contain reprocessing procedures that have not been reviewed or approved by the FDA. All 185 units distributed nationwide are affected.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27001K, Ureteroscope, 9.5 Fr., 6¿, 34 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1022-2025·2025-02-05

    Medtronic Percutaneous Reference Pins May Not Fit Properly

    Medtronic is recalling sterile percutaneous reference pins used in image-guided surgery because the pins may not fit into the patient reference frame or adapter, potentially affecting proper surgical tracking.

    Product
    Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 97733236 The Sterile Percutaneous Reference Pin Set is intended to be used to place and remove a pin percutaneously into bony patient anatomy for rigid attachment of a pat
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1030-2025·2025-02-05

    AXIOS Stent and Delivery System Recalled for Improper Expansion Risk

    The AXIOS Stent and Delivery System is being recalled because the outer sheath can detach and prevent proper stent expansion, requiring device replacement during the procedure.

    Product
    AXIOS Stent and Delivery System 10mmx10mm UPN: M00553740
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1039-2025·2025-02-05

    Knee Implant Component Recalled for Incorrect Metal Post Assembly

    Zimmer, Inc. is recalling 155 units of NexGen LCCK Legacy knee implant components because the metal support post was assembled incorrectly in some units. The defective assembly could compromise implant function.

    Product
    NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0455-2025·2025-02-05

    Alpha Test supplement capsules recalled for unapproved cathine ingredient

    Iovate Health Sciences is recalling Alpha Test 120-count capsules containing cathine, an unapproved ingredient not permitted in dietary supplements. Affected products were distributed across the United States and internationally.

    Product
    Alpha Test 120 count Black labeled plastic bottle, 12 bottles per case of 120 count capsules each per bottle, Store in a cool, dry place (60 F to 80 F), UPC 6-31656-60676-8, SKU0001191, SKU0002075
    Category
    Drug
    Distribution
    30 states
  • HighFDA (Devices)·Z-1066-2025·2025-02-05

    Infinity Central Station speakers may fail to produce critical patient alarms

    Edifier speakers in Draeger's Infinity Central Station may not produce audible alarms when in analog mode, potentially delaying patient treatment. Approximately 1,013 units are affected worldwide.

    Product
    Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of dat
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1050-2025·2025-02-05

    BD Pyxis MedStation Door Latch Assembly Failures in Automated Dispensing Cabinets

    The FDA is recalling approximately 33,589 BD Pyxis MedStation automated dispensing cabinets due to door latch assembly failures that may leave doors open, potentially delaying medication access and enabling medication diversion.

    Product
    BD Pyxis MedStation ES AUX Tower REF 343 BD Pyxis " MedStation 4000 AUX Tower REF 314 The BD Pyxis" MedStation ES and Pyxis" MedStation 4000 are automated dispensing cabinets (ADC) that are intended to securely store and dispense medications to a qualified and authorized he
    Category
    Medical Device
    Distribution
    54 states
  • HighFDA (Devices)·Z-1035-2025·2025-02-05

    Stent system outer sheath may detach, preventing proper expansion

    Boston Scientific's HOT AXIOS Stent has a reported defect where the outer sheath distal black tip can detach, preventing proper stent expansion and requiring device exchange.

    Product
    HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553550
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1027-2025·2025-02-05

    AXIOS Stent Delivery System Outer Sheath May Detach During Use

    The AXIOS Stent with Electrocautery-Enhanced Delivery System may have a defective outer sheath that can detach and prevent proper stent expansion, requiring device replacement during the procedure.

    Product
    AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553670
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1032-2025·2025-02-05

    Outer Sheath Detachment in HOT AXIOS Stent Delivery System

    Boston Scientific's HOT AXIOS stent delivery system may have an outer sheath that detaches and prevents proper stent expansion. The defective tip can remain around the stent, requiring device exchange during the procedure.

    Product
    HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553520
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1057-2025·2025-02-05

    Karl Storz Uretero-Renoscope with Unapproved Reprocessing Instructions Recalled

    Karl Storz Endoscopy is recalling 59 units of its uretero-renoscope due to instructions for use containing reprocessing procedures that the FDA has not reviewed or approved for safety and efficacy.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27001L, Uretero-Renoscope, 8 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1028-2025·2025-02-05

    AXIOS Stent Distal Tip Detachment Prevents Proper Expansion

    Boston Scientific is recalling AXIOS Stent devices due to outer sheath distal tip detachment that can prevent proper stent expansion and require device replacement during the procedure.

    Product
    AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553680
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1038-2025·2025-02-05

    Knee implant articular surface recalled for incorrect metal post assembly

    Zimmer, Inc. is recalling NexGen LCCK knee implant articular surface components due to incorrect assembly of the metal support post. The defect was identified internally in a small number of units.

    Product
    NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/C,D - Height 14 mm, Item Number 00-5994-022-14
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1068-2025·2025-02-05

    Implantable Vagus Nerve Stimulation Device May Stop Delivering Therapy

    LivaNova's SenTiva DUO vagus nerve stimulation device may stop delivering therapy due to an internal component issue. The recall affects 84 units distributed worldwide.

    Product
    Brand Name: SenTiva DUO" Product Name: VNS Therapy¿ SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-D Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, le
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1034-2025·2025-02-05

    Boston Scientific HOT AXIOS Stent Delivery Systems Recalled for Component Detachment

    Boston Scientific is recalling HOT AXIOS Stent systems because the outer sheath's tip can detach and prevent proper stent expansion, requiring device replacement during procedures.

    Product
    HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1051-2025·2025-02-05

    Karl Storz Uretero-Renoscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling a uretero-renoscope surgical device (REF 27003L) because the reprocessing instructions in the product documentation have not been reviewed or approved by the FDA.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27003L, MICHEL Uretero-Renoscope, 9.5 Fr., 43cm, RxONLY, CE0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1041-2025·2025-02-05

    Medline procedure kits recalled for surgical masks failing sterilization tests

    Medline SHORT TERM INFUSAPORT ACCESS KIT with reorder number DYNDC3188 is recalled because included surgical masks failed bioburden testing, potentially compromising sterility during procedures.

    Product
    Medline procedure kits labeled as: 1) SHORT TERM INFUSAPORT ACCESS KIT, Reorder number DYNDC3188; surgery tray
    Category
    Medical Device
    Distribution
    Distributed nationwide