  • HighFDA (Devices)·Z-1749-2026·2026-04-08

    Paradigm REAL-Time insulin pump delivery varies with pump height

    Medtronic is recalling Paradigm REAL-Time insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0420-2026·2026-04-08

    Prazosin hydrochloride tablets recalled over nitrosamine impurity above limits

    Five lots of Prazosin Hydrochloride tablets are being recalled after detection of the nitrosamine impurity N-nitroso-prazosin impurity C above acceptable limits.

    Product
    PRAZOSIN HYDROCHLORIDE — PRAZOSIN HYDROCHLORIDE (PRAZOSIN HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0409-2026·2026-04-08

    Quality Choice irritation relief eye drops recalled over sterility assurance

    Quality Choice Eye Drops Irritation Relief, containing tetrahydrozoline HCl and zinc sulfate, are being recalled for lack of assurance of sterility. 182,424 bottles are affected.

    Product
    QUALITY CHOICE EYE DROPS IRRITATION RELIEF — QUALITY CHOICE EYE DROPS IRRITATION RELIEF (TETRAHYDROZOLINE HCL, ZINC SULFATE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1742-2026·2026-04-08

    MiniMed 770G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 770G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1683-2026·2026-04-08

    CPX 4 medium height tissue expander infusion needles may be blunt

    Mentor is recalling CPX 4 US Breast Tissue Expanders with Suture Tabs in the smooth medium height range because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-107MH, SCPX-117MH,SCPX-127MH, SCPX-135MH, SCPX-146MH, SCPX-156MH Smooth Medium Height Tissue Expander, 275cc, 350cc, 450cc, 550cc, 650cc, 800cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0414-2026·2026-04-08

    GNP redness and dry eye relief drops recalled over sterility assurance

    GNP Eye Drops Redness and Dry Eye Relief, containing glycerin and naphazoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    LEADER EYE DROPS REDNESS RELIEF — LEADER EYE DROPS REDNESS RELIEF (GLYCERIN, NAPHAZOLINE HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1752-2026·2026-04-08

    Paradigm REAL-Time Veo pump delivery varies with pump height changes

    Medtronic is recalling Paradigm REAL-Time Veo insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0419-2026·2026-04-08

    Prazosin hydrochloride 100-count bottles recalled over nitrosamine impurity

    A further quantity of Prazosin Hydrochloride tablets is being recalled after detection of the nitrosamine impurity N-nitroso-prazosin impurity C above acceptable limits.

    Product
    PRAZOSIN HYDROCHLORIDE — PRAZOSIN HYDROCHLORIDE (PRAZOSIN HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1755-2026·2026-04-08

    Burlington Medical kilts may lose radiation attenuation over time

    Burlington Medical is recalling kilts made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Kilt
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1674-2026·2026-04-08

    Philips Azurion 3 M15 X-ray may not start from foot switch

    Philips is recalling the Azurion 3 M15 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1707-2026·2026-04-08

    Medline thoracic robot kits contain recalled Olympus biopsy valves

    Medline is recalling Thoracic Robots kits, model DYNJ908777B, because they contain recalled Olympus biopsy valves where rubber fragments may detach from the valve slit.

    Product
    THORACIC ROBOTS, DYNJ908777B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1697-2026·2026-04-08

    da Vinci double fenestrated grasper recalled over broken grip cables

    Intuitive Surgical is recalling da Vinci S and Si Double Fenestrated Grasper Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including bl
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1751-2026·2026-04-08

    MiniMed 530G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 530G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 530G Insulin Pump (MMT-551, MMT-751)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1662-2026·2026-04-08

    Philips Allura Xper FD10/10 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10/10 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0415-2026·2026-04-08

    Rugby and Walgreens lubricant eye drops recalled over sterility assurance

    K.C. Pharmaceuticals is recalling Sterile Eye Drops Soothing Tears, sold as Rugby Lubricating Eye Drops and Walgreens Lubricant Eye Drops, for lack of assurance of sterility.

    Product
    STERILE EYE DROPS SOOTHING TEARS (polyethylene glycol 400 0.4% and propylene glycol 0.3%), 0.5 OZ 0.5 fl oz (15 mL) bottles; a) Rugby LUBRICATING EYE DROPS, Distributed by: Rugby Laboratories, Livonia, MI 48152, NDC 0536-1219-94; b) Walgreens, Lubricant Eye Drops, DISTRIBUTED BY
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1741-2026·2026-04-08

    MiniMed 780G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 780G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1690-2026·2026-04-08

    Medline ENFit G-tube connectors may leak from dimensional specification failure

    Medline is recalling ENFit G-tube connectors because they were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices.

    Product
    ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1695-2026·2026-04-08

    Olympus SOLTIVE Pro laser power supply may fail producing smoke

    Olympus is recalling SOLTIVE Pro SuperPulsed Laser Systems because a potential defect in the 24V power supply module may render the system inoperable and produce smoke or a burning smell.

    Product
    Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tis
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1655-2026·2026-04-08

    Paradise renal denervation catheter distributed despite being marked for scrapping

    Recor Medical is recalling one lot of the Paradise Ultrasound Renal Denervation System Catheter because a nonconforming product that should have been scrapped was distributed instead.

    Product
    Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1711-2026·2026-04-08

    Medline anterior hip pack contains recalled Olympus biopsy valves

    Medline is recalling the Anterior Hip Pack because it contains recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    MEDLINE ANTERIOR HIP PACK DYNJ64672B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1702-2026·2026-04-08

    VOLT locking screw taps swapped between lots carry incorrect thread

    The 2.4 VOLT Tap Locking Screw Tap is being recalled after a full lot swap with the Cortical Screw Tap left the bone taps with an incorrect thread.

    Product
    2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0413-2026·2026-04-08

    GoodSense original formula eye drops recalled over sterility assurance

    GoodSense Eye Drops Original Formula, containing tetrahydrozoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    GOODSENSE EYE DROPS ORIGINAL FORMULA — GOODSENSE EYE DROPS ORIGINAL FORMULA (TETRAHYDROZOLINE HCL)
    Category
    Drug
    Distribution
    Distributed nationwide