73,363 recalls have nationwide distribution and so reach New Jersey. 9,769 additional recalls listed New Jersey specifically in their distribution scope.

What the numbers show

Total
9,769
Critical
1,249
Severe
2,486
Most recent
2026-10-07

Of the 9,769 recalls naming New Jersey we have scored against our rubric, 1,249 (13%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,486 (25%) are Severe.

These figures cover only the 9,769 recalls whose distribution notice names New Jersey specifically. The 73,363 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

17351–17375 of 83132

  • SevereFDA (Devices)·Z-1772-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough design issue. This could prevent life-saving electrical therapy from being delivered.

    Product
    CRTD DTMC2D4 COMPIA MRI OUS DF4, Model Number DTMC2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1748-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapy During Critical Events

    Certain Medtronic implantable defibrillators may fail to deliver energy during high-voltage therapy due to a feedthrough manufacturing defect. 523 units distributed worldwide are affected.

    Product
    CRT-D DTBB2QQ VIVA QUAD S IS4/DF4 INTL, Model Number DTBB2QQ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1789-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    Medtronic implantable defibrillators with certain feedthroughs may fail to deliver high-voltage therapy when needed. This rare malfunction could prevent life-saving electrical treatment.

    Product
    ICD-VR DVAB1D1 VISIA AF US IS1/DF1, Model Number DVAB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1749-2023·2023-06-28

    Implantable Cardiac Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Certain Medtronic implantable cardiac defibrillators with specific glassed feedthrough components may deliver reduced energy during high-voltage therapy. This could prevent the device from functioning when needed in cardiac emergencies.

    Product
    CRT-D DTBC2D1 BRAVA IS1/DF1 INTL, Model Number DTBC2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1794-2023·2023-06-28

    FDA Recalls Medtronic Implantable Defibrillators Due to Potential Output Risk

    Medtronic's ICD-VR VISIA AF S implantable defibrillators may fail to deliver energy during high-voltage therapy due to a feedthrough component defect. The FDA Class I recall affects 194 units distributed nationwide and worldwide.

    Product
    ICD-VR VISIA AF S US/OUS DF4, Model Number DVAC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1717-2023·2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced or No Therapy Output

    Certain Medtronic implantable cardioverter defibrillators and cardiac resynchronization defibrillators may fail to deliver energy during high-voltage therapy due to a specific feedthrough design issue. The recall affects 3,801 units distributed in the US and worldwide.

    Product
    ICD-DR DDMD3D1 PRIMO MRI, Model Number DDMD3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1785-2023·2023-06-28

    Medtronic implantable defibrillators recalled for potential therapy delivery failure

    Medtronic has recalled certain implantable cardioverter defibrillators due to a rare risk of reduced or no energy output during high-voltage therapy.

    Product
    CRTD CROME HF MRI IS1 DF1, Model Number DTPC2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1773-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Critical Therapy

    Medtronic implantable defibrillators may produce reduced or no energy output during high-voltage therapy delivery. Affected devices were manufactured with a specific feedthrough component.

    Product
    CRTD DTMC2QQ COMPIA MRI QUAD OUS DF4, Model Number DTMC2QQ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1724-2023·2023-06-28

    Medtronic ICD Devices Potential Energy Output Failure During Therapy

    Medtronic's implantable cardioverter defibrillators (ICDs) may produce reduced or no energy output during high voltage therapy due to a feedthrough defect. Patients with affected lot numbers should contact their healthcare provider.

    Product
    ICD DDPA2D4G COBALT XT DR MRI DF4 GOLD, Model Number DDPA2D4G; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1816-2023·2023-06-28

    Medtronic Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic recalls certain ICD and CRT-D defibrillators due to a rare potential for reduced or no-energy output during high-voltage therapy. Affected devices were manufactured with a specific glassed feedthrough component.

    Product
    ICD-VR DVME3D1 MIRRO MRI, Model Number DVME3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1817-2023·2023-06-28

    Implantable Cardiac Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic's ICD-VR DVME3D4 defibrillators are recalled due to a rare risk of reduced or no energy output during high-voltage therapy. The defect involves a specific feedthrough component and affects approximately 14,989 units distributed nationwide and worldwide.

    Product
    ICD-VR DVME3D4 MIRRO MRI, Model Number DVME3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1798-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail To Deliver Therapy; Class I Recall

    Medtronic implantable cardioverter defibrillators (ICDs) may fail to deliver high-voltage therapy due to a manufacturing defect in a specific feedthrough component. 447 units are affected worldwide.

    Product
    ICD-VR DVBB2D4 EVERA XT DF4 INTL, Model Number DVBB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1750-2023·2023-06-28

    Medtronic Implantable Defibrillators Recall Due to Potential Therapy Failure

    Medtronic is recalling certain implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) due to a rare potential for reduced or no-energy output during high voltage therapy. The recalled devices contain a specific feedthrough component.

    Product
    CRT-D DTBC2D4 BRAVA IS1/DF4 INTL, Model Number DTBC2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1711-2023·2023-06-28

    Medtronic Cardiac Defibrillators May Fail to Deliver Therapy

    Certain Medtronic implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) may produce reduced or no energy output during high voltage therapy due to a defective feedthrough component.

    Product
    ICD-DR DDMB1D1 EVERA MRI XT US DF1, Model Number DDMB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1781-2023·2023-06-28

    Implantable Cardiac Devices Risk Therapy Failure From Feedthrough Defect

    Certain Medtronic implantable cardioverter defibrillators may fail to deliver high-voltage therapy due to a defective feedthrough component. The FDA classified this as a Class I recall affecting 2,884 devices.

    Product
    CRTD COBALT HF MRI IS1 DF1, Model Number DTPB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1793-2023·2023-06-28

    Implantable Defibrillators May Produce Reduced or No Energy Output

    Medtronic implantable cardioverter defibrillators may produce reduced or no-energy output during high voltage therapy due to a defective feedthrough component. The defect affects 176 units distributed worldwide.

    Product
    ICD-VR VISIA AF S US/OUS IS1/DF1, Model Number DVAC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1730-2023·2023-06-28

    Implantable defibrillators may fail to deliver emergency therapy

    Medtronic is recalling CRT-D defibrillators that may fail to deliver electrical therapy during life-threatening heart rhythms due to a defective feedthrough component. Affected devices were distributed nationwide and worldwide.

    Product
    CRT-D DTBA1D1G DF1 VIVA XT US GOLD CTD, Model Number DTBA1D1G; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1723-2023·2023-06-28

    ICD COBALT XT Defibrillators Recalled for Potential Reduced Energy Output

    Medtronic recalled 22,114 ICD COBALT XT defibrillators due to a rare potential for reduced or no energy output during high-voltage therapy delivery caused by a specific feedthrough defect.

    Product
    ICD COBALT XT DR MRI IS1 DF4, Model Number DDPA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1805-2023·2023-06-28

    Implantable Cardioverter Defibrillators at Risk of Reduced Energy Output

    Medtronic ICDs and CRT-Ds with a specific feedthrough defect may produce reduced or no energy output during high-voltage therapy. The recall involves approximately 4,546 units distributed nationwide and worldwide.

    Product
    ICD-VR DVFB2D1 VISIA AF MRI XT OUS DF1, Model Number DVFB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1721-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver High Voltage Therapy

    Certain Medtronic implantable cardioverter defibrillators may deliver reduced or no energy during high voltage therapy due to a defective feedthrough. This defect could prevent the device from properly treating life-threatening heart rhythms.

    Product
    ICD COBALT XT DR MRI IS1 DF1, Model Number DDPA2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1811-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapy Due to Feedthrough Defect

    Medtronic's ICD-VR DVMB2D4 implantable defibrillators may fail to deliver electrical therapy due to a defect in the feedthrough component. The recall affects 6,477 units distributed worldwide.

    Product
    ICD-VR DVMB2D4 EVERA MRI XT DF4 INTL, Model Number DVMB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1737-2023·2023-06-28

    Medtronic Implantable Heart Defibrillators Risk Therapy Delivery Failure

    Medtronic CRT-D and ICD implantable defibrillators may fail to deliver therapy due to a rare defect in a specific feedthrough component. The recalled devices carry a potential for reduced or no-energy output during high-voltage heart rhythm treatment.

    Product
    CRT-D DTBA2D4 VIVA XT IS1/DF4 INTL, Model Number DTBA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1778-2023·2023-06-28

    Implantable Defibrillators May Deliver Reduced Energy During Therapy

    Medtronic ICDs and CRT-Ds with a specific feedthrough may deliver reduced or no energy during high voltage therapy. Approximately 2,669 units are affected in the US and worldwide.

    Product
    CRTD COBALT XT HF QUAD MRI IS4 DF1, Model Number DTPA2Q1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1707-2023·2023-06-28

    Implantable Defibrillators Risk Reduced or No Energy Output During Therapy

    Certain Medtronic implantable defibrillators may fail to deliver high voltage therapy due to a feedthrough defect. The FDA Class I recall affects 2,282 units distributed nationwide and worldwide.

    Product
    ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1764-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapy in Rare Cases

    Medtronic implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators may rarely fail to deliver high-voltage therapy due to a manufacturing defect in the device feedthrough.

    Product
    CRTD DTMB1QQ AMPLIA MRI QUAD US DF4, Model Number DTMB1QQ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide