The Recall Desk

State

New Jersey product recalls

73,219 recalls have nationwide distribution and so reach New Jersey. 9,768 additional recalls listed New Jersey specifically in their distribution scope.

What the numbers show

Total
9,768
Critical
1,249
Severe
2,486
Most recent
2026-09-30

Of the 9,768 recalls naming New Jersey we have scored against our rubric, 1,249 (13%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,486 (25%) are Severe.

These figures cover only the 9,768 recalls whose distribution notice names New Jersey specifically. The 73,219 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

17176–17200 of 82987

  • SevereFDA (Devices)·Z-1731-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Potential Therapy Failure

    Medtronic is recalling 661 CRT-D implantable defibrillators due to a rare manufacturing defect that may reduce or eliminate energy output during therapy delivery. Affected devices may fail to deliver necessary electrical treatment during cardiac emergencies.

    Product
    CRT-D DTBA1D4 VIVA XT IS1/DF4 US, Model Number DTBA1D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1728-2023·2023-06-28

    Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic is recalling certain ICD and CRT-D devices that may fail to deliver high-voltage therapy due to a manufacturing defect in the device feedthrough. The recall affects 2,018 units distributed nationwide and worldwide.

    Product
    ICD CROME DR MRI IS1 DF4, Model Number DDPC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1819-2023·2023-06-28

    Implantable defibrillators may fail to deliver high-voltage therapy during treatment

    Certain Medtronic ICD and CRT-D devices with a specific feedthrough may fail to deliver energy during high-voltage therapy. This rare defect affects 9,325 units worldwide and could compromise cardiac protection.

    Product
    ICD COBALT XT VR MRI DF4, Model Number DVPA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1777-2023·2023-06-28

    Implantable Defibrillators Recalled for Rare Risk of Energy Output Failure

    Medtronic implantable defibrillators may deliver reduced or no energy during therapy due to a rare electrical defect. Approximately 1,914 units with a specific feedthrough component are affected and distributed worldwide.

    Product
    CRTD COBALT XT HF MRI IS1 DF4, Model Number DTPA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1710-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Certain Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may produce reduced or no energy output during high-voltage therapy, potentially compromising treatment effectiveness.

    Product
    ICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1746-2023·2023-06-28

    Medtronic Implantable Cardiac Defibrillators Risk Therapy Delivery Failure

    Medtronic is recalling certain CRT-D and ICD models due to a potential feedthrough defect that may prevent energy delivery during high-voltage therapy. The FDA Class I recall affects 435 units distributed in the United States and worldwide.

    Product
    CRT-D DTBB2D4 VIVA S IS1/DF4 INTL, Model Number DTBB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1782-2023·2023-06-28

    Implantable Defibrillators May Fail During High-Voltage Cardiac Therapy

    Medtronic implantable defibrillators may fail to deliver energy during high-voltage cardiac therapy due to a defective glassed feedthrough component. The recall affects 2,968 units distributed nationwide and worldwide.

    Product
    CRTD COBALT HF MRI IS1 DF4, Model Number DTPB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1796-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High-Voltage Therapy

    Medtronic recalls implantable cardioverter defibrillators due to rare potential for reduced or no-energy output during high-voltage therapy, which could prevent life-saving defibrillation.

    Product
    ICD-VR DVBB1D4 EVERA XT DF4 US, Model Number DVBB1D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1732-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Energy During Therapy

    Medtronic implantable cardioverter defibrillators (CRT-Ds) may fail to deliver adequate energy during high-voltage treatment due to a manufacturing defect in the feedthrough component, affecting 373 units.

    Product
    CRT-D DTBA1Q1 VIVA QUAD XT IS4/DF1 US, Model Number DTBA1Q1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1745-2023·2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic's implantable cardioverter defibrillators may produce reduced or no energy during high voltage therapy due to a feedthrough defect. A total of 1,076 units are affected worldwide.

    Product
    CRT-D DTBB2D1 VIVA S IS1/DF1 INTL, Model Number DTBB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1801-2023·2023-06-28

    Medtronic implantable cardioverter defibrillators recalled for potential reduced-output defect

    Medtronic has recalled certain ICD and CRT-D cardiac devices due to a rare risk of reduced or no energy output during high-voltage therapy. The affected models contain a specific glassed feedthrough component.

    Product
    ICD-VR DVEX2E4 EV ICD OUS EV4, Model Number DVEX2E4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1753-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver High-Voltage Therapy Energy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough defect. Approximately 11,794 units worldwide are affected.

    Product
    CRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1821-2023·2023-06-28

    Implantable cardiac defibrillators recalled for potential therapy failure risk

    Medtronic recalls ICD COBALT VR MRI DF4 defibrillators due to a rare potential for reduced or no energy output during high-voltage therapy. The defect affects devices manufactured with a specific glassed feedthrough component.

    Product
    ICD COBALT VR MRI DF4, Model Number DVPB3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1786-2023·2023-06-28

    Implantable Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic implantable defibrillators with a specific feedthrough may fail to deliver adequate energy during high voltage therapy. Approximately 476 affected units distributed nationwide and worldwide.

    Product
    CRTD CROME HF MRI IS1 DF4, Model Number DTPC2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1791-2023·2023-06-28

    Implantable Defibrillators With Defective Feedthrough May Fail to Deliver Therapy

    Certain Medtronic implantable defibrillators may produce reduced or no energy output during high-voltage therapy due to a defective feedthrough component affecting 71 units.

    Product
    ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1815-2023·2023-06-28

    Implantable Defibrillators With Defective Feedthrough May Fail During Critical Therapy

    Medtronic is recalling 14,270 implantable cardioverter defibrillators with a specific feedthrough that may produce reduced or no energy output during high-voltage therapy. The affected devices may fail to deliver critical treatment when needed.

    Product
    ICD-VR DVMD3D4 PRIMO MRI, Model Number DVMD3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1709-2023·2023-06-28

    Medtronic Implantable Defibrillators May Fail to Deliver Therapy

    Certain Medtronic ICDs manufactured with a specific feedthrough may fail to deliver high-voltage therapy. Patients should contact their healthcare provider immediately.

    Product
    ICD-DR DDBC3D1 EVERA S IS1/DF1 GLOB, Model Number DDBC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1736-2023·2023-06-28

    Medtronic Implantable Defibrillators May Fail to Deliver Therapy

    Medtronic is recalling certain implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators due to a rare risk of reduced or no energy output during high voltage therapy.

    Product
    CRTD DTBA2D1G VIVA XT IS1 DF1 OUS GOLD, Model Number DTBA2D1G; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1716-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Potential Energy Output Failure

    Medtronic is recalling certain implantable cardioverter defibrillators and cardiac resynchronization therapy devices due to a rare potential for reduced or no energy output during high voltage therapy. Approximately 37,547 units are affected.

    Product
    ICD-DR DDMC3D4 EVERA MRI S IS-1/DF4 GLOB, Model Number DDMC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Drugs)·D-0893-2023·2023-06-28

    Prescription Drug Recall: Ziprasidone Cartons Contain Wrong Medication

    Unit Dose cartons labeled as Ziprasidone Hydrochloride 20 mg were found to contain Dronabinol 2.5 mg instead. Lot T04769 (Exp. 2024/12) was distributed nationwide.

    Product
    ZIPRASIDONE — ZIPRASIDONE (ZIPRASIDONE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1705-2023·2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic is recalling 522 implantable cardioverter defibrillators that may deliver reduced or no energy during high-voltage therapy due to a feedthrough defect.

    Product
    ICD-DR DDBB1D4 EVERA XT IS1/DF4 US, Model Number DDBB1D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1729-2023·2023-06-28

    Medtronic CRT-D Implants Recalled for Rare Energy Output Failure Risk

    Medtronic is recalling certain CRT-D (cardiac resynchronization therapy defibrillator) models due to a rare risk of reduced or no energy output during high-voltage therapy. The FDA has classified this as a Class I recall affecting 1,473 units.

    Product
    CRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1734-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Reduced Therapy Output

    Medtronic is recalling 14 units of CRTD VIVA QUAD XT implantable cardioverter defibrillators due to rare potential for reduced or no energy output during high voltage therapy. Affected devices were manufactured with a specific feedthrough defect.

    Product
    CRTD VIVA QUAD XT CUST PARY GOLD, Model Number DTBA1QQPG; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1713-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may experience reduced or no energy output during high-voltage therapy, potentially preventing treatment of life-threatening heart rhythms.

    Product
    ICD-DR DDMB2D1 EVERA MRI XT OUS DF1, Model Number DDMB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1720-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High Voltage Therapy

    Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during high voltage therapy due to a specific feedthrough manufacturing defect. Approximately 8,219 units were affected worldwide.

    Product
    ICD-DR DDME3D4 MIRRO MRI, Model Number DDME3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide