The Recall Desk

State

North Carolina product recalls

73,100 recalls have nationwide distribution and so reach North Carolina. 10,294 additional recalls listed North Carolina specifically in their distribution scope.

What the numbers show

Total
10,294
Critical
1,486
Severe
2,536
Most recent
2026-09-23

Of the 10,294 recalls naming North Carolina we have scored against our rubric, 1,486 (14%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,536 (25%) are Severe.

These figures cover only the 10,294 recalls whose distribution notice names North Carolina specifically. The 73,100 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

10176–10200 of 83394

  • HighFDA (Drugs)·D-0666-2024·2024-10-02

    Herbal Dietary Supplement Recalled for Undeclared Pharmaceuticals

    Infla-650 Herbal Dietary Supplement is being recalled because it contains undeclared acetaminophen, diclofenac, and phenylbutazone. The product was marketed without FDA approval and poses a risk of drug interactions and adverse effects.

    Product
    Infla-650 Herbal Dietary Supplement, 700mg, packaged in pouches containing 60 capsules, Distributed by: Guru Inc., 4370 Lawrenceville Hwy NW, P.O. Box 2204, Lilburn, GA 30047-1100, USA. Ph.: 872 216 1566, http://www.dalayurevda.com, Made in India, UPC: 042535355019
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3267-2024·2024-10-02

    Boston Scientific esophageal stents recalled due to catheter tip detachment risk

    Boston Scientific is recalling Wallflex and Agile esophageal stent systems due to potential detachment of the delivery catheter tip during stent placement. These devices are used to maintain patency in esophageal strictures caused by malignant tumors or esophageal fistulas.

    Product
    WALLFLEX FC ESO STENT RMV 18X153MM-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3301-2024·2024-10-02

    Cardinal Health Recalls Salem Sump Stomach Tubes Due to Valve Breakage

    Cardinal Health is recalling Salem Sump Dual Lumen Stomach Tubes because the Anti-Reflux Valve may break due to improper use. Labeling changes were made to address the issue.

    Product
    (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888264929, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888264945, Sterile; (c) CardinalHealth Salem Sump…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0665-2024·2024-10-02

    Refresh LACRI-LUBE Eye Ointment Recalled for Potential Tube Seal Breach

    Allergen is recalling specific lots of Refresh LACRI-LUBE eye ointment (763,426 tubes) due to potential breaches in the tube seal that may compromise sterility. Consumers should verify their product lot number against the recall list.

    Product
    Refresh LACRI-LUBE, (42.5% Mineral Oil, 57.3% White Petrolatum) Lubricant Eye Ointment, Net wt, 0.12 oz (3.5g) Tube, Distributed by: Allergen, an AbbVie company, Madison, NJ 07940, UPC code: 300230312042
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3224-2024·2024-10-02

    Karl Storz Nephroscope Recall Due to Unapproved Reprocessing Instructions

    Karl Storz Endoscopy is recalling 5 nephroscopes because their instructions for use list reprocessing methods not reviewed or approved by the FDA.

    Product
    Karl Storz - Endoskope , REF: 27830KA, Nephroscope for MIP M, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3294-2024·2024-10-02

    Boston Scientific Recalls Agile Esophageal Stent Systems Due to Catheter Tip Detachment

    Boston Scientific is recalling Agile Esophageal Stent Systems because the delivery catheter tip may detach. The recall affects 10 units distributed worldwide.

    Product
    AGILE ESO FC RMV 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3227-2024·2024-10-02

    Endovascular AAA Packs Recalled Due to Sterilization Process Defect

    American Contract Systems is recalling 93 units of Endovascular AAA Packs due to inability to confirm sterilization assurance requirements were met following an environmental monitoring failure.

    Product
    Endovascular AAA Pack, ESED52B; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0668-2024·2024-10-02

    Prescription immunosuppressant tablets recalled due to failed dissolution specifications

    Ascend Laboratories recalled 117,493 bottles of Mycophenolic Acid tablets nationwide because the tablets failed to meet dissolution specifications. Patients taking this medication should consult their doctor.

    Product
    MYCOPHENOLIC ACID — MYCOPHENOLIC ACID (MYCOPHENOLIC ACID)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3300-2024·2024-10-02

    Custodiol HTK Solution Recalled for Potential Particle Contamination

    Custodiol HTK Solution is subject to a Class II recall due to the potential for particles in the solution. The affected product consists of 48 cartons distributed in North Carolina.

    Product
    Custodiol HTK Solution, 500ml bottle. 10 per carton.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3286-2024·2024-10-02

    WallFlex and Agile esophageal stent systems recalled for delivery catheter tip detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems due to potential for delivery catheter tip detachment during insertion.

    Product
    WALLFLEX PC ESOPH STENT 23/28MM X 155MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3246-2024·2024-10-02

    BD Phoenix M50 may misidentify E. coli bacteria in clinical samples

    The BD Phoenix M50 instruments may misidentify Escherichia coli (E. coli) in clinical samples, potentially causing misdiagnosis and inappropriate treatment. About 4,295 affected units were distributed worldwide.

    Product
    BD Phoenix M50 instrument -Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD Phoenix System provides rapid results for most aerobic and facultative anaerobic Grampositive bacteria. Catalog…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3209-2024·2024-10-02

    Ultrasonic Cleaner Solution Recalled for Bacterial Contamination and Infection Risk

    Ultrasonic cleaner solutions contaminated with pathogenic bacteria may cause serious infections if used in medical or dental procedures, and may leak or become bloated. Patients and healthcare workers handling the products are at risk.

    Product
    Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultrasonic Cleaner Solution, Models: GVUS128-1, GVUS128-1EA, GVUS550-1, GVUS24-1, GVUS24-1EA, JGVUS128-1, VGVUSNP128-1, VGVUSNP128-1EA
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0664-2024·2024-10-02

    FDA Recalls Refresh P.M. Eye Ointment Due to Potential Sterility Breach

    AbbVie is recalling Refresh P.M. lubricant eye ointment due to potential breaches in tube seals that may compromise product sterility. Approximately 2.4 million tubes were distributed nationwide and internationally.

    Product
    Refresh P.M., (Mineral Oil 42.5%, White Petrolatum 57.3%) Lubricant Eye Ointment, Net wt, 0.12 oz (3.5g) Tube, Distributed by: Allergen, an AbbVie company, Madison, NJ 07940, UPC code: 300230667043,
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3217-2024·2024-10-02

    Siemens Diagnostic Ultrasound Systems Recalled for Incorrect Unit Conversion

    Siemens ACUSON Maple ultrasound systems may perform incorrect unit conversion in software, producing values 10 times smaller than actual. This could lead to misdiagnosis or inappropriate patient management.

    Product
    ACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software
    Category
    Medical Device
    Distribution
    42 states
  • HighFDA (Devices)·Z-3221-2024·2024-10-02

    Medical endoscope recalled for unapproved sterilization instructions

    Karl Storz Hopkins Telescope 6 endoscopes are recalled because their Instructions for Use contain reprocessing methods that lack FDA review and approval.

    Product
    Karl Storz - Endoskope , REF: 27293AA, Hopkins Telescope 6, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3223-2024·2024-10-02

    Karl Storz Hopkins Telescope 6 Endoscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz Hopkins Telescope 6 endoscopes (59 units, nationwide) are recalled because their reprocessing instructions lack FDA review and approval for safety and efficacy.

    Product
    Karl Storz - Endoskope , REF: 27295AA, Hopkins Telescope 6, 18 Fr, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3229-2024·2024-10-02

    Arthroscopy Pack Recalled Due to Unconfirmed Sterilization Assurance

    American Contract Systems is recalling 48 Arthroscopy Packs because humidity readings were out of specification, preventing confirmation of sterilization.

    Product
    Arthroscopy Pack, MSKA45A; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3226-2024·2024-10-02

    Medical Pacemaker Pack Recalled Due to Sterilization Assurance Failure

    American Contract Systems is recalling the PACEMAKER PACK (Lot 2405112) due to an inoperable chart recorder that prevented confirmation of sterilization requirements. Affected units were distributed in MO, MN, MA, OH, and NE.

    Product
    PACEMAKER PACK, SLCV01J; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3277-2024·2024-10-02

    Boston Scientific Esophageal Stent Systems Recalled for Catheter Detachment Risk

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems due to potential delivery catheter tip detachment. The recall affects 79 units distributed worldwide.

    Product
    WALLFLEX ESOPHAGEAL FC 18/23-25MMX15CM- Agile Esophageal Partially Covered and Fully Covered OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3199-2024·2024-10-02

    MRI Diagnostic Device Recalled for Loose Assembly Screws Blocking Equipment

    Philips Ingenia Elition S MRI machines (10 units) are recalled because assembly screws may come loose and interfere with the horizontal tabletop, potentially delaying diagnosis and causing patient anxiety.

    Product
    Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782150 2) 782106
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3238-2024·2024-10-02

    Vaginal Delivery Medical Convenience Kit Recalled for Unconfirmed Sterilization Assurance

    American Contract Systems, Inc. is recalling the Vaginal Delivery ASDV25B medical convenience kit due to an inoperable chart recorder that prevented confirmation of sterilization assurance. The device may not meet required sterilization standards.

    Product
    Vaginal Delivery, ASDV25B; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3257-2024·2024-10-02

    Biodesign Fistula Plug implants recalled for premature expiration defect

    Cook Biotech is recalling 130 units of Biodesign Fistula Plug C-FPS-0.7-2 surgical implants because the devices expire prior to the expiration date printed on the labeling. Devices were distributed worldwide including the US and 11 other countries.

    Product
    Biodesign Fistula Plug C-FPS-0.7-2 - For implantation to reinforce soft tissue for repair of recto-vaginal or anorectal fistulas. Order Number (GPN): G46373
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3244-2024·2024-10-02

    Pacemaker COPM11B Medical Kits Recalled for Sterilization Verification Issue

    American Contract Systems is recalling 3 units of the COPM11B pacemaker medical kits due to inability to confirm sterilization requirements were met. An inoperable chart recorder caused humidity readings to exceed specifications during manufacturing.

    Product
    Pacemaker, COPM11B; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3266-2024·2024-10-02

    WallFlex and Agile Esophageal Stent Systems Recalled for Catheter Tip Detachment

    Boston Scientific recalls WallFlex and Agile Esophageal Stent Systems due to potential delivery catheter tip detachment. The affected stents treat esophageal strictures and may cause procedural complications if detachment occurs.

    Product
    WALLFLEX FC ESO STENT RMV 18X123MM-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3248-2024·2024-10-02

    Hologic Recalls Selenia Dimensions and 3Dimensions Breast Imaging Software

    Hologic is recalling Selenia Dimensions and 3Dimensions software due to a defect that may cause blurry images and potential misdiagnosis.

    Product
    Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as follows Model/Catalog Number: 1. ASY-16824 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.1 SYSTEMS 2. ASY-16825…
    Category
    Medical Device
    Distribution
    Distributed nationwide