2501–2525 of 76486

  • HighFDA (Devices)·Z-1667-2026·2026-04-08

    Philips Allura Xper FD10 OR Table X-ray may not start reliably

    Philips is recalling the Allura Xper FD10 OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1682-2026·2026-04-08

    CPX 4 low height tissue expander infusion needles may be blunt

    Mentor is recalling the CPX 4 US Breast Tissue Expander with Suture Tabs, smooth low height 650 cc, because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-157LH Smooth Low Height Tissue Expander, 650cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1670-2026·2026-04-08

    Philips Allura Xper FD20/20 OR Table X-ray may not start

    Philips is recalling the Allura Xper FD20/20 OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/20 OR Table; System Code: 722039;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1677-2026·2026-04-08

    Philips Azurion 7 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 7 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1669-2026·2026-04-08

    Philips Allura Xper FD20/20 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/20; System Code: 722038;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1664-2026·2026-04-08

    Philips Allura Xper FD20 Biplane X-ray may not start reliably

    Philips is recalling the Allura Xper FD20 Biplane because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1673-2026·2026-04-08

    Philips Azurion 3 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 3 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1660-2026·2026-04-08

    Philips Allura Xper FD10F X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10F because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10F; System Code: 722002;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1740-2026·2026-04-08

    MiniMed 630G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 630G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1679-2026·2026-04-08

    Philips Azurion 5 M12 X-ray may not start from foot switch

    Philips is recalling the Azurion 5 M12 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 5 M12; System Code: (1)722227, (2)722231;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1674-2026·2026-04-08

    Philips Azurion 3 M15 X-ray may not start from foot switch

    Philips is recalling the Azurion 3 M15 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1694-2026·2026-04-08

    Olympus SOLTIVE Premium laser power supply may fail producing smoke

    Olympus is recalling SOLTIVE Premium SuperPulsed Laser Systems because a potential defect in the 24V power supply module may render the system inoperable and produce smoke or a burning smell.

    Product
    Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1739-2026·2026-04-08

    MiniMed 670G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 670G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1680-2026·2026-04-08

    Philips Azurion 5 M20 X-ray may not start from foot switch

    Philips is recalling the Azurion 5 M20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1701-2026·2026-04-08

    da Vinci permanent cautery hook recalled over frayed pitch cables

    Intuitive Surgical is recalling da Vinci S and Si Permanent Cautery Hook Instruments following increased complaints regarding frayed or broken pitch cables on reusable surgical instruments.

    Product
    Brand Name: da Vinci S, Si Permanent Cautery Hook Instrument Product Name: da Vinci S, Si Permanent Cautery Hook Instrument Model/Catalog Number: 420183 Software Version: NA Product Description: EndoWrist Instruments or da Vinci S/Si Reusable instruments, including blunt and
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0420-2026·2026-04-08

    Prazosin hydrochloride tablets recalled over nitrosamine impurity above limits

    Five lots of Prazosin Hydrochloride tablets are being recalled after detection of the nitrosamine impurity N-nitroso-prazosin impurity C above acceptable limits.

    Product
    PRAZOSIN HYDROCHLORIDE — PRAZOSIN HYDROCHLORIDE (PRAZOSIN HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1666-2026·2026-04-08

    Philips Allura Xper FD20 Biplane OR Table X-ray may not start

    Philips is recalling the Allura Xper FD20 Biplane OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0408-2026·2026-04-08

    Amneal tramadol tablets failed nitrosamine impurity specification at stability

    Amneal Pharmaceuticals is recalling two lots of Tramadol Hydrochloride Tablets 50 mg after a nitrosamine impurity was found out of specification at the 24-month stability interval.

    Product
    TRAMADOL HYDROCHLORIDE — TRAMADOL HYDROCHLORIDE (TRAMADOL HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0413-2026·2026-04-08

    GoodSense original formula eye drops recalled over sterility assurance

    GoodSense Eye Drops Original Formula, containing tetrahydrozoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    GOODSENSE EYE DROPS ORIGINAL FORMULA — GOODSENSE EYE DROPS ORIGINAL FORMULA (TETRAHYDROZOLINE HCL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1757-2026·2026-04-08

    Burlington Medical wrap aprons may lose radiation attenuation over time

    Burlington Medical is recalling wrap aprons made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Wrap Aprons.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1665-2026·2026-04-08

    Philips Allura Xper FD20 OR Table X-ray may not start

    Philips is recalling the Allura Xper FD20 OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1700-2026·2026-04-08

    da Vinci tenaculum forceps recalled over frayed or broken pitch cables

    Intuitive Surgical is recalling da Vinci S and Si Tenaculum Forceps Instruments following increased complaints regarding frayed or broken pitch cables on reusable surgical instruments.

    Product
    Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument Model/Catalog Number: 420207 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp end
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1684-2026·2026-04-08

    CPX 4 tall height tissue expander infusion needles may be blunt

    Mentor is recalling CPX 4 US Breast Tissue Expanders with Suture Tabs in the smooth tall height range because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-113TH, SCPX-123TH, SCPX-140TH, SCPX-146TH Smooth Tall Height Tissue Expander, 350cc, 450cc, 650cc, 750cc
    Category
    Medical Device
    Distribution
    Distributed nationwide