8201–8225 of 79908

  • HighFDA (Devices)·Z-1052-2025·2025-02-05

    Karl Storz Cysto-Urethroscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling 221 cysto-urethroscopes because their reprocessing instructions have not been FDA-approved. Medical facilities should stop using these devices until receiving updated instructions from the manufacturer.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27030KA, Cysto-Urethroscope, NON STERILE, CE
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1028-2025·2025-02-05

    AXIOS Stent Distal Tip Detachment Prevents Proper Expansion

    Boston Scientific is recalling AXIOS Stent devices due to outer sheath distal tip detachment that can prevent proper stent expansion and require device replacement during the procedure.

    Product
    AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553680
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1037-2025·2025-02-05

    HOT AXIOS Stent System Recalled Due to Outer Sheath Detachment Risk

    Boston Scientific is recalling HOT AXIOS Stent systems due to potential outer sheath detachment that can prevent proper stent expansion. This defect may require device replacement during the procedure.

    Product
    HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553570
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1068-2025·2025-02-05

    Implantable Vagus Nerve Stimulation Device May Stop Delivering Therapy

    LivaNova's SenTiva DUO vagus nerve stimulation device may stop delivering therapy due to an internal component issue. The recall affects 84 units distributed worldwide.

    Product
    Brand Name: SenTiva DUO" Product Name: VNS Therapy¿ SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-D Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, le
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1021-2025·2025-02-05

    Design Options Combined Spinal/Epidural Tray Recalled for Connector Defect

    B. Braun Medical is recalling 1,090 units of the Design Options Combined Spinal/Epidural Tray due to an incorrect catheter connector that cannot securely connect to the epidural catheter, preventing medication administration.

    Product
    Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. Model/Catalog Number: 530159
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1029-2025·2025-02-05

    AXIOS Stent Delivery System Outer Sheath Detachment Recall (47 Units)

    Boston Scientific's AXIOS Stent Delivery System (47 units affected) is recalled because the outer sheath can detach and prevent proper stent expansion, requiring device exchange.

    Product
    AXIOS Stent and Electrocautery- Enhanced Delivery System 8mmx8mm UPN: M00553690
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1055-2025·2025-02-05

    Karl Storz Uretero-Renoscope recalled for unapproved reprocessing instructions

    Karl Storz Endoscopy has recalled the KARL STORZ ENDOSKOPE Uretero-Renoscope nationwide because its instructions for use contain reprocessing procedures that lack FDA review and approval for safety and efficacy.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27010L, Uretero-Renoscope, 7 Fr., 43cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1057-2025·2025-02-05

    Karl Storz Uretero-Renoscope with Unapproved Reprocessing Instructions Recalled

    Karl Storz Endoscopy is recalling 59 units of its uretero-renoscope due to instructions for use containing reprocessing procedures that the FDA has not reviewed or approved for safety and efficacy.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27001L, Uretero-Renoscope, 8 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1040-2025·2025-02-05

    Medline Procedure Kits Recalled for Surgical Mask Bioburden Testing Failure

    Medline has recalled 749 procedure kits containing surgical masks that failed to meet bioburden testing standards under EN 14683. The affected kits were distributed nationwide.

    Product
    Medline procedure kits labeled as: 1) Port Access Kit, Reorder number CVI5130; 2) VANTEX 7FR 3L 20CM CVC STOCK+ BUNDLE US. Reorder number STCVC01USGV; general surgical tray
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1023-2025·2025-02-05

    CyberKnife Treatment System: Internal Snap Ring Detachment Risk

    Snap rings may detach from internal shafts in CyberKnife Treatment Delivery Systems, potentially causing uncontrolled rotation on treatment axes. This affects 212 units distributed worldwide for stereotactic radiosurgery.

    Product
    CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1033-2025·2025-02-05

    HOT AXIOS Stent Delivery System Outer Sheath May Detach, Preventing Stent Expansion

    Boston Scientific's HOT AXIOS stent delivery system has a defect where the outer sheath can detach and prevent proper stent expansion, requiring device replacement.

    Product
    HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 8mmx8mm UPN: M00553530
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1031-2025·2025-02-05

    AXIOS Stent Delivery System recalled due to distal sheath detachment

    Boston Scientific is recalling 38 units of AXIOS Stent and Delivery System (15mmx10mm) because the outer sheath's distal tip may detach, preventing proper stent expansion and requiring device replacement.

    Product
    AXIOS Stent and Delivery System 15mmx10mm UPN: M00553750
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1051-2025·2025-02-05

    Karl Storz Uretero-Renoscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling a uretero-renoscope surgical device (REF 27003L) because the reprocessing instructions in the product documentation have not been reviewed or approved by the FDA.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27003L, MICHEL Uretero-Renoscope, 9.5 Fr., 43cm, RxONLY, CE0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1060-2025·2025-02-05

    Karl Storz Ureteroscope Recall Due to Unapproved Reprocessing Instructions

    Karl Storz ureteroscopes are being recalled because their instructions contain reprocessing procedures that have not been reviewed or approved by the FDA. The recall affects 3,069 units distributed nationwide.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27010K, Ureteroscope, 7 Fr., 34 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1025-2025·2025-02-05

    AXIOS Stent Delivery System: Detaching Outer Sheath May Prevent Stent Expansion

    Boston Scientific is recalling 1,144 AXIOS Stent Delivery System units because the outer sheath's distal black tip can detach. If this occurs, the tip may remain around the stent saddle and prevent proper stent expansion, requiring device exchange.

    Product
    AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553650
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1050-2025·2025-02-05

    BD Pyxis MedStation Door Latch Assembly Failures in Automated Dispensing Cabinets

    The FDA is recalling approximately 33,589 BD Pyxis MedStation automated dispensing cabinets due to door latch assembly failures that may leave doors open, potentially delaying medication access and enabling medication diversion.

    Product
    BD Pyxis MedStation ES AUX Tower REF 343 BD Pyxis " MedStation 4000 AUX Tower REF 314 The BD Pyxis" MedStation ES and Pyxis" MedStation 4000 are automated dispensing cabinets (ADC) that are intended to securely store and dispense medications to a qualified and authorized he
    Category
    Medical Device
    Distribution
    54 states
  • HighFDA (Devices)·Z-1035-2025·2025-02-05

    Stent system outer sheath may detach, preventing proper expansion

    Boston Scientific's HOT AXIOS Stent has a reported defect where the outer sheath distal black tip can detach, preventing proper stent expansion and requiring device exchange.

    Product
    HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553550
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1027-2025·2025-02-05

    AXIOS Stent Delivery System Outer Sheath May Detach During Use

    The AXIOS Stent with Electrocautery-Enhanced Delivery System may have a defective outer sheath that can detach and prevent proper stent expansion, requiring device replacement during the procedure.

    Product
    AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553670
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0456-2025·2025-02-05

    BakeMark USA Recalls Bavarian Creme Filling Due to Thermal Processing Deviation

    BakeMark USA recalls WESTCO 3065 Golden Bavarian Creme Filling due to a thermal processing deviation that may result in inadequate heat treatment. The product was distributed to 22 states.

    Product
    WESTCO 3065 GOLDEN BAVARIAN CREME FILLING ARTIFICIALLY FLAVORED Ingredients: Water, Sugar, Modified Food Starch, Contains 2% or Less of: Artificial Flavor, Salt, Color (Titanium Dioxide, Yellow 5, Yellow 6), Preservatives (Sodium Benzoate, Sorbic Acid, Potassium Sorbate, Phosphor
    Category
    Food
    Distribution
    22 states
  • HighFDA (Food)·F-0455-2025·2025-02-05

    Alpha Test supplement capsules recalled for unapproved cathine ingredient

    Iovate Health Sciences is recalling Alpha Test 120-count capsules containing cathine, an unapproved ingredient not permitted in dietary supplements. Affected products were distributed across the United States and internationally.

    Product
    Alpha Test 120 count Black labeled plastic bottle, 12 bottles per case of 120 count capsules each per bottle, Store in a cool, dry place (60 F to 80 F), UPC 6-31656-60676-8, SKU0001191, SKU0002075
    Category
    Drug
    Distribution
    30 states
  • HighFDA (Devices)·Z-1062-2025·2025-02-05

    Karl Storz Ureteroscope Model 27001K Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling its Ureteroscope (Model 27001K) because the Instructions for Use contain reprocessing procedures that have not been reviewed or approved by the FDA. All 185 units distributed nationwide are affected.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27001K, Ureteroscope, 9.5 Fr., 6¿, 34 cm, NON STERILE, RxONLY, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1053-2025·2025-02-05

    Karl Storz Pediatric Cysto-Urethroscope Recalled for Unapproved Reprocessing Instructions

    Karl Storz is recalling 185 Pediatric Cysto-Urethroscope instruments nationwide. The product instructions contain reprocessing methods that have not been reviewed or approved by the FDA for safety and effectiveness.

    Product
    KARL STORZ - ENDOSKOPE, REF: 27030KB, Pediatric Cysto-Urethroscope, NON STERILE, CE
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1067-2025·2025-02-05

    Implantable VNS Therapy Generators May Stop Delivering Treatment

    LivaNova is recalling approximately 8,223 SenTiva VNS Therapy generators due to a potential internal component failure that may cause the implanted device to stop delivering seizure therapy. Affected patients should contact their physician.

    Product
    Brand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and exte
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1041-2025·2025-02-05

    Medline procedure kits recalled for surgical masks failing sterilization tests

    Medline SHORT TERM INFUSAPORT ACCESS KIT with reorder number DYNDC3188 is recalled because included surgical masks failed bioburden testing, potentially compromising sterility during procedures.

    Product
    Medline procedure kits labeled as: 1) SHORT TERM INFUSAPORT ACCESS KIT, Reorder number DYNDC3188; surgery tray
    Category
    Medical Device
    Distribution
    Distributed nationwide