DxI 800 with Spot B analyzer motor current resets below requirement
DxI 800 Access Immunoassay Analyzers with Spot B are being recalled because software overwrites the factory PnP X-motor current setting, causing increased motion errors. 451 systems are affected.
- Product
- DxI 800 Access Immunoassay Analyzer W/Spot B, Part Number A71456
- Category
- Medical Device
- Distribution
- Distributed nationwide