Hawaii product recalls

72,992 recalls have nationwide distribution and so reach Hawaii. 2,767 additional recalls listed Hawaii specifically in their distribution scope.

What the numbers show

Total
2,767
Critical
333
Severe
567
Most recent
2026-09-16

Of the 2,767 recalls naming Hawaii we have scored against our rubric, 333 (12%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 567 (20%) are Severe.

These figures cover only the 2,767 recalls whose distribution notice names Hawaii specifically. The 72,992 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

7326–7350 of 75759

  • HighFDA (Devices)·Z-1367-2025·2025-03-26

    Medline Proxima Sterile Surgical Drapes Recalled for Packaging Breach Risk

    Medline Industries is recalling Proxima Drape sterile surgical drapes due to potential packaging breach that could compromise sterility. The recall affects 16 units distributed nationwide.

    Product
    Proxima Drape labeled as FACE LIFT. Sterile surgical drape.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1375-2025·2025-03-26

    BD Pyxis Medication Dispensing Systems Software Sync Error Recall

    BD Pyxis medication dispensing systems may disconnect from their server due to a software error, preventing them from receiving medication updates. This could lead to incorrect medication administration or delayed access to medications.

    Product
    BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES Desktop PC, REF: 1156-00, CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01, CII Safe ES Desktop PC, W-less SCNR, REF: 107-255-01,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1383-2025·2025-03-26

    Microstream CO2 Sampling Line May Not Disconnect From Intubation Tube

    Microstream intubated CO2 sampling lines may be difficult or impossible to disconnect from endotracheal tubes. This could delay necessary procedures and cause unintended extubation, potentially leading to respiratory failure.

    Product
    Microstream Instructions for Use and Part Number used with - Product Description, REF: Microstream Luer Intubated Adult Pediatric, PT00156355 - Microstream Luer Adult-Pediatric Intubated CO2 Sampling Line 2m/6.5ft Short-term use: Procedural/Emergency, MLAI
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1336-2025·2025-03-26

    C-Section Surgical Packs Recalled Over Sterilization Assurance Failure

    American Contract Systems is recalling C-Section surgical packs (252 kits) after being unable to confirm sterilization assurance. Lack of verified sterilization could delay or prolong surgical treatment.

    Product
    C-SECTION PACK, Model Nos LLCS44F LLCS44F-01
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1376-2025·2025-03-26

    CO2 Sampling Lines May Fail to Disconnect During Neonatal Care

    Microstream CO2 sampling lines and airway adapters may be difficult or impossible to disconnect from endotracheal tubes, potentially delaying treatment or causing unintended extubation. The defect affects neonatal and pediatric patients worldwide.

    Product
    Microstream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 006324; MICROSTREAM
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0280-2025·2025-03-26

    Prasugrel 5 mg Tablets Recalled Due to Failed Dissolution Specifications

    Mylan Pharmaceuticals is recalling specific lots of Prasugrel 5 mg tablets due to failed dissolution specifications. The affected tablets may not dissolve properly in the body.

    Product
    PRASUGREL — PRASUGREL (PRASUGREL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1362-2025·2025-03-26

    Vein Tray Models UTVT21C Unable to Confirm Sterilization Assurance

    American Contract Systems Inc is recalling certain Vein Tray models because sterilization assurance requirements could not be confirmed. The affected units may lack functionality and could lead to delayed treatment.

    Product
    VEIN TRAY, Model No UTVT21C UTVT21C02 UTVT21C-03
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1348-2025·2025-03-26

    Delivery Pack Medical Device Sterilization Assurance Not Confirmed

    American Contract Systems Inc is recalling Delivery Pack medical devices (Models SGDV22D-01 and SGDV22E) because sterilization assurance requirements were not confirmed. Unsterilized devices may not function properly and could delay treatment.

    Product
    DELIVERY PACK , Model No SGDV22D-01 SGDV22E
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1360-2025·2025-03-26

    Lumbar Spine Pack Recalled Due to Sterilization Assurance Failure

    American Contract Systems Inc is recalling 106 SPINE LUMBAR PACK kits (Models UTSL72AA, UTSL72AB) because the manufacturer cannot confirm sterilization assurance requirements were met, potentially causing functionality loss and treatment delays.

    Product
    SPINE LUMBAR PACK , Model No UTSL72AA UTSL72AB
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·25V190000·2025-03-26

    Micro Bird Recalls G5 School Buses for Missing Seat Reinforcements

    Micro Bird is recalling certain G5 school buses because seat reinforcement supports may be missing, which could cause seats to detach in a crash.

    Product
    MICRO BIRD — MICRO BIRD G5 SCHOOL BUS
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1350-2025·2025-03-26

    Laceration Tray Kits Recalled Due to Sterilization Assurance Failure

    American Contract Systems Inc recalls 690 laceration tray kits due to inability to confirm sterilization requirements were met. The trays may lack functionality and delay patient treatment.

    Product
    LACERATION TRAY , Model No SGLA07A SGLA07B-01 SGLA07B-02 SGLA07B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0287-2025·2025-03-26

    FDA Recalls Prasugrel Tablets for Failed Dissolution Specifications

    Golden State Medical Supply Inc. is recalling 2,601 bottles of Prasugrel 5 mg tablets due to failed dissolution specifications. The tablets may not dissolve properly, potentially affecting drug absorption.

    Product
    PRASUGREL — PRASUGREL (PRASUGREL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1358-2025·2025-03-26

    Vascular Pack Sterilization Assurance Requirements Cannot Be Confirmed

    American Contract Systems cannot confirm Major Vascular Pack surgical kits were properly sterilized. Sterilization assurance failure could result in loss of functionality and delayed patient treatment.

    Product
    MAJOR VASCULAR PACK , Model No UTMV78X-01 UTMV78Y UTMV78Y-01 UTMV78X-01 UTMV78Y-02
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1365-2025·2025-03-26

    Sterile Surgical Drape Recalled Due to Potential Packaging Breach

    Medline Industries is recalling 114 units of Proxima Drape surgical drapes due to a potential breach in pouch packaging that could compromise sterility. Affected units were distributed nationwide.

    Product
    Proxima Drape labeled as ENT-PLASTICS PACK. Sterile surgical drape.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1352-2025·2025-03-26

    Laparoscopy Surgical Packs Recalled Due to Unconfirmed Sterilization

    American Contract Systems Inc is recalling 237 laparoscopy surgical packs because sterilization assurance could not be confirmed. The packs may experience loss of functionality and require treatment delays.

    Product
    LAPAROSCOPY PACK , Model No SFLA02A SFLA02B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1379-2025·2025-03-26

    CO2 Sampling and Airway Adapter Devices May Be Difficult to Disconnect

    Oridion Medical's CO2 sampling and airway adapter devices may be difficult to disconnect from endotracheal tubes, potentially causing treatment delays and serious respiratory complications in neonatal, infant, and pediatric patients.

    Product
    Microstream Instructions for Use and Part Number used with - Product Description, REF: Airway Adapter and CO2 Sampling Line, 01158 - Microstream Luer Adult-Pediatric Airway Adapter, 010989; Omnistream CO2 Sampling Line X25, 010991
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1343-2025·2025-03-26

    PICC Line Catheters Recalled for Sterilization Assurance Issues

    American Contract Systems is recalling PICC line catheters because sterilization assurance requirements could not be confirmed. Unsterilized devices may lose functionality and delay patient treatment.

    Product
    CENTRAL LINE PICC , Model No LLCL44H LLCL44H-01
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0286-2025·2025-03-26

    Injectable Drug Recalled for Stability Issues at Expiration

    Amgen recalls 258,750 vials of Neupogen (filgrastim) injection due to stability concerns. Stability data shows products may be out of specification at expiration; patients should consult healthcare providers about affected lots.

    Product
    NEUPOGEN — NEUPOGEN (FILGRASTIM)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1364-2025·2025-03-26

    Surgical Drape Packs Recalled for Potential Sterility Breach in Packaging

    Medline Industries is recalling Proxima Sterile Surgical Drape packs due to a potential breach in the pouch packaging that could lead to loss of sterility. The recall affects 1,490 units distributed nationwide.

    Product
    Proxima Drape labeled as PACK,BASIC,SIRUS,V,5/CS; PACK,BASIC IV,SIRUS,5/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,UNIVERSAL II,SIRUS,7/CS; PACK,LAPAROTOMY,NO GOWN,10/CS; PAC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1386-2025·2025-03-26

    Access 2 Immunoassay Analyzer motor alignment issue may delay results

    Beckman Coulter is recalling Access 2 Immunoassay Analyzers due to sample motor misalignment that may delay patient test result reporting. The recall affects 873 units distributed worldwide.

    Product
    Access 2 Immunoassay Analyzer, Catalog Number 81600N
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1372-2025·2025-03-26

    Modular Stem Orthopedic Implant Recalled Due to Incorrect Product Identifier

    37 units of a modular stem orthopedic implant (Product Code 880-601/11) have been recalled due to an incorrect GTIN product identifier on the carton label, which belongs to a different product. No illnesses or injuries have been reported.

    Product
    Modular Stem cylindrical, Tilastan¿, press-fit cementless, L= 80 mm, DIA = 11 mm, Product Code: 880-601/11
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0282-2025·2025-03-26

    Ciprofloxacin Ophthalmic Solution Recalled Nationwide for Defective Container Cap

    Ciprofloxacin Ophthalmic Solution is being recalled because the cap spike is lodged in the bottle nozzle, preventing patients from accessing the medication. Affected lot 084A067 was distributed nationwide.

    Product
    CIPROFLOXACIN — CIPROFLOXACIN (CIPROFLOXACIN)
    Category
    Drug
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0288-2025·2025-03-26

    Medline Alcohol Prep Pads Recalled for Subpotent Isopropyl Alcohol

    Medline Industries has voluntarily recalled over 6.6 million alcohol prep pads nationwide because the product contains less isopropyl alcohol than labeled. The subpotent product was first reported on March 26, 2025.

    Product
    MEDLINE ALCOHOL PREP — MEDLINE ALCOHOL PREP (ISOPROPYL ALCOHOL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1387-2025·2025-03-26

    Beckman Coulter DxI 9000 Analyzer Communication Failure Recall

    Beckman Coulter recalls certain DxI 9000 immunoassay analyzers due to firmware issues that can cause communication failures, potentially delaying patient result reporting.

    Product
    DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1371-2025·2025-03-26

    Helix Elite SARS-CoV-2 QC Material Recall Due to Potential Ct Value Issues

    Microbiologics Inc is recalling Helix Elite Inactivated SARS-CoV-2 quality control material (Lot HE0065-29) because the A549 human cell target may produce late Ct values, potentially affecting test result validity.

    Product
    Helix Elite" Inactivated Standard: Inactivated SARS-CoV-2 Whole Virus (Pellet), Multi-analyte, unassayed quality control material, Catalog Number HE0065N
    Category
    Medical Device
    Distribution
    Distributed nationwide