Guam product recalls

72,919 recalls have nationwide distribution and so reach Guam. 385 additional recalls listed Guam specifically in their distribution scope.

What the numbers show

Total
385
Critical
14
Severe
37
Most recent
2026-01-28

Of the 385 recalls naming Guam we have scored against our rubric, 14 (4%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 37 (10%) are Severe.

These figures cover only the 385 recalls whose distribution notice names Guam specifically. The 72,919 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

2451–2475 of 73304

  • HighFDA (Drugs)·D-0410-2026·2026-04-08

    Quality Choice moisturizing relief eye drops recalled over sterility assurance

    Quality Choice Moisturizing Relief Eye Drops, containing dextran 70, polyethylene glycol 400, povidone and tetrahydrozoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    QUALITY CHOICE MOISTURIZING RELIEF EYE DROPS — QUALITY CHOICE MOISTURIZING RELIEF EYE DROPS (DEXTRAN 70, POLYETHYLENE GLYCOL 400, POVIDONE, TETRAHYDROZOLINE HCL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1707-2026·2026-04-08

    Medline thoracic robot kits contain recalled Olympus biopsy valves

    Medline is recalling Thoracic Robots kits, model DYNJ908777B, because they contain recalled Olympus biopsy valves where rubber fragments may detach from the valve slit.

    Product
    THORACIC ROBOTS, DYNJ908777B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1683-2026·2026-04-08

    CPX 4 medium height tissue expander infusion needles may be blunt

    Mentor is recalling CPX 4 US Breast Tissue Expanders with Suture Tabs in the smooth medium height range because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-107MH, SCPX-117MH,SCPX-127MH, SCPX-135MH, SCPX-146MH, SCPX-156MH Smooth Medium Height Tissue Expander, 275cc, 350cc, 450cc, 550cc, 650cc, 800cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1743-2026·2026-04-08

    MiniMed 640G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 640G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1741-2026·2026-04-08

    MiniMed 780G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 780G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0413-2026·2026-04-08

    GoodSense original formula eye drops recalled over sterility assurance

    GoodSense Eye Drops Original Formula, containing tetrahydrozoline hydrochloride, are being recalled for lack of assurance of sterility.

    Product
    GOODSENSE EYE DROPS ORIGINAL FORMULA — GOODSENSE EYE DROPS ORIGINAL FORMULA (TETRAHYDROZOLINE HCL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1674-2026·2026-04-08

    Philips Azurion 3 M15 X-ray may not start from foot switch

    Philips is recalling the Azurion 3 M15 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1709-2026·2026-04-08

    Medline laryngoscopy bronchoscopy kits contain Olympus valves shedding rubber fragments

    Medline is recalling four bronchoscopy and laryngoscopy kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1760-2026·2026-04-08

    Burlington Medical caps may lose radiation attenuation over time

    Burlington Medical is recalling protective caps made with Xenolite 800 attenuation material because of the potential for attenuation degradation over time.

    Product
    Burlington Medical, Caps.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1749-2026·2026-04-08

    Paradigm REAL-Time insulin pump delivery varies with pump height

    Medtronic is recalling Paradigm REAL-Time insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1687-2026·2026-04-08

    Artoura ultra high profile expander infusion needles may be blunt

    Mentor is recalling Artoura Breast Tissue Expanders in the smooth ultra high profile range because the infusion sets may contain a dull or blunt needle tip.

    Product
    Artoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc, 700cc, 850cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1699-2026·2026-04-08

    da Vinci grasping retractor instruments recalled over broken grip cables

    Intuitive Surgical is recalling da Vinci S and Si Grasping Retractor Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Grasping Retractor Instrument Product Name: da Vinci S, Si Grasping Retractor Instrument Model/Catalog Number: 420278 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp en
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0408-2026·2026-04-08

    Amneal tramadol tablets failed nitrosamine impurity specification at stability

    Amneal Pharmaceuticals is recalling two lots of Tramadol Hydrochloride Tablets 50 mg after a nitrosamine impurity was found out of specification at the 24-month stability interval.

    Product
    TRAMADOL HYDROCHLORIDE — TRAMADOL HYDROCHLORIDE (TRAMADOL HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1694-2026·2026-04-08

    Olympus SOLTIVE Premium laser power supply may fail producing smoke

    Olympus is recalling SOLTIVE Premium SuperPulsed Laser Systems because a potential defect in the 24V power supply module may render the system inoperable and produce smoke or a burning smell.

    Product
    Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1657-2026·2026-04-08

    Atellica hemoglobin A1c assay may return depressed patient results

    Siemens is recalling Atellica CH Enzymatic Hemoglobin A1c because results may be depressed when the RCRP reagent is processed on the same analyzer, with bias increasing over time.

    Product
    Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1693-2026·2026-04-08

    Sterile radiology procedure kits may have a breached sterile pouch

    One lot of Sterile Radiology Procedure Kits, model DYNDH1491B, is being recalled because of a potential breach in the sterile pouch packaging that may compromise sterility.

    Product
    Sterile Radiology Procedure Kits, Model Number DYNDH1491B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1740-2026·2026-04-08

    MiniMed 630G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 630G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1692-2026·2026-04-08

    Vue Motion V12 may display cine image frames out of sequence

    Philips is recalling Vue Motion V12 because mis-ordered frames during dynamic cine runs may cause image frames to display out of sequence.

    Product
    Vue Motion V12. Product Number: 1017979.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1764-2026·2026-04-08

    Burlington Medical blockers may lose radiation attenuation over time

    Burlington Medical is recalling blockers because of the potential for attenuation degradation over time, decreasing the lifespan of the product.

    Product
    Burlington Medical, Blockers.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1680-2026·2026-04-08

    Philips Azurion 5 M20 X-ray may not start from foot switch

    Philips is recalling the Azurion 5 M20 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1708-2026·2026-04-08

    Medline endoscopy kits contain recalled Olympus biopsy valves shedding fragments

    Medline is recalling BAL and ENDO kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1753-2026·2026-04-08

    Burlington Medical thyroid shields may lose radiation attenuation over time

    Burlington Medical is recalling thyroid shields over the potential for attenuation degradation over time, which decreases the product's usable lifespan. 41,962 units are affected.

    Product
    Burlington Medical, Thyroid Shield.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1702-2026·2026-04-08

    VOLT locking screw taps swapped between lots carry incorrect thread

    The 2.4 VOLT Tap Locking Screw Tap is being recalled after a full lot swap with the Cortical Screw Tap left the bone taps with an incorrect thread.

    Product
    2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.
    Category
    Medical Device
    Distribution
    Distributed nationwide