  • SevereFDA (Devices)·Z-1959-2023·2023-07-05

    Endotracheal Tubes Recalled for Potential 15mm Connector Disconnection

    TELEFLEX is recalling 4,160 endotracheal tubes due to reports of the 15mm connector disconnecting from the tube. Affected units were distributed nationwide.

    Product
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080060
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1865-2023·2023-07-05

    Preformed Endotracheal Tubes Recalled for Connector Disconnection

    Teleflex is recalling 4,146 preformed endotracheal tubes due to reports of 15mm connector disconnection. The affected devices were distributed nationwide.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180050
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1961-2023·2023-07-05

    TELEFLEX Endotracheal Tubes Recalled for Connector Disconnection Risk

    TELEFLEX is recalling 17,950 endotracheal tubes due to reports of the 15mm connector disconnecting from the device. These are critical airway support devices.

    Product
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080070
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1863-2023·2023-07-05

    Teleflex Endotracheal Tube Connector Disconnection Recall

    Teleflex is recalling Preformed AGT Oral Endotracheal Tubes due to reports that the 15mm connector may disconnect from the tube. The recall affects 5,120 units distributed nationwide.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180040
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1851-2023·2023-07-05

    Flexi-Set Endotracheal Tube Connector May Disconnect During Use

    Teleflex is recalling 244,120 Flexi-Set Cuffed Endotracheal Tubes due to reports that the 15mm connector may disconnect from the tube. Affected units were distributed nationwide.

    Product
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504580
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereNHTSA·23E050000·2023-07-05

    Lippert Recalls Curt Axles Due to Improperly Tightened U-Bolts

    Lippert is recalling certain Curt 7000 and 8000 lb axles because improperly tightened U-bolts may allow the axle to move out of position, increasing crash risk.

    Product
    SUSPENSION:REAR:SPRINGS:LEAF SPRING ASSEMBLY:U-BOLT, LEAF SPRING TO AXLE recall
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1998-2023·2023-07-05

    Patient Return Electrode Recalled Due to Burn Reports During Surgery

    Megadyne Medical Products is recalling MEGA SOFT Universal Patient Return Electrodes after reports of patient burns during electrosurgery procedures. The FDA classified this as a Class I recall.

    Product
    MEGADYNE MEGA SOFT Universal Patient Return Electrode. used during electrosurgery
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1877-2023·2023-07-05

    Endotracheal Tube Connector Disconnection Risk, FDA Class I Recall

    Teleflex LLC is recalling approximately 15,810 Preformed AGT Oral Endotracheal Tubes due to reports of 15mm connector disconnection. The affected units were distributed nationwide, including Puerto Rico.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780075
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1868-2023·2023-07-05

    TELEFLEX Preformed Endotracheal Tube Recall for Connector Disconnection Risk

    TELEFLEX is recalling 1,570 units of Preformed AGT Oral Endotracheal Tubes nationwide due to reports of 15mm connector disconnection from the tubes. The disconnection could compromise airway management during patient ventilation.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180065
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1853-2023·2023-07-05

    Flexi-Set Cuffed Endotracheal Tube and Stylet Connector Disconnection Recall

    TELEFLEX LLC is recalling 13,710 units of Flexi-Set Cuffed Endotracheal Tubes due to reports of connector disconnection. The 15mm connector may separate from the tube during use.

    Product
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504590
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1891-2023·2023-07-05

    Teleflex Endotracheal Tubes Recalled for 15mm Connector Disconnection

    Teleflex is recalling approximately 39,550 endotracheal tubes due to reports of 15mm connector disconnection. The FDA Class I recall affects units distributed nationwide including Puerto Rico.

    Product
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082085
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1940-2023·2023-07-05

    Endotracheal Tube 15mm Connector Disconnection Recall

    Teleflex is recalling endotracheal tubes after reports that the 15mm connector may disconnect from the tube. The recall affects 36,135 units distributed nationwide.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780035
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1858-2023·2023-07-05

    Flexi-Set Endotracheal Tubes recalled due to connector disconnection risk

    Teleflex is recalling 12,400 Flexi-Set Uncuffed Endotracheal Tube and Stylet Sets nationwide due to reports of 15mm connector disconnection. The affected batches and lot numbers are identified above.

    Product
    Flexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506545
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1950-2023·2023-07-05

    Endotracheal Tube Connector May Disconnect During Patient Use

    Teleflex recalls preformed AGT oral endotracheal tubes because the 15mm connector may disconnect from the tube, posing risk during patient use.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780090
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1852-2023·2023-07-05

    Flexi-Set Endotracheal Tubes Recalled Due to Connector Disconnection Risk

    TELEFLEX LLC is recalling Flexi-Set Cuffed Endotracheal Tube and Stylet Sets nationwide due to reports of the 15mm connector disconnecting from the tube, posing a serious risk in critical care settings.

    Product
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504585
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1837-2023·2023-07-05

    Slick Set Endotracheal Tubes recalled due to reported 15mm connector disconnection

    Teleflex is recalling Slick Set Uncuffed Endotracheal Tubes due to reports of the 15mm connector disconnecting from the tube. The recall affects 14,160 units distributed nationwide.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170120
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1827-2023·2023-07-05

    Endotracheal Tube and Stylet Set Connector May Disconnect During Use

    TELEFLEX is recalling Slick Set Cuffed Endotracheal Tube and Stylet Sets due to reports that the 15mm connector may disconnect from the tube during use.

    Product
    Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170050
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1995-2023·2023-07-05

    MEGADYNE MEGA SOFT Reusable Patient Return Electrode Recalled for Patient Burn Hazard

    Megadyne Medical Products is recalling 21,100 MEGADYNE MEGA SOFT Reusable Patient Return Electrodes nationwide due to reports of patient burns during electrosurgical procedures.

    Product
    MEGADYNE MEGA SOFT Reusable Patient Return Electrode. used during electrosurgery
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2001-2023·2023-07-05

    Megadyne Mega Soft Universal Patient Return Electrode Recalled for Burn Risk

    Megadyne Medical Products is recalling 21,100 MEGA SOFT electrosurgery patient return electrodes distributed nationwide due to reports of patient burns during surgical procedures.

    Product
    MEGADYNE MEGA SOFT Universal Dual Patient Return Electrode. used during electrosurgery
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1924-2023·2023-07-05

    Teleflex Endotracheal Tubes Recalled for 15mm Connector Disconnection

    Teleflex is recalling Preformed AGT Oral Endotracheal Tubes due to reports of the 15mm connector disconnecting from the tube. Approximately 2,980 units were distributed nationwide.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181065
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1889-2023·2023-07-05

    Endotracheal Tube Connectors Recalled for Potential Disconnection Risk

    TELEFLEX LLC is recalling 300,613 endotracheal tubes due to reports of 15mm connector disconnection. The affected tubes were distributed nationwide and could pose patient safety risks.

    Product
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082075
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1934-2023·2023-07-05

    Endotracheal Tubes With 15mm Connector Disconnection Recalled Nationwide

    TELEFLEX is recalling 25,630 endotracheal tubes nationwide because the 15mm connector can disconnect. The affected products were distributed throughout the United States and Puerto Rico.

    Product
    Endotracheal Tube oral/nasal uncuffed/plain - Murphy, REF 100382025
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1894-2023·2023-07-05

    Endotracheal tubes recalled for 15mm connector disconnection

    Teleflex is recalling 19,250 endotracheal tubes because the 15mm connector may disconnect during use, potentially compromising airway management in patients.

    Product
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082100
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1886-2023·2023-07-05

    Teleflex Endotracheal Tubes Recalled for Connector Disconnection Risk

    Teleflex is recalling approximately 170,610 endotracheal tubes due to reports of 15mm connector disconnection. The disconnection could compromise airway management during mechanical ventilation.

    Product
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082060
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1839-2023·2023-07-05

    Endotracheal Tube Connector Disconnection Risk Prompts Teleflex Recall

    Teleflex is recalling Slick Set Uncuffed Endotracheal Tubes due to reported 15mm connector disconnection. The recall affects 22,400 units distributed nationwide.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170130
    Category
    Medical Device
    Distribution
    Distributed nationwide