Syringe Pump Force Sensor Drift May Cause Detection Delays or False Alarms
Smiths Medical Medfusion syringe pumps may have force sensors that drift out of calibration, causing delays in detecting line occlusions or triggering false alarms. Devices produced before April 2022 are at higher risk.
- Product
- Medfusion Syringe Pump, Model 3500-0600-XX: a) 0600-00; b) 0600-01; c) 0600-50; d) 0600-51; e) 0600-82; f) 0600-249; software versions: v3, v4, v5, and v6
- Category
- Medical Device
- Distribution
- Distributed nationwide