The Recall Desk

State

Florida product recalls

73,104 recalls have nationwide distribution and so reach Florida. 12,855 additional recalls listed Florida specifically in their distribution scope.

What the numbers show

Total
12,855
Critical
1,775
Severe
2,991
Most recent
2026-10-02

Of the 12,855 recalls naming Florida we have scored against our rubric, 1,775 (14%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2,991 (23%) are Severe.

These figures cover only the 12,855 recalls whose distribution notice names Florida specifically. The 73,104 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

10501–10525 of 85959

  • HighFDA (Food)·F-1823-2024·2024-10-02

    M&P Food Products Cheese Blintzes Recalled for Undeclared Wheat

    M&P Food Products Cheese Blintzes are recalled due to undeclared wheat in the product. Consumers with wheat allergies should not consume this frozen food.

    Product
    M&P Food Products Cheese Blintzes; Hand Made Russian Style; 6 Classic; NET WT. 14 OZ. (396g); KEEP FROZEN; INGREDIENTS: Flour, low fat milk, eggs, margarine, sugar, salt; FILLING INGREDIENTS: Farmer Cheese; Cheese Blintzes; UPC: 707163303251
    Category
    Food
    Distribution
    7 states
  • HighFDA (Food)·F-1829-2024·2024-10-02

    M&P Food Vareniki Products Recalled for Undeclared Wheat

    M&P Food Productions is recalling Potato & Mushroom Vareniki because the products contain undeclared wheat, which poses a risk to consumers with wheat allergies.

    Product
    Potato & Mushroom Vareniki; M&P Food Production; NET WT. 14oz; HANDMADE; Keep Frozen; DOUGH INGREDIENTS: Flour, eggs, water, oil (soybean), salt; FILLING INGREDIENTS: Potato, Mushroom, Fried Onion; UPC: 707163026174
    Category
    Food
    Distribution
    7 states
  • HighFDA (Devices)·Z-3274-2024·2024-10-02

    WallFlex Esophageal Stent System Recalled Due to Catheter Tip Detachment Risk

    Boston Scientific recalls WallFlex and Agile esophageal stent systems due to potential delivery catheter tip detachment affecting 77 units distributed worldwide.

    Product
    WALLFLEX FC ESO STENT RMV LL 23X155- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3245-2024·2024-10-02

    FDA Recalls Surgical Instrument Handle That May Break During Use

    The FDA is recalling LINK Universal Handles due to risk of material failure when first used. The surgical instrument may break during surgery, compromising patient safety.

    Product
    LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3205-2024·2024-10-02

    MRI diagnostic device screws may loosen and block tabletop

    Philips is recalling the Ingenia Elition X MRI device due to loose screws in the RF carrier assembly that could block the tabletop and delay diagnosis or cause patient anxiety.

    Product
    SmartPath to Ingenia Elition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782118
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3258-2024·2024-10-02

    Biodesign Recto-Vaginal Fistula Plug expires before labeled expiration date

    Cook Biotech is recalling 3 units of Biodesign Recto-Vaginal Fistula Plug RVP-0.2 because the devices expire prior to the expiration date printed on the labeling.

    Product
    Biodesign Recto-Vaginal Fistula Plug RVP-0.2 - For implantation to reinforce soft tissue for repair of recto-vaginal fistulas. Order Number (GPN): G46601
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3192-2024·2024-10-02

    Biopsy instrument packaging holes compromise sterile barrier integrity

    Argon Medical Devices recalls BioPince Ultra and Tru-Core II biopsy instruments due to holes in packaging trays compromising the sterile barrier, which may allow microorganism introduction and infection risk.

    Product
    BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370-1080-03, 370-1580-01, 370-1580-02, 370-1580-03; Tru-Core II Biopsy…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-1827-2024·2024-10-02

    Sour Cherry Vareniki Recalled for Undeclared Wheat Allergen

    M & P Food Productions is recalling Sour Cherry Vareniki frozen dumplings due to undeclared wheat, a major allergen not listed on the label.

    Product
    Sour Cherry Vareniki; M&P Food Production; NET WT. 14oz; HANDMADE; Keep Frozen; DOUGH INGREDIENTS: Flour, eggs, water, oil (soybean), salt; FILLING INGREDIENTS: Sour Cherry, Sugar; UPC: 707163026150
    Category
    Food
    Distribution
    7 states
  • HighFDA (Devices)·Z-3197-2024·2024-10-02

    MRI Device Ingenia Ambition S Screws May Loosen and Block Tabletop

    Assembly screws in Philips Ingenia Ambition S MRI systems may loosen and protrude, potentially blocking the patient tabletop and delaying diagnosis. Philips is recalling 64 affected units worldwide.

    Product
    Ingenia Ambition S - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782108 2) 781359 3) 782133
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3201-2024·2024-10-02

    Philips MR 5300 MRI System Recalled for Loose Assembly Screws

    Philips is recalling 68 MR 5300 MRI diagnostic systems worldwide because loose screws in the RF carrier assembly could block equipment movement and delay patient diagnosis.

    Product
    MR 5300 - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782110
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3213-2024·2024-10-02

    Cervical disc implant recalled due to foreign contamination during assembly

    Centinel Spine is recalling the Prodisc C SK cervical disc implant due to blue discoloration identified as foreign contamination on the inlay during assembly. The contamination was not fully evaluated in manufacturing validations.

    Product
    prodisc C SK, Total cervical disc replacement, Model/Catalog Number: PDSM5
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3290-2024·2024-10-02

    Esophageal stent systems recalled due to delivery catheter defect

    Boston Scientific has recalled AGILE and WallFlex esophageal stent systems due to potential delivery catheter tip detachment. This defect could occur during stent deployment and affect proper stent placement.

    Product
    AGILE ESO OTW PC 23MM X 15.0CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-1830-2024·2024-10-02

    Farmer Cheese Vareniki recalled for undeclared wheat and milk allergens

    M&P Food Productions is recalling Farmer Cheese Vareniki due to undeclared wheat and milk allergens. The frozen product was distributed across seven states from August 2023 to August 2024.

    Product
    Farmer Cheese Vareniki; M&P Food Production; NET WT. 14oz; HANDMADE; Keep Frozen; DOUGH INGREDIENTS: Flour, eggs, water, oil (soybean), salt; FILLING INGREDIENTS: Farmer Cheese; UPC: 707163026181
    Category
    Food
    Distribution
    7 states
  • HighFDA (Devices)·Z-3275-2024·2024-10-02

    Boston Scientific Esophageal Stents Recalled for Catheter Tip Detachment Risk

    Boston Scientific is recalling WallFlex and Agile Esophageal Stent Systems due to potential delivery catheter tip detachment. The recall affects 231 units distributed worldwide.

    Product
    WALLFLEX ESOPHAGEAL FC 18/23-25MMX10CM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-1826-2024·2024-10-02

    M&P Delicacies Pirozhki Pies Recalled for Undeclared Wheat and Milk

    M&P Food Productions recalls frozen Pirozhki pies distributed across multiple states due to undeclared wheat and milk in the product. Consumers with wheat or milk allergies should not consume these products.

    Product
    M&P Delicacies PIROZHKI (PIES) with CABBAGE; INGREDIENTS: Flour, eggs, butter, vegetable oil, dry yeast, sugar, salt; Filling: cabbage, carrot, sauerkraut, onion, salt, pepper; 1 SERVING READY TO EAT; KEEP FROZEN
    Category
    Food
    Distribution
    7 states
  • HighFDA (Devices)·Z-3299-2024·2024-10-02

    Arrow Arterial Kits Recalled Due to Open Packaging Seals

    Arrow International recalls over 14,000 arterial kits due to reports of open packaging seals. The recall affects multiple models distributed nationwide.

    Product
    Arrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-BW1; (2) Arrow Arterial Access Tray with 0.021" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-CK; (3) Arrow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-1825-2024·2024-10-02

    M&P Delicacies Pirozhki Frozen Pies Recalled for Undeclared Wheat and Milk

    M&P Delicacies frozen pirozhki pies contain undeclared wheat and milk allergens. Consumers with allergies to these ingredients should not consume the product.

    Product
    M&P Delicacies PIROZHKI (PIES) with POTATOES; INGREDIENTS: Flour, eggs, butter, vegetable oil, dry yeast, sugar, salt; Filling: Potatoe, onion, salt, pepper; 1 SERVING READY TO EAT; KEEP FROZEN
    Category
    Food
    Distribution
    7 states
  • HighFDA (Drugs)·D-0660-2024·2024-10-02

    Mupirocin Antibiotic Ointment Recalled Nationwide Due to Subpotency

    Glenmark Pharmaceuticals recalls Mupirocin Ointment nationwide due to subpotency, meaning some lots contain less active ingredient than specified. Affected customers should contact their pharmacy or prescriber.

    Product
    MUPIROCIN — MUPIROCIN (MUPIROCIN)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3302-2024·2024-10-02

    Cardinal Health Salem Sump Stomach Tubes Recalled for Anti-Reflux Valve Breakage

    Cardinal Health is recalling Salem Sump Dual Lumen Stomach Tubes due to reports of anti-reflux valve breakage caused by improper use. Product labeling has been updated.

    Product
    (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888266114, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888266122…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3291-2024·2024-10-02

    Boston Scientific Esophageal Stent Systems Recalled for Catheter Tip Detachment Risk

    Boston Scientific is recalling WallFlex and Agile Esophageal Stent Systems for potential delivery catheter tip detachment during insertion. Eighteen units were distributed worldwide including the United States.

    Product
    AGILE ESO OTW FC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3158-2024·2024-10-02

    Mobile X-Ray System Exposure Settings May Not Apply Correctly

    Fujifilm mobile x-ray systems may fail to apply configured exposure settings or default to adult chest imaging parameters. This could result in inappropriate radiation exposure levels.

    Product
    The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-1828-2024·2024-10-02

    Potato and Onion Vareniki Recalled Due to Undeclared Wheat

    M&P Food Productions is recalling Potato & Onion Vareniki due to undeclared wheat. Consumers with wheat allergies or celiac disease should not consume this product.

    Product
    Potato & Onion Vareniki; M&P Food Production; NET WT. 14oz; HANDMADE; Keep Frozen; DOUGH INGREDIENTS: Flour, eggs, water, oil (soybean), salt; FILLING INGREDIENTS: Potato, Fried Onion; UPC: 707163026167
    Category
    Food
    Distribution
    7 states
  • HighFDA (Devices)·Z-3297-2024·2024-10-02

    CooperSurgical Recalls Origio Sperm Wash Medium Due to Low Sodium Pyruvate

    CooperSurgical is recalling Origio Sperm Wash Medium because low sodium pyruvate levels could compromise gamete development.

    Product
    Origio Sperm Wash Medium Pack of Five 60ml bottles - Intended for washing of spermatozoa. Part Number: 84055060A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3215-2024·2024-10-02

    Siemens ACUSON Juniper ultrasound systems software unit conversion error

    Siemens is recalling ACUSON Juniper ultrasound systems that may perform incorrect unit conversions when configured with non-default settings, potentially producing measurements 10 times smaller than actual values and risking patient misdiagnosis.

    Product
    ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11653093, with software.
    Category
    Medical Device
    Distribution
    42 states