  • HighFDA (Devices)·Z-2214-2025·2025-08-06

    Philips MR 7700 gradient coils may act as a heat source

    MR 7700 systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 82 units are affected.

    Product
    MR 7700 Model Numbers (REF): (1) 782120, (2) 782153;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2232-2025·2025-08-06

    LVP dual-inlet administration sets have potential for external cassette leaks

    LVP Primary Administration Sets, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site, are being recalled because of the potential for external cassette leaks. 19,225 units are affected.

    Product
    LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site (Qty 25). Product Code: SET-0013-25.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2208-2025·2025-08-06

    Philips Ingenia Ambition S gradient coils may act as heat source

    Ingenia Ambition S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 611 units are affected.

    Product
    Ingenia Ambition S Model Numbers (REF): (1) 781359, (2) 782108, (3) 782133 (China ONLY), (4) 782139;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2226-2025·2025-08-06

    Blood MacConkey biplates may carry Listeria monocytogenes surface contamination

    Blood/MacConkey Biplates, product number R02049, are being recalled because the products may contain surface and subsurface contamination of Listeria monocytogenes. 611 units are affected.

    Product
    Blood/MacConkey Biplate 100/PK, Product Number R02049
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2189-2025·2025-08-06

    Philips Allura R8.2.x systems may temporarily lose x-ray imaging functionality

    Philips Allura R8.2.x systems are being recalled because of the potential for temporary loss of imaging (X-ray) functionality due to a software issue. 4,508 units are affected.

    Product
    Philips Allura R8.2.x Systems. Software Verion: R8.2.x. 1. Allura Xper FD10, Model Number: 722026. 2. Allura Xper FD10/10, Model Number: 722027. 3. Allura Xper FD20, Model Number: 722028. 4. Allura Xper FD20/10, Model Number: 722029. 5. Allura Xper FD10 OR Table,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2212-2025·2025-08-06

    Philips MRI Machine Intera 3.0T Quasar Dual Recalled for Fire Risk

    Philips is recalling its Intera 3.0T Quasar Dual MRI system due to a potential fire hazard from gradient coil failures. The recall affects one unit distributed worldwide.

    Product
    Intera 3.0T Quasar Dual Model Number (REF): 781150;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2218-2025·2025-08-06

    Philips MR 7700 upgrade gradient coils may act as a heat source

    Upgrade to MR 7700 systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 26 units are affected.

    Product
    Upgrade to MR 7700 Model Number (REF): 782130;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2203-2025·2025-08-06

    Philips Evolution Upgrade 3.0T gradient coils may act as heat source

    Evolution Upgrade 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 50 units are affected.

    Product
    Evolution Upgrade 3.0T Model Numbers (REF): (1) 782117, (2) 782143;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2211-2025·2025-08-06

    Philips Ingenia Elition X gradient coils may act as heat source

    Ingenia Elition X MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 632 units are affected.

    Product
    Ingenia Elition X Model Numbers (REF): (1) 781358, (2) 782107, (3) 782119, (4) 782136, (5) 782151 (China ONLY);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2205-2025·2025-08-06

    Philips Ingenia 1.5T S gradient coils may act as heat source

    Ingenia 1.5T S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 136 units are affected.

    Product
    Ingenia 1.5T S Model Number (REF): 781347;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2177-2025·2025-08-06

    HarmonyAIR dual flat panel monitor yokes may detach from spring arm

    Dual Flat Panel Monitor accessories for HarmonyAIR A-Series Surgical Lighting Systems are being recalled because the yoke assembly may detach from the spring arm. 167 units are affected.

    Product
    Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2204-2025·2025-08-06

    Philips Ingenia 1.5T gradient coils may act as a heat source

    Ingenia 1.5T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 2,279 units are affected.

    Product
    Ingenia 1.5T Model Numbers (REF): (1) 781315, (2) 718641, (3) 781396, (4) 782101, (5) 782115, (6) 782140;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2200-2025·2025-08-06

    Philips Achieva 3.0T gradient coils may act as a heat source

    Achieva 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 460 units are affected.

    Product
    Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2201-2025·2025-08-06

    Philips Achieva XR gradient coils may act as a heat source

    Achieva XR MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 40 units are affected.

    Product
    Achieva XR; Model Numbers (REF): (1) 781153, (2) 781253;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2206-2025·2025-08-06

    Philips Ingenia 3.0T gradient coils may act as a heat source

    Ingenia 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 1,467 units are affected.

    Product
    Ingenia 3.0T Model Numbers (REF): (1) 781342, (2) 781377, (3) 782103;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2175-2025·2025-08-06

    Inseego dongle kit flash drives carry outdated patient monitor software

    Inseego USB8 4G Dongle Kits for the Tempus Pro Monitor are being recalled because the kit's USB flash drive contains outdated software that could revert the monitor and reintroduce a previous issue. 55 units affected.

    Product
    Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2224-2025·2025-08-06

    Strep Selective II Agar plates may carry Listeria monocytogenes contamination

    Strep Selective II Agar, product number R01859, is being recalled because the products may contain surface and subsurface contamination of Listeria monocytogenes. 371 units are affected.

    Product
    Strep Selective II Agar, Product Number R01859
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2209-2025·2025-08-06

    Philips Ingenia Ambition X gradient coils may act as heat source

    Ingenia Ambition X MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 690 units are affected.

    Product
    Ingenia Ambition X Model Numbers (REF): (1) 781356, (2) 782109, (3) 782138;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2233-2025·2025-08-06

    LVP blood administration sets have potential for external cassette leaks

    LVP Blood Products Administration Sets are being recalled because of the potential for external cassette leaks. 14,280 units are affected.

    Product
    LVP Blood Products Administration Set Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter (Qty 20). Product Code: SET-0014-20.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2234-2025·2025-08-06

    LVP primary administration sets have potential for external cassette leaks

    LVP Primary Administration Sets, Single Outlet Low Sorbing, are being recalled because of the potential for external cassette leaks. 35,925 units are affected.

    Product
    LVP Primary Administration Set, Single Outlet Low Sorbing (Qty 25). Product Code: SET-0021-25.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2217-2025·2025-08-06

    SmartPath to Ingenia Elition X gradient coils may overheat

    SmartPath to Ingenia Elition X systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 33 units affected.

    Product
    SmartPath to Ingenia Elition X Model Numbers (REF): (1) 782118, (2) 782144;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2198-2025·2025-08-06

    CoSense sampling sets may carry residual carbon monoxide from insufficient aeration

    Precision Sampling Sets for the CoSense End-Tidal Carbon Monoxide Monitor are being recalled because residual carbon monoxide may remain in the cannula filter, which may raise ETCOc scores. 2,160 units affected.

    Product
    Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0545-2025·2025-08-06

    Bicillin L-A 2.4 million unit vials recalled after particulates found

    Bicillin L-A (penicillin G benzathine injectable suspension) 2,400,000 units per 4 mL is being recalled because of CGMP deviations, with particulates identified during visual inspection. 19,279 vials affected.

    Product
    BICILLIN L-A — BICILLIN L-A (PENICILLIN G BENZATHINE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2230-2025·2025-08-06

    RAPIDPoint 500 wash waste cartridge lot biases seven blood analytes

    A lot of RAPIDPoint 500 Systems Wash/Waste Cartridges is being recalled because a manufacturing error causes positive and negative bias across seven analytes. 1,477 units are affected.

    Product
    RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2210-2025·2025-08-06

    Philips Ingenia Elition S gradient coils may act as a heat source

    Ingenia Elition S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 242 units are affected.

    Product
    Ingenia Elition S Model Numbers (REF): (1) 781357, (2) 782106, (3) 782137, (4) 782150 (China ONLY);
    Category
    Medical Device
    Distribution
    Distributed nationwide