  • HighFDA (Devices)·Z-3201-2024·2024-10-02

    Philips MR 5300 MRI System Recalled for Loose Assembly Screws

    Philips is recalling 68 MR 5300 MRI diagnostic systems worldwide because loose screws in the RF carrier assembly could block equipment movement and delay patient diagnosis.

    Product
    MR 5300 - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782110
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3258-2024·2024-10-02

    Biodesign Recto-Vaginal Fistula Plug expires before labeled expiration date

    Cook Biotech is recalling 3 units of Biodesign Recto-Vaginal Fistula Plug RVP-0.2 because the devices expire prior to the expiration date printed on the labeling.

    Product
    Biodesign Recto-Vaginal Fistula Plug RVP-0.2 - For implantation to reinforce soft tissue for repair of recto-vaginal fistulas. Order Number (GPN): G46601
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3283-2024·2024-10-02

    Wallflex and Agile Esophageal Stent Systems Recalled for Potential Catheter Tip Detachment

    Boston Scientific is recalling WallFlex and Agile Esophageal Stent Systems due to the potential for delivery catheter tip detachment. The recall affects 39 units distributed worldwide.

    Product
    WALLFLEX PC ESOPH STENT 18/23MM X 153MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophag
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3226-2024·2024-10-02

    Medical Pacemaker Pack Recalled Due to Sterilization Assurance Failure

    American Contract Systems is recalling the PACEMAKER PACK (Lot 2405112) due to an inoperable chart recorder that prevented confirmation of sterilization requirements. Affected units were distributed in MO, MN, MA, OH, and NE.

    Product
    PACEMAKER PACK, SLCV01J; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3191-2024·2024-10-02

    Custom Procedure Kit Sheath Introducers Mislabeled by Size

    Merit Medical Systems is recalling Custom Procedure Kits (REF: K12T-11077) with 7F sheath introducers that are incorrectly labeled as 7.5F, a labeling discrepancy affecting 24 units distributed nationwide.

    Product
    Custom Procedure Kit, REF: K12T-11077
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·24V737000·2024-10-02

    2025 Chevrolet Equinox EV door strikers recalled for possible breakage

    General Motors is recalling 2025 Chevrolet Equinox EV vehicles (along with Blazer and Blazer EV models) because door strikers may not be properly heat-treated and could break, causing doors to open unexpectedly. All affected vehicles remain in dealerships.

    Product
    CHEVROLET — CHEVROLET BLAZER EV, EQUINOX EV, BLAZER
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3288-2024·2024-10-02

    Boston Scientific Recalls Agile Esophageal Stent Systems Due to Catheter Detachment Risk

    Boston Scientific is recalling Agile Esophageal Stent Systems because the delivery catheter tip may detach. The recall affects 15 units distributed worldwide.

    Product
    AGILE ESO OTW PC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3230-2024·2024-10-02

    American Contract Systems Recalls Open Shoulder Medical Convenience Kits

    American Contract Systems is recalling 12 Open Shoulder medical convenience kits because a faulty chart recorder prevented verification of sterilization requirements.

    Product
    Open Shoulder, LWOS39L; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3294-2024·2024-10-02

    Boston Scientific Recalls Agile Esophageal Stent Systems Due to Catheter Tip Detachment

    Boston Scientific is recalling Agile Esophageal Stent Systems because the delivery catheter tip may detach. The recall affects 10 units distributed worldwide.

    Product
    AGILE ESO FC RMV 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3269-2024·2024-10-02

    Boston Scientific Recalls WallFlex Esophageal Stent Systems Due to Catheter Tip Detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems because the delivery catheter tip may detach. The recall covers 176 units distributed worldwide.

    Product
    WALLFLEX FC ESO STENT RMV 23X125MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esoph
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3281-2024·2024-10-02

    Boston Scientific Recalls WallFlex Esophageal Stents Due to Catheter Tip Detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems because the delivery catheter tip may detach during use.

    Product
    WALLFLEX PC ESOPH STENT 18/23MM X 103MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophag
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3202-2024·2024-10-02

    Philips MR 7700 MRI Scanner Recall for Loose Screws

    Philips is recalling 7 MR 7700 MRI scanners because loose screws may block the tabletop, potentially delaying diagnosis or causing patient anxiety.

    Product
    MR 7700 -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782120
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3301-2024·2024-10-02

    Cardinal Health Recalls Salem Sump Stomach Tubes Due to Valve Breakage

    Cardinal Health is recalling Salem Sump Dual Lumen Stomach Tubes because the Anti-Reflux Valve may break due to improper use. Labeling changes were made to address the issue.

    Product
    (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888264929, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888264945, Sterile; (c) CardinalHealth Salem Sump Dua
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3224-2024·2024-10-02

    Karl Storz Nephroscope Recall Due to Unapproved Reprocessing Instructions

    Karl Storz Endoscopy is recalling 5 nephroscopes because their instructions for use list reprocessing methods not reviewed or approved by the FDA.

    Product
    Karl Storz - Endoskope , REF: 27830KA, Nephroscope for MIP M, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3221-2024·2024-10-02

    Medical endoscope recalled for unapproved sterilization instructions

    Karl Storz Hopkins Telescope 6 endoscopes are recalled because their Instructions for Use contain reprocessing methods that lack FDA review and approval.

    Product
    Karl Storz - Endoskope , REF: 27293AA, Hopkins Telescope 6, Rx only, CE 0123
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3204-2024·2024-10-02

    MRI Diagnostic Device Loose Screws May Block Patient Tabletop

    Philips SmartPath dStream MRI devices have loose screws in the RF carrier assembly that may interfere with the moving tabletop, potentially delaying diagnosis. Affected units should be inspected and repaired immediately.

    Product
    SmartPath to dStream for XR and 3.0T- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782113; 2) 781270; 3) 782129
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-1822-2024·2024-10-02

    M&P Food Products Frozen Blintzes Recalled for Undeclared Wheat

    M&P Food Products Cheese & Cherry Blintzes are recalled because they contain undeclared wheat. Consumers with wheat allergies should not consume this product.

    Product
    M&P Food Products Cheese & Cherry Blintzes; Hand Made Russian Style; 6 Classic; NET WT. 14 OZ. (396g); KEEP FROZEN; INGREDIENTS: Flour, low fat milk; eggs, margarine; sugar, salt; FILLING INGREDIENTS: Farmer Cheese, cherries; Cheese & Cherry Blintzes Net Wt. 14 OZ. (396g) UPC: 70
    Category
    Food
    Distribution
    7 states
  • HighFDA (Food)·F-1825-2024·2024-10-02

    M&P Delicacies Pirozhki Frozen Pies Recalled for Undeclared Wheat and Milk

    M&P Delicacies frozen pirozhki pies contain undeclared wheat and milk allergens. Consumers with allergies to these ingredients should not consume the product.

    Product
    M&P Delicacies PIROZHKI (PIES) with POTATOES; INGREDIENTS: Flour, eggs, butter, vegetable oil, dry yeast, sugar, salt; Filling: Potatoe, onion, salt, pepper; 1 SERVING READY TO EAT; KEEP FROZEN
    Category
    Food
    Distribution
    7 states
  • HighFDA (Devices)·Z-3243-2024·2024-10-02

    Medical device kit recalled due to inability to confirm sterilization assurance

    American Contract Systems recalled its Cysto medical convenience kit (36 units, Lot 2405033) due to a humidity monitoring failure that prevented confirmation of adequate product sterilization.

    Product
    Cysto, FTCY03L; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3298-2024·2024-10-02

    Origio Sperm Wash Medium Recalled for Low Sodium Pyruvate Concentration

    CooperSurgical is recalling Origio Sperm Wash Medium due to low sodium pyruvate concentration. The deficiency could compromise gamete development in assisted reproductive procedures.

    Product
    Origio Sperm Wash Medium Pack of Five 60ml bottles - Intended for washing of spermatozoa. Part Number: 84055060D
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3286-2024·2024-10-02

    WallFlex and Agile esophageal stent systems recalled for delivery catheter tip detachment

    Boston Scientific is recalling WallFlex and Agile esophageal stent systems due to potential for delivery catheter tip detachment during insertion.

    Product
    WALLFLEX PC ESOPH STENT 23/28MM X 155MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophag
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3187-2024·2024-10-02

    Roche Creatine Kinase test reagents recalled for calibration errors

    Roche Diagnostics is recalling Creatine Kinase test reagents that may exhibit abnormal calibrations, producing inaccurate results. Lot 755471 distributed nationwide.

    Product
    Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination of creatine kinase (CK) in human serum and plasma. Catalog Number: 07190794190
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3240-2024·2024-10-02

    Surgical convenience kit sterilization assurance cannot be confirmed

    American Contract Systems is recalling the Essentia MAJ Laparotomy Std Pack because an equipment failure prevented confirmation of sterilization requirements for this medical convenience kit.

    Product
    Essentia MAJ Laparotomy Std Pack, ESLT53B; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3245-2024·2024-10-02

    FDA Recalls Surgical Instrument Handle That May Break During Use

    The FDA is recalling LINK Universal Handles due to risk of material failure when first used. The surgical instrument may break during surgery, compromising patient safety.

    Product
    LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3237-2024·2024-10-02

    Medical Convenience Kit Recalled for Failed Sterilization Assurance Validation

    American Contract Systems, Inc. is recalling the Hand Pack AKHD97C medical convenience kit because an inoperable chart recorder caused humidity readings to fall outside specifications, preventing confirmation of sterilization assurance.

    Product
    Hand Pack, AKHD97C; Medical convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide