  • HighFDA (Devices)·Z-2199-2025·2025-08-06

    SwabFlush prefilled syringes carry SwabCaps with a compromised seal

    SwabFlush Prefilled Syringes with ICU Medical SwabCaps are being recalled because a manufacturing issue affects the SwabCap seal, which may reduce disinfection of the luer access valve. 1,427,100 units affected.

    Product
    SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2233-2025·2025-08-06

    LVP blood administration sets have potential for external cassette leaks

    LVP Blood Products Administration Sets are being recalled because of the potential for external cassette leaks. 14,280 units are affected.

    Product
    LVP Blood Products Administration Set Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter (Qty 20). Product Code: SET-0014-20.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2224-2025·2025-08-06

    Strep Selective II Agar plates may carry Listeria monocytogenes contamination

    Strep Selective II Agar, product number R01859, is being recalled because the products may contain surface and subsurface contamination of Listeria monocytogenes. 371 units are affected.

    Product
    Strep Selective II Agar, Product Number R01859
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2235-2025·2025-08-06

    LVP dual Y-site administration sets have potential for external cassette leaks

    LVP Primary Administration Sets, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site, are being recalled because of the potential for external cassette leaks. 348,150 units are affected.

    Product
    LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Product Code: SET-0032-25.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2178-2025·2025-08-06

    HarmonyAIR E-Series monitor yokes may detach from the spring arm

    Dual Flat Panel Monitor accessories for HarmonyAIR E-Series Surgical Lighting Systems are being recalled because the yoke assembly may detach from the spring arm. 59 units are affected.

    Product
    Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2223-2025·2025-08-06

    Blood agar plates may carry Listeria monocytogenes surface contamination

    Blood Agar with 5% Sheep Blood, product number R01202, is being recalled because the products may contain surface and subsurface contamination of Listeria monocytogenes. 1,552 units affected.

    Product
    Blood Agar,5% Sheep Blood 100/PK, Product Number R01202
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2226-2025·2025-08-06

    Blood MacConkey biplates may carry Listeria monocytogenes surface contamination

    Blood/MacConkey Biplates, product number R02049, are being recalled because the products may contain surface and subsurface contamination of Listeria monocytogenes. 611 units are affected.

    Product
    Blood/MacConkey Biplate 100/PK, Product Number R02049
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2177-2025·2025-08-06

    HarmonyAIR dual flat panel monitor yokes may detach from spring arm

    Dual Flat Panel Monitor accessories for HarmonyAIR A-Series Surgical Lighting Systems are being recalled because the yoke assembly may detach from the spring arm. 167 units are affected.

    Product
    Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0544-2025·2025-08-06

    Bicillin L-A penicillin vials recalled after particulates found during visual inspection

    Bicillin L-A (penicillin G benzathine injectable suspension) 1,200,000 units per 2 mL is being recalled because of CGMP deviations, with particulates identified during visual inspection. 50,855 vials affected.

    Product
    BICILLIN L-A — BICILLIN L-A (PENICILLIN G BENZATHINE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2202-2025·2025-08-06

    Philips Evolution Upgrade 1.5T gradient coils may act as a heat source

    Evolution Upgrade 1.5T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 93 units are affected.

    Product
    Evolution Upgrade 1.5T Model Numbers (REF): (1) 782116, (2) 782148;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2213-2025·2025-08-06

    Philips MR 5300 gradient coils may act as a heat source

    MR 5300 systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 551 units are affected.

    Product
    MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2205-2025·2025-08-06

    Philips Ingenia 1.5T S gradient coils may act as heat source

    Ingenia 1.5T S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 136 units are affected.

    Product
    Ingenia 1.5T S Model Number (REF): 781347;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2222-2025·2025-08-06

    Osteotec silicone finger implants may contain foreign material

    Osteotec Silicone Finger Implants in sizes 3, 4 and 5 are being recalled because there is the potential that the silicone implant may contain foreign material. 15,573 units are affected.

    Product
    Brand Name: Osteotec Product Name: Osteotec Silicone Finger Implant Model/Catalog Number: OSTF-3 (Sz. 3), OSTF-4 (Sz. 4), OSTF-5 (Sz. 5) Software Version: N/A Product Description: The Osteotec Silicone Finger Implant is a one-piece, sterile, flexible silicone elastomer implan
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2230-2025·2025-08-06

    RAPIDPoint 500 wash waste cartridge lot biases seven blood analytes

    A lot of RAPIDPoint 500 Systems Wash/Waste Cartridges is being recalled because a manufacturing error causes positive and negative bias across seven analytes. 1,477 units are affected.

    Product
    RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2201-2025·2025-08-06

    Philips Achieva XR gradient coils may act as a heat source

    Achieva XR MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 40 units are affected.

    Product
    Achieva XR; Model Numbers (REF): (1) 781153, (2) 781253;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2200-2025·2025-08-06

    Philips Achieva 3.0T gradient coils may act as a heat source

    Achieva 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 460 units are affected.

    Product
    Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2203-2025·2025-08-06

    Philips Evolution Upgrade 3.0T gradient coils may act as heat source

    Evolution Upgrade 3.0T MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 50 units are affected.

    Product
    Evolution Upgrade 3.0T Model Numbers (REF): (1) 782117, (2) 782143;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2198-2025·2025-08-06

    CoSense sampling sets may carry residual carbon monoxide from insufficient aeration

    Precision Sampling Sets for the CoSense End-Tidal Carbon Monoxide Monitor are being recalled because residual carbon monoxide may remain in the cannula filter, which may raise ETCOc scores. 2,160 units affected.

    Product
    Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2210-2025·2025-08-06

    Philips Ingenia Elition S gradient coils may act as a heat source

    Ingenia Elition S MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 242 units are affected.

    Product
    Ingenia Elition S Model Numbers (REF): (1) 781357, (2) 782106, (3) 782137, (4) 782150 (China ONLY);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2209-2025·2025-08-06

    Philips Ingenia Ambition X gradient coils may act as heat source

    Ingenia Ambition X MR systems are being recalled because component failures in the gradient coil may act as a heat source with the potential to produce smoke or fire. 690 units are affected.

    Product
    Ingenia Ambition X Model Numbers (REF): (1) 781356, (2) 782109, (3) 782138;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2195-2025·2025-08-06

    LYFO-DISK M. canis single packs may fail to recover the organism

    LYFO-DISK M. canis preparations, REF 0894L, are being recalled because they may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.

    Product
    LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894L
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2231-2025·2025-08-06

    3M Ranger irrigation fluid warming set notice clarifies flow rates

    The 3M Ranger Irrigation Fluid Warming Set is subject to a notice issued to clarify flow rates related to the inlet fluid temperature. 186,880 units are affected.

    Product
    3M Ranger Irrigation Fluid Warming Set, REF 24750
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0547-2025·2025-08-06

    Sucralfate tablets recalled because bankruptcy left the quality program unmonitored

    Sucralfate Tablets USP 1 gram from Nostrum Laboratories are being recalled across all lots within expiry because the firm cannot monitor its quality program after filing for Chapter 11. 60,608 bottles affected.

    Product
    Sucralfate Tablets, USP 1 gram, packaged in a) 100-count bottles NDC 29033-0003-01, and b) 500-count bottles, NDC 29033-0003-05), Rx Only, Manufactured by: Nostrum Laboratories, Inc., Kansas City, MO 64120.
    Category
    Drug
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0542-2025·2025-08-06

    Amlodipine and benazepril capsules released with the wrong expiry date

    Amlodipine and Benazepril HCl Capsules USP 2.5 mg/10 mg are being recalled because they were released with the wrong expiry date, printed as February 2027 instead of January 2027. 7,668 bottles affected.

    Product
    AMLODIPINE BESYLATE AND BENAZEPRIL HYDROCHLORIDE — AMLODIPINE BESYLATE AND BENAZEPRIL HYDROCHLORIDE (AMLODIPINE BESYLATE AND BENAZEPRIL HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide