  • HighFDA (Devices)·Z-1511-2025·2025-04-09

    Cordis Vascular Stent System Recalled Due to Incorrect Product Size

    Cordis US Corp recalls Smart Control Vascular Stent System ILIAC devices due to product mix-up where some units are labeled 9mm X 60mm but contain different stent sizes. Patients should consult their healthcare provider immediately.

    Product
    Cordis, REF: C09060ML, Smart Control Vascular Stent System ILIAC, 9mm X 60mm 6F, SterileEO, Rx Only
    Category
    Medical Device
    Distribution
    5 states
  • HighFDA (Devices)·Z-1480-2025·2025-04-09

    Nasopore Nasal Dressing Recall Due to Compromised Sterile Packaging Seals

    Stryker Corporation is recalling Nasopore nasal dressings due to potentially compromised sterile packaging seals. Affected units totaling 182,344 may have lost sterility, posing a contamination risk.

    Product
    Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and Nasopore 4cm Firm 2PK Catalog Number 5400-222-004S
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1516-2025·2025-04-09

    Surgical Suture Recall Due to Silicone Curing Defect

    Ethicon Endo Surgery recalls ETHIBOND EXCEL surgical sutures worldwide due to curing issues with the silicone coating on the needles identified during manufacturing.

    Product
    ETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1517-2025·2025-04-09

    Clinical Chemistry Analyzer Recalled for Software Defect in Test Result Calculation

    Beckman Coulter is recalling certain DxC 500 AU Clinical Chemistry Analyzers due to a software defect that prevents calculated test results from being recalculated when constituent tests are rerun, potentially delaying patient result reporting.

    Product
    DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1474-2025·2025-04-09

    Nasopore nasal dressing recall: compromised sterile barrier seals

    Stryker recalls Nasopore 4cm fragmentable nasal dressings due to potential blister seal damage that compromises the sterile barrier. Nationwide distribution affected.

    Product
    Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·25V226000·2025-04-09

    Audi Q6 rearview camera image may be delayed or deactivated

    Volkswagen is recalling certain 2025 Audi Q6 and SQ6 vehicles because the rearview camera image may be delayed or deactivated after shifting into reverse, failing FMVSS 111.

    Product
    AUDI — AUDI SQ6 SPORTBACK E-TRON, Q6 E-TRON
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0308-2025·2025-04-09

    Prescription Duloxetine Capsules Recalled for Nitrosamine Impurity

    Breckenridge Pharmaceutical is recalling Duloxetine 30 mg capsules distributed nationwide due to a nitrosamine impurity above acceptable limits. Consumers should consult their doctor about their medication.

    Product
    DULOXETINE — DULOXETINE (DULOXETINE HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighNHTSA·25E025000·2025-04-09

    Hiab Recalls Waltco ZS-MK2 Lift Gates Due to Pin Seizure Risk

    Hiab USA Inc. is recalling specific Waltco ZS-MK2 series lift gates because insufficient lubrication may cause the cylinder pin to seize, leading to lift gate failure and injury risk.

    Product
    STRUCTURE:BODY:HATCHBACK/LIFTGATE recall
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1477-2025·2025-04-09

    Hemopore nasal wound dressing packaging seal defect may compromise sterility

    Stryker Corporation is recalling 182,344 units of Hemopore nasal/sinus wound dressing due to potential packaging seal defects that could compromise sterility. A bubble on the blister seal may indicate the sterile barrier has been breached.

    Product
    Hemopore 8cm nasal/sinus temporary wound dressing, Catalog Numbers 5400-020-208 and 5400-020-208ITL (OUS)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1471-2025·2025-04-09

    Nasopore Standard Nasal Dressing Recalled Due to Sterility Seal Defects

    Stryker is recalling 182,344 units of Nasopore Standard nasal dressings nationwide due to potential bubbles in blister seals that may breach the sterility barrier and increase contamination risk.

    Product
    Nasopore Standard 8cm fragmentable nasal dressing, Catalog Numbers 5400-010-008 and 5400-010-008ITL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0306-2025·2025-04-09

    Biocon Pharma atorvastatin calcium tablets recalled for failed dissolution specifications

    Biocon Pharma is recalling 2,184 bottles of Atorvastatin Calcium Tablets USP 40 mg nationwide. The product failed to meet required dissolution specifications, which may affect medication absorption.

    Product
    ATORVASTATIN CALCIUM — ATORVASTATIN CALCIUM (ATORVASTATIN CALCIUM)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0685-2025·2025-04-09

    SeaBear Clam Chowder Recalled for Potential Botulinum Contamination

    SeaBear Smokedhouse Alehouse Clam Chowder pouches are recalled nationwide due to potential seal failure that could allow Clostridium botulinum contamination. Consumers should not consume the product.

    Product
    SeaBear Smokedhouse Alehouse Clam Chowder is packaged in retort pouch, net wt. 12-oz., UPC 0 34507 07021 1. Case label, item 60104, Alehouse Chowder 12/CS.
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1470-2025·2025-04-09

    Nasopore nasal dressing may have compromised sterile packaging

    Stryker is recalling Nasopore Standard 4cm nasal dressings because the blister seals may develop bubbles, potentially compromising the sterility barrier. Affected units should not be used.

    Product
    Nasopore Standard 4cm fragmentable nasal dressing, Catalog Numbers 5400-010-004 and 5400-010-004ITL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1509-2025·2025-04-09

    BD Connecta Stopcocks Recalled for Non-U.S. Instructions Shipped to U.S. Market

    BD Connecta Luer-Lok 360 stopcocks were distributed with non-U.S. instructions for use. Healthcare facilities should verify they have the correct U.S. instructions before using affected lot numbers.

    Product
    BD Connecta BD Luer-Lok 360, REF 394910 UDI-DI code: 00382903949106 Connecta¿ Plus 1 and Connecta¿ Plus 3 2-way and 3-way Stopcocks are control valves for use in IV. therapy and hemodynamic pressure monitoring.
    Category
    Medical Device
    Distribution
    35 states
  • HighFDA (Devices)·Z-1506-2025·2025-04-09

    Medtronic A820 myPTM Infusion Pump Software Shows Delayed Patient Response

    Medtronic is recalling the A820 myPTM Software Application used with SynchroMed infusion pumps because the app is taking longer than expected to respond to patient interactions. Approximately 4365 units are affected worldwide.

    Product
    A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1478-2025·2025-04-09

    Nasopore Nasal Dressing Recall Due to Breached Sterile Packaging

    Stryker Corporation is recalling Nasopore nasal dressings due to defects in blister seals that may compromise sterility. Affected products may have bubbles on the seal, indicating the sterile barrier has been breached.

    Product
    Nasopore Ex Firm 4cm fragmentable nasal dressing, Catalog Number 5400-030-004 and Nasopore Forte plus 4cm Catalog Number 5400-030-004ITL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1515-2025·2025-04-09

    Ethicon PROLENE BLU Surgical Sutures Recalled for Manufacturing Defect

    Ethicon is recalling specific lots of PROLENE BLU surgical sutures due to silicone curing issues during needle manufacturing.

    Product
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1510-2025·2025-04-09

    Cordis Smart Control Vascular Stent Size Mismatch Recalled

    Cordis recalled 20 units of its Smart Control Vascular Stent System ILIAC due to a labeling mix-up where the external label does not match the actual stent size. Wrong-sized stents could cause serious vessel complications if implanted.

    Product
    Cordis, REF C10040ML, Smart Control Vascular Stent System ILIAC, 10mm X 40mm 6F, SterileEO, Rx Only
    Category
    Medical Device
    Distribution
    5 states
  • HighFDA (Devices)·Z-1472-2025·2025-04-09

    Otopore Cylinder Wound Dressing Recalled for Potential Sterility Seal Defect

    Stryker Corporation recalls Otopore Cylinder outer ear wound dressing due to potential bubbles in blister seals that may breach the sterile barrier. Affected units were distributed nationwide.

    Product
    Otopore Cylinder outer ear wound dressing, Firm Catalog Number 5400-020-000 and Forte Catalog Number 5400-020-000TL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1476-2025·2025-04-09

    Nasopore FD nasal dressing recalled for potential sterile barrier breach

    Stryker Corporation is recalling 182,344 units of Nasopore FD nasal dressing nationwide due to potentially breached sterile seals that may allow product contamination.

    Product
    Nasopore FD fragmentable nasal dressing, Catalog Numbers 5400-020-108 and 5400-020-108ITL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1479-2025·2025-04-09

    Nasal dressing recall: potential sterile barrier breach from seal defect

    Stryker is recalling over 182,000 Nasopore nasal dressings due to potential bubbles in the blister seal that could compromise the sterile barrier. The affected units were distributed nationwide.

    Product
    Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1482-2025·2025-04-09

    Stryker nasal wound dressing recalled for defective sterile packaging seal

    Stryker Corporation is recalling 182,344 units of Hemopore 2PK nasal/sinus wound dressing due to potential bubbles in blister seals that indicate sterility breach. Affected lot numbers are 2024020221, 2024051720, and 2024082321.

    Product
    Hemopore 2PK nasal/sinus temporary wound dressing, Catalog Number 5400-222-208S
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0693-2025·2025-04-09

    Good & Gather Baby Vegetable Puree Recalled for Elevated Lead Levels

    Good & Gather baby vegetable puree (4 oz) is being recalled due to elevated levels of lead. The recall affects 25,600 units distributed nationwide by Target.

    Product
    Good & Gather baby Pea, Zucchini, Kale & Thyme Vegetable Puree. Net Wt 4 oz. (113g). 6+ months. UPC 1 91907-99314 1. Distributed by Target Corporation, Minneapolis, MN 55403. Product of Colombia
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1500-2025·2025-04-09

    Phoroptor VRx Digital Refraction System head may become loose or detach

    Reichert, Inc. is recalling 122 units of the Phoroptor VRx Digital Refraction System Model 16241 due to an assembly defect that could cause the device head to become loose and potentially detach.

    Product
    Phoroptor VRx Digital Refraction System Model Numbers: 16241
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1481-2025·2025-04-09

    Nasopore Nasal Dressings Recalled Due to Breached Sterility Seals

    Stryker Corporation is recalling Nasopore nasal dressings due to potential breaches in packaging sterility seals that could compromise product sterilization. Approximately 182,344 units have been distributed nationwide.

    Product
    Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and Nasopore 4cm Firm 2PK Catalog Number 5400-222-004S
    Category
    Medical Device
    Distribution
    Distributed nationwide