  • HighFDA (Devices)·Z-0946-2025·2025-01-22

    Gynecological dilator kits recalled for potential sterile barrier compromise

    Sklar Instruments is recalling ECONO STERILE gynecological dilator kits nationwide due to packaging issues that may breach the sterile barrier, potentially affecting product safety.

    Product
    ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE HEGAR DILATOR 11&12MM STRLCS25Model Number 96-0960; 2) ECONO STERILE HEGAR DILATOR 9&10MM STRL CS25Model Number 96-0961; 3) ECONO STERILE HEGAR DILATOR 7&8MM STRL CS25Model Number 96-0962; 4) ECONO STERILE
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0954-2025·2025-01-22

    Surgical instrument kits recalled for potential sterile packaging defects

    Sklar Instruments is recalling 41,850 units of ECONO STERILE surgical procedure kits due to packaging defects that may compromise the sterile barrier, affecting multiple clamp and hemostat models used in general orthopedic surgery.

    Product
    ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE OLSEN-HEG NH SMO DEL 5.5STER25, Model Number 96-2352A; 2) ECONO STERILE HALSEY NH SERR 5" STRL CS/25, Model Number 96-2353; 3) ECONO STERILE OLSEN-HEGAR NH 5.5" STRL CS/25, Model Number 96-2355A; 4) ECO
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0862-2025·2025-01-22

    Surgical Retractor Grip Cables May Fray or Break During Use

    Intuitive Surgical is recalling 738 units of the 8MM Atrial Retractor (Model 470246) due to reports of frayed or broken grip cables on reusable instruments. The defect could compromise instrument performance during surgical procedures.

    Product
    8MM,ATRIAL RETRACTOR SHORT RIGHT,IS4000¿ REF 470246
    Category
    Medical Device
    Distribution
    52 states
  • HighFDA (Devices)·Z-0850-2025·2025-01-22

    Intuitive Surgical Recalls 8MM Surgical Scissors for Grip Cable Defects

    Intuitive Surgical is recalling 4,259 units of 8MM round tip scissors due to complaints of frayed or broken grip cables. The defect could affect instrument reliability during surgical procedures.

    Product
    8MM, ROUND TIP SCISSORS, IS4000 ¿ REF 470007
    Category
    Medical Device
    Distribution
    52 states
  • HighFDA (Devices)·Z-0865-2025·2025-01-22

    Reusable Surgical Instruments Recalled for Frayed or Broken Grip Cables

    Intuitive Surgical is recalling reusable surgical instruments due to reports of frayed or broken grip cables that may affect safe operation during procedures.

    Product
    8MM,MEGA SUTURECUT ND,IS4000¿ REF 470309
    Category
    Medical Device
    Distribution
    52 states
  • HighFDA (Devices)·Z-0854-2025·2025-01-22

    Prograsp Surgical Forceps Recalled Due to Frayed Grip Cables

    Intuitive Surgical is recalling 2,852 units of 8MM Prograsp Forceps due to complaints of frayed or broken grip cables. These reusable surgical instruments may lose grip control during use.

    Product
    8MM,PROGRASP FORCEPS,IS4000 ¿ REF 470093
    Category
    Medical Device
    Distribution
    52 states
  • HighFDA (Devices)·Z-0852-2025·2025-01-22

    Intuitive Surgical DeBakey Forceps Recalled for Frayed or Broken Grip Cables

    Intuitive Surgical is recalling 1,739 8MM DeBakey Forceps (Model 470036) used in surgical procedures due to increased reports of frayed or broken grip cables. Cable failure could impair instrument function during surgery.

    Product
    8MM,DEBAKEY FORCEPS,IS4000 ¿ REF 470036
    Category
    Medical Device
    Distribution
    52 states
  • HighFDA (Devices)·Z-0917-2025·2025-01-22

    Cardinal Health Presource Surgical Kits Recalled for Potential Endotoxin Contamination

    Cardinal Health is recalling Presource surgical kits due to potential endotoxin contamination in non-sterile surgical strips and patties. Endotoxins could pose a risk if contaminated components are used in sterile surgical procedures.

    Product
    Cardinal Health Presource Kits: 1) KIT, H N FREE FLAP, Catalog Number: PEHFHNAMHJ; 2) KIT,H N FREE FLAP , Catalog Number: PWHFHNAMH1; 3) KIT,H N FREE FLAP , Catalog Number: PWHFHNAMH2; 4) KIT,H N FREE FLAP , Catalog Number: PWHFHNAMH3; 5) KIT,H N FREE FLAP , Catalog Numb
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0878-2025·2025-01-22

    Fenestrated Bipolar Forceps Recalled Due to Frayed or Broken Grip Cables

    Intuitive Surgical is recalling 147,006 units of 8MM Fenestrated Bipolar Forceps due to increased complaints of frayed or broken grip cables, which could impair surgical control.

    Product
    8MM,FENESTRATED BIPOLAR FORCEPS,IS4000 REF 471205
    Category
    Medical Device
    Distribution
    52 states
  • HighFDA (Devices)·Z-0936-2025·2025-01-22

    ECONO STERILE Medical Procedure Kits Recalled for Sterile Barrier Compromise

    Sklar Instruments recalls ECONO STERILE medical procedure kits (300 units nationwide) due to packaging issues that may compromise sterile barrier integrity.

    Product
    ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE BACKHAUS TWL CLP STER 3.5 CS50, Model Number 96-2630; 2) ECONO STERILE BACKHAUS TWL CLP STER 5.25 25, Model Number 96-2636; general purpose clips
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0906-2025·2025-01-22

    BD Pyxis MedStation ES software issue may delay medication dispensing

    A software bug in BD Pyxis MedStation ES medication dispensing cabinets may prevent staff from locating alternate supply locations, potentially delaying patient therapy.

    Product
    BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0940-2025·2025-01-22

    ECONO STERILE Procedure Kits Recalled for Sterile Barrier Breach Risk

    Sklar Instruments is recalling ECONO STERILE medical procedure kits (Model 96-4921A) due to packaging issues that may result in a breach of the sterile barrier. The kits should not be used.

    Product
    ECONO STERILE medical procedure kits labeled as: ECONO STERILE BLAKE GALLSTONE FCP CVD 8"CS25, Model Number 96-4921A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0924-2025·2025-01-22

    FDA Recalls Small Graptor Retractor Instruments Due to Cable Failures

    Intuitive Surgical is recalling 42,028 Small Graptor surgical retractor instruments used with da Vinci X/Xi systems due to pitch cable failures. The recalled devices affect surgical facilities nationwide and internationally.

    Product
    8MM, Small Graptor REF 470318 Is a grasping retractor Instrument used with the da Vinci X/Xi Surgical Systems in a wide variety of procedures to perform a variety of surgical tasks including manipulation and retraction of tissue.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0901-2025·2025-01-22

    Digoxin Test Slides May Report Inaccurately Due to Hemoglobin Interference

    QUIDEL ORTHO is recalling VITROS DGXN test slides because hemoglobin can interfere with digoxin measurements at lower-than-expected levels, potentially causing inaccurate test results.

    Product
    Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxi
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0923-2025·2025-01-22

    Tenaculum Forceps for da Vinci Surgical Systems Recalled Due to Pitch Cable Failures

    Intuitive Surgical is recalling 8MM Tenaculum Forceps used with da Vinci X/Xi Surgical Systems due to pitch cable failures. These instruments are used for tissue manipulation and retraction during surgical procedures.

    Product
    8MM, Tenaculum Forceps REF 470207 Instrument used with the da Vinci X/Xi Surgical Systems in a wide variety of procedures to perform a variety of surgical tasks including manipulation and retraction of tissue.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0844-2025·2025-01-22

    Folysil Silicone Foley Catheters Recalled for Possible Sterility Issue

    Coloplast recalls Folysil Silicone Foley Catheters due to a possible sterility defect detected at the manufacturing facility. The affected units may not maintain proper sterilization, posing a risk of infection.

    Product
    Folysil Silicone Foley Catheters, Device Reference Numbers AA6106, AA6108, AA6110, AA6112, AA6114, AA6116, AA6118, AA6120, AA6122, AA6124, AA6308, AA6310, AA6312, AA6314, AA6316, AA6318, AA6320, AA6322, AA6324, AA6408, AA6410, AA6412, AA6414, AA6416, AA6418, AA6420, AA6422, AA642
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0201-2025·2025-01-22

    Clobazam tablets recalled for contamination with foreign tablets

    Amerisource Health Services LLC is recalling Clobazam 10 mg tablets nationwide due to contamination with foreign tablets or capsules. Patients should contact their pharmacist or healthcare provider.

    Product
    CLOBAZAM — CLOBAZAM (CLOBAZAM)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0930-2025·2025-01-22

    Medical procedure kit recall due to sterile barrier packaging breach risk

    Sklar Instruments is recalling ECONO STERILE nasal procedure kits due to packaging issues that may compromise sterility. The recall affects 175 units distributed nationwide.

    Product
    ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE VIENNA NSL SPEC MD STRL CS/25, Model Number 96-2236; 2) ECONO STERILE VIENNA NSL SPEC MD STRL CS/25, Model Number 96-2237M; 3) ECONO STERILE VIENNA NSL SPEC LG STRL CS/25, Model Number 96-2238; for nasal
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0939-2025·2025-01-22

    Sklar Instruments Recalls ECONO STERILE Medical Procedure Kits for Sterile Barrier Breach Risk

    Sklar Instruments is recalling 245,015 ECONO STERILE medical procedure kits nationwide due to packaging issues that may compromise the sterile barrier. No illnesses or injuries have been reported.

    Product
    ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ROCH-PEAN FCP 8" CVD CS/25,Model Number 94-0286; 2) ECONO STERILE DRESSING FCP 5.5" CS/25LG,Model Number 94-0290; 3) ECONO STERILE TISSUE FCP 1X2 5.5" STER CS25,Model Number 94-0291; 4) ECONO STERILE FRCP
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0860-2025·2025-01-22

    Intuitive Surgical Tenaculum Forceps Recalled for Frayed or Broken Grip Cables

    Intuitive Surgical is recalling 11,162 units of 8MM Tenaculum Forceps (Model 470207) due to increased complaints of frayed or broken grip cables on reusable surgical instruments. Users should discontinue use immediately.

    Product
    8MM,TENACULUM FORCEPS,IS4000¿ REF 470207
    Category
    Medical Device
    Distribution
    52 states
  • HighFDA (Devices)·Z-0935-2025·2025-01-22

    Sklar Instruments Recalls Sterile Vein Hooks Due to Packaging Issues

    Sklar Instruments is recalling 1,050 units of ECONO STERILE Vein Hooks due to packaging issues that may breach the sterile barrier. Products were distributed nationwide.

    Product
    ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE VEIN HOOK 1.5MM STERILE CS/25, Model Number 96-4020M; 2) ECONO STERILE VEIN HOOK 2MM STERILE CS/25, Model Number 96-4021M; 3) ECONO STERILE VEIN HOOK 2.5MM STERILE CS/25, Model Number 96-4022M; venous proced
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0937-2025·2025-01-22

    Sklar Instruments Recalls ECONO STERILE Surgical Scalpel Handles for Sterile Barrier Risk

    Sklar Instruments recalls ECONO STERILE SCALPEL HDL #3 surgical handles due to packaging issues that may compromise the sterile barrier. The 150-unit recall affects units distributed nationwide with lot code SK-125.

    Product
    ECONO STERILE medical procedure kits labeled as: ECONO STERILE SCALPEL HDL #3 W/METRIC CS/25, Model Number 96-3500M
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0913-2025·2025-01-22

    Cardinal Health Presource Surgical Kits Recalled for Potential Endotoxin Contamination

    Cardinal Health is recalling 512,786 Presource surgical kits due to potential endotoxin contamination in non-sterile surgical strips and patties. The kits were distributed to the US, Canada, and Saudi Arabia.

    Product
    Cardinal Health Presource Kits: 0 ) DESCRIPTION Catalog Number: CATALOG #; 1 ) KIT, NASAL , Catalog Number: PE23NSMEA; 2 ) KIT, NASAL , Catalog Number: PE23NSMEC; 3 ) KIT,SEPTOPLASTY NEW HAN , Catalog Number: PE23SENLA; 4 ) KIT,SEPTOPLASTY NEW HAN , Catalog Number:
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0950-2025·2025-01-22

    Strabismus Surgical Scissors Kits Recalled Due to Sterile Barrier Packaging Issues

    Sklar Instruments is recalling 350 strabismus surgical scissors kits nationwide due to packaging issues that may compromise the sterile barrier.

    Product
    Medical procedure kits labeled as: 1) ECONO STERILE STRABIS SCISS STR 4.5"STER 50, Model Number 96-2540; 2) ECONO STERILE STRABIS SCISS CVD 4.5"STER 50, Model Number 96-2541; 3) ECONO STERILE STRABIS SCISS STR 4.5"STER 25, Model Number 96-2542; 4) ECONO STERILE STRABISM
    Category
    Medical Device
    Distribution
    Distributed nationwide