Severity 4 (Severe) recalls have a real, documented risk of significant injury, hospitalization, or property damage even if no deaths have been reported. The agency classification is FDA Class II with hospitalization reports, USDA Class I, or a structural vehicle defect. These need attention quickly — not five-alarm urgency but not something to put off. Examples include allergen mislabeling on common allergens for products eaten by children, drug potency out of specification by a margin that could cause measurable harm, and vehicle defects affecting braking or steering.

The full rubric is on the methodology page. Severity is computed deterministically from the agency’s own classification plus the source text — no editorial vibes.

4876–4900 of 29102

  • SevereFDA (Devices)·Z-1720-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High Voltage Therapy

    Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during high voltage therapy due to a specific feedthrough manufacturing defect. Approximately 8,219 units were affected worldwide.

    Product
    ICD-DR DDME3D4 MIRRO MRI, Model Number DDME3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1710-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Certain Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may produce reduced or no energy output during high-voltage therapy, potentially compromising treatment effectiveness.

    Product
    ICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1735-2023·2023-06-28

    Implantable defibrillators risk energy output failure during high voltage therapy

    Certain Medtronic cardiac defibrillators may produce reduced or no energy output during high voltage therapy. The defect involves a specific feedthrough component used in the manufacturing process.

    Product
    CRT-D DTBA2D1 VIVA XT IS1/DF1 INTL, Model Number DTBA2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1749-2023·2023-06-28

    Implantable Cardiac Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Certain Medtronic implantable cardiac defibrillators with specific glassed feedthrough components may deliver reduced energy during high-voltage therapy. This could prevent the device from functioning when needed in cardiac emergencies.

    Product
    CRT-D DTBC2D1 BRAVA IS1/DF1 INTL, Model Number DTBC2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1714-2023·2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential Reduced Energy Output

    Medtronic implantable cardioverter defibrillators (ICDs) and CRT-Ds with specific feedthrough designs may produce reduced or insufficient energy during high voltage therapy, affecting the device's ability to treat heart arrhythmias.

    Product
    ICD-DR DDMB2D4 EVERA MRI XT IS-1/DF4 INT, Model Number DDMB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1808-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High Voltage Therapy

    Medtronic implantable cardioverter defibrillators may fail to deliver therapy due to a defect in the electrical feedthrough. The devices risk reduced or no energy output during high-voltage treatment.

    Product
    ICD-VR VISIA MRI AF S US/OUS DF4, Model Number DVFC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Food)·F-0991-2023·2023-06-28

    Kanimi Easy Roll Crab Flavored Seafood Product Recalled for Undeclared Wheat and Eggs

    Aquamar Inc is recalling Kanimi Easy Roll Crab Flavored Seafood due to undeclared wheat and egg allergens that pose a serious health risk to allergic consumers.

    Product
    Kanimi Easy Roll Crab Flavored Seafood Shred Style 91406 VCM Ingredients: Fish Protein (Pollock Whiting and or Threadfin Bream), Water, Potato Starch, Contains less than 2% of the following: Sugar, Modified Food Starch, Liquid Egg White, Salt, Sorbitol, Natural and Artificial Fla
    Category
    Food
    Distribution
    12 states
  • SevereFDA (Devices)·Z-1782-2023·2023-06-28

    Implantable Defibrillators May Fail During High-Voltage Cardiac Therapy

    Medtronic implantable defibrillators may fail to deliver energy during high-voltage cardiac therapy due to a defective glassed feedthrough component. The recall affects 2,968 units distributed nationwide and worldwide.

    Product
    CRTD COBALT HF MRI IS1 DF4, Model Number DTPB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1704-2023·2023-06-28

    Implantable defibrillators may fail to deliver high-voltage therapy due to defect

    Medtronic's EVERA XT implantable defibrillators may fail to deliver high-voltage therapy in certain models manufactured with a defective feedthrough. Patients should consult their cardiologist.

    Product
    ICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1796-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High-Voltage Therapy

    Medtronic recalls implantable cardioverter defibrillators due to rare potential for reduced or no-energy output during high-voltage therapy, which could prevent life-saving defibrillation.

    Product
    ICD-VR DVBB1D4 EVERA XT DF4 US, Model Number DVBB1D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1737-2023·2023-06-28

    Medtronic Implantable Heart Defibrillators Risk Therapy Delivery Failure

    Medtronic CRT-D and ICD implantable defibrillators may fail to deliver therapy due to a rare defect in a specific feedthrough component. The recalled devices carry a potential for reduced or no-energy output during high-voltage heart rhythm treatment.

    Product
    CRT-D DTBA2D4 VIVA XT IS1/DF4 INTL, Model Number DTBA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1791-2023·2023-06-28

    Implantable Defibrillators With Defective Feedthrough May Fail to Deliver Therapy

    Certain Medtronic implantable defibrillators may produce reduced or no energy output during high-voltage therapy due to a defective feedthrough component affecting 71 units.

    Product
    ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1717-2023·2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced or No Therapy Output

    Certain Medtronic implantable cardioverter defibrillators and cardiac resynchronization defibrillators may fail to deliver energy during high-voltage therapy due to a specific feedthrough design issue. The recall affects 3,801 units distributed in the US and worldwide.

    Product
    ICD-DR DDMD3D1 PRIMO MRI, Model Number DDMD3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1778-2023·2023-06-28

    Implantable Defibrillators May Deliver Reduced Energy During Therapy

    Medtronic ICDs and CRT-Ds with a specific feedthrough may deliver reduced or no energy during high voltage therapy. Approximately 2,669 units are affected in the US and worldwide.

    Product
    CRTD COBALT XT HF QUAD MRI IS4 DF1, Model Number DTPA2Q1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1794-2023·2023-06-28

    FDA Recalls Medtronic Implantable Defibrillators Due to Potential Output Risk

    Medtronic's ICD-VR VISIA AF S implantable defibrillators may fail to deliver energy during high-voltage therapy due to a feedthrough component defect. The FDA Class I recall affects 194 units distributed nationwide and worldwide.

    Product
    ICD-VR VISIA AF S US/OUS DF4, Model Number DVAC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1774-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High-Voltage Therapy

    Some Medtronic implantable defibrillators contain a specific feedthrough that may rarely cause reduced or no energy output during high-voltage therapy. The recall affects approximately 4,151 units distributed nationwide.

    Product
    CRTD COBALT XT HF MRI IS1 DF1, Model Number DTPA2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1773-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Critical Therapy

    Medtronic implantable defibrillators may produce reduced or no energy output during high-voltage therapy delivery. Affected devices were manufactured with a specific feedthrough component.

    Product
    CRTD DTMC2QQ COMPIA MRI QUAD OUS DF4, Model Number DTMC2QQ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1746-2023·2023-06-28

    Medtronic Implantable Cardiac Defibrillators Risk Therapy Delivery Failure

    Medtronic is recalling certain CRT-D and ICD models due to a potential feedthrough defect that may prevent energy delivery during high-voltage therapy. The FDA Class I recall affects 435 units distributed in the United States and worldwide.

    Product
    CRT-D DTBB2D4 VIVA S IS1/DF4 INTL, Model Number DTBB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1803-2023·2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential High Voltage Therapy Failure

    Medtronic implantable cardioverter defibrillators (ICDs) with a specific feedthrough component may have reduced or absent energy output during high voltage therapy. The FDA Class I recall affects 12,099 units distributed nationwide and worldwide.

    Product
    ICD-VR DVFB1D1 VISIA AF MRI US DF1, Model Number DVFB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1713-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may experience reduced or no energy output during high-voltage therapy, potentially preventing treatment of life-threatening heart rhythms.

    Product
    ICD-DR DDMB2D1 EVERA MRI XT OUS DF1, Model Number DDMB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1716-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Potential Energy Output Failure

    Medtronic is recalling certain implantable cardioverter defibrillators and cardiac resynchronization therapy devices due to a rare potential for reduced or no energy output during high voltage therapy. Approximately 37,547 units are affected.

    Product
    ICD-DR DDMC3D4 EVERA MRI S IS-1/DF4 GLOB, Model Number DDMC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1817-2023·2023-06-28

    Implantable Cardiac Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic's ICD-VR DVME3D4 defibrillators are recalled due to a rare risk of reduced or no energy output during high-voltage therapy. The defect involves a specific feedthrough component and affects approximately 14,989 units distributed nationwide and worldwide.

    Product
    ICD-VR DVME3D4 MIRRO MRI, Model Number DVME3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1727-2023·2023-06-28

    Implantable cardioverter defibrillators may output reduced or no energy during therapy

    Medtronic ICDs and CRT-Ds with specific feedthroughs may fail to deliver adequate high-voltage therapy. This FDA Class I recall affects 487 units distributed nationwide and worldwide.

    Product
    ICD CROME DR MRI IS1 DF1, Model Number DDPC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1740-2023·2023-06-28

    Implantable Defibrillators Risk Loss of Therapy from Feedthrough Defect

    Medtronic recalls certain CRT-D VIVA QUAD XT GOLD implantable defibrillators due to a rare risk of reduced or no electrical output during therapy. Affected devices contain a specific glassed feedthrough component.

    Product
    CRT-D VIVA QUAD XT GOLD CTD, Model Number DTBA2QQG; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1753-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver High-Voltage Therapy Energy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough defect. Approximately 11,794 units worldwide are affected.

    Product
    CRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide