The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

1626–1650 of 27375

  • HighFDA (Devices)·Z-0662-2026·2025-12-03

    IV set 490367 recalled over medication backflow and priming failure

    B. Braun Medical is recalling 6,768 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/60 DR/ML, CA
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0687-2026·2025-12-03

    IV set 490407 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 10,896 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/CARESITE EXT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0711-2026·2025-12-03

    SafeDAY set 352635 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 200,917 SafeDAY sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 60 DROP W/2 SAFEDAY; Cat
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0677-2026·2025-12-03

    IV set 490612 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 5,952 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. 15DR 2SAFELINE 1CARESITE 1S
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0723-2026·2025-12-03

    Imaging software recalled over a bug that can lose clinical information

    Intelerad Medical Systems is recalling 19 IntelePACS InteleConnect and TechPortal installations because a bug can cause data loss in the Relevant Clinical Info field.

    Product
    IntelePACS - InteleConnect / TechPortal
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0709-2026·2025-12-03

    Outlook set 354212 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 162,648 units outlook sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK IV SET 15DROP W/2 CARESITE; C
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0685-2026·2025-12-03

    IV set 490312 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 226,632 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMN SET W/15 DROP, ULTR
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0697-2026·2025-12-03

    IV set 480231 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 85,176 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0608-2026·2025-12-03

    Anesthesia IV sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 14,712 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV AD. SET,15 DROPS/ML, W/EXT S
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0690-2026·2025-12-03

    IV set 490303 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 32,304 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET,3 CARESITE LAD
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0661-2026·2025-12-03

    IV set 490309 recalled over backflow into the primary bag

    B. Braun Medical is recalling 259,392 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 2 CARESITE,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0614-2026·2025-12-03

    Anesthesia set 490468 recalled over backflow and inability to prime

    B. Braun Medical is recalling 58,752 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 117 IN.; Catalog N
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0630-2026·2025-12-03

    IV extension sets recalled over backflow between piggyback and primary containers

    B. Braun Medical is recalling 43,452 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. EXTENSION SET, 18 IN; Catalog Numb
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0657-2026·2025-12-03

    IV set 490427 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 224,856 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15 DROPS/ML
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0708-2026·2025-12-03

    Outlook pump sets recalled over backflow into the primary container

    B. Braun Medical is recalling 2,352 Outlook pump sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK PUMP SET 3 CARESITE LADS 133
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0647-2026·2025-12-03

    Single lot IV sets recalled over backflow and priming occlusion

    B. Braun Medical is recalling 48 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN 15D, 2 B/C VLV, 3
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0684-2026·2025-12-03

    Anesthesia set 490472 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 10,536 units anesthesia iv sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADM 15D 2 CARESITE, SAFE
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0658-2026·2025-12-03

    IV set 470005 recalled over medication backflow and priming failure

    B. Braun Medical is recalling 29,900 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET 15DR 3 CARESITE, 124
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0609-2026·2025-12-03

    Anesthesia set 490477 recalled over backflow and inability to prime

    B. Braun Medical is recalling 36,552 units anesthesia iv sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 134
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0629-2026·2025-12-03

    Extension set 456520 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 2,100 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. ANES SET W/DUAL FEM.LL SAFEPORT MAN
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0652-2026·2025-12-03

    IV set 490241 recalled over backflow and inability to prime

    B. Braun Medical is recalling 66,792 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET HF STOPCOCK,10
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0613-2026·2025-12-03

    Anesthesia set 490476 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 19,320 units anesthesia iv sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 2-HIGH FLOW STPCK
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0681-2026·2025-12-03

    IV set 490426 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 63,192 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/10 DROPS, 13
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0638-2026·2025-12-03

    Caresite extension sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 32,500 Caresite extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Caresite Extension Set_Export; Cata
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0689-2026·2025-12-03

    Secondary IV sets recalled over backflow into the primary container

    B. Braun Medical is recalling 25,728 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, QUAD MANIFOLD
    Category
    Medical Device
    Distribution
    0 states