How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

10726–10750 of 27470

  • HighFDA (Devices)·Z-0161-2022·2021-10-27

    Easi-Care Gait Belt Metal Buckle may fail at lower loads than rated

    Kinsman Enterprises is recalling approximately 13,123 Easi-Care Gait Belts because the metal buckle may fail at loads below the labeled 600-pound capacity, creating a fall risk for users.

    Product
    Easi-Care Gait Belt Metal Buckle (54", 60", and 72" lengths)
    Category
    Medical Device
    Distribution
    14 states
  • HighFDA (Devices)·Z-0139-2022·2021-10-27

    GE SIGNA Artist MRI System May Produce Incomplete Image Scans

    GE Healthcare recalls 18 SIGNA Artist MRI systems with a software defect that may cause missing image slices, creating gaps in 3D scans that could affect diagnostic accuracy.

    Product
    SIGNA Artist nuclear magnetic resonance imaging system
    Category
    Medical Device
    Distribution
    19 states
  • HighFDA (Devices)·Z-0124-2022·2021-10-27

    Atellica IM 1300 Analyzer Reagent Overflow at High Altitudes

    Siemens has recalled approximately 1,193 Atellica IM 1300 analyzers due to reagent overflow that occurs at altitudes above 350 meters, causing diagnostic test errors.

    Product
    Atellica IM 1300 Analyzer, SMN 11066001
    Category
    Medical Device
    Distribution
    23 states
  • HighFDA (Devices)·Z-0133-2022·2021-10-27

    MRI System Software Defect May Cause Missing Anatomical Image Data

    GE Healthcare's SIGNA Architect MRI system may fail to generate complete image slices under certain software conditions, resulting in gaps within 3D volume images that could affect diagnostic imaging integrity.

    Product
    SIGNA Architect nuclear magnetic resonance imaging system
    Category
    Medical Device
    Distribution
    19 states
  • HighFDA (Devices)·Z-0137-2022·2021-10-27

    GE SIGNA MRI Systems Software Issue Causes Missing Image Slices

    GE Healthcare's SIGNA Creator and SIGNA Explorer MRI systems may produce incomplete 3D images due to missing slices in software version DV29.1. Missing anatomical information could impact diagnostic accuracy.

    Product
    SIGNA Creator, SIGNA Explorer nuclear magnetic resonance imaging system
    Category
    Medical Device
    Distribution
    19 states
  • HighFDA (Devices)·Z-0141-2022·2021-10-27

    MRI System Recall: GE Healthcare SIGNA HDxt Missing Image Slices

    GE Healthcare is recalling the 1.5T SIGNA HDxt MR System due to a software issue that may cause missing slices in 3D volume images. Three devices distributed across the United States and internationally are affected.

    Product
    1.5T SIGNA HDxt MR System nuclear magnetic resonance imaging system
    Category
    Medical Device
    Distribution
    19 states
  • HighFDA (Devices)·Z-0153-2022·2021-10-27

    Jamshidi Bone Marrow Biopsy Tray Recalled for Misbranding

    Bard Peripheral Vascular Inc is recalling an Economy Jamshidi Bone Marrow Biopsy and Aspiration Tray due to misbranding. A total of 360 devices were distributed to healthcare facilities across multiple U.S. states.

    Product
    Catalog # BEK4511, Qty: 10, Ecomomy Jamshidi Bone Marrow Biopsy and Aspiration Tray, 11 G X 4"Biopsy/Aspiration needle, 15 G Aspiration Needle (Adjustable Length 15/16" -1-7/8") (24mm-48mm) sterile, For Single Use Only, RX only. Exp. Date Lidocaine Vendor lot. CareFusion. O
    Category
    Medical Device
    Distribution
    16 states
  • HighFDA (Devices)·Z-0152-2022·2021-10-27

    Smiths Medical Portex Loss of Resistance Device Lacks Sterilization and Reuse Labeling

    Smiths Medical ASD Inc. is recalling the Portex Loss of Resistance Device because product labeling lacks information on sterilization procedures and does not prohibit reuse. The recall affects 3,420 devices distributed internationally.

    Product
    smiths medical portex Loss of Resistance Device, REF 100/398/000
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0138-2022·2021-10-27

    GE Healthcare SIGNA Voyager MRI System Recall for Missing Image Slices

    GE Healthcare is recalling the SIGNA Voyager nuclear magnetic resonance imaging system due to a software issue that can cause missing slices in 3D volume images, potentially creating gaps in anatomical information.

    Product
    SIGNA Voyager nuclear magnetic resonance imaging system
    Category
    Medical Device
    Distribution
    19 states
  • HighFDA (Devices)·Z-0122-2022·2021-10-27

    MPR Slotted Mallet recalled for potential head separation

    Implant Resource Inc is recalling the MPR Slotted Mallet (Model MPR-1120-1000, Lot 200911-1-1) due to potential separation of the mallet head from the handle and unexpected deformations on the striking surface.

    Product
    MPR Slotted Mallet, labeling etched onto devices as follows: Medical Products Resource 800.524.5194 www.m-p-r.com MPR-1120-1000 (01)00840001405868
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0146-2022·2021-10-27

    Herniatome Surgical Instrument Recalled Due to Sterilization Process Failure

    ADRIA SRL recalls Herniatome cervical surgical instruments (Model 5091749) due to sterilization process parameter failures. Products distributed to NJ and NY.

    Product
    Herniatome, 17G x 9cm, Cervical Type, Model Number 5091749
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0157-2022·2021-10-27

    Medtronic Surgical Blades Recalled Due to Tip Breaks and Vibration

    Medtronic is recalling surgical blades used in ENT procedures due to increasing complaints of tip breaks and vibration. Approximately 274,991 devices distributed worldwide are affected.

    Product
    Medtronic sterile single-use blades that are components of the StraightShot M Series Handpiece and Fusion ENT Navigation Software, and the Midas Rex Microsaw and Triton System with the following product numbers: (1) REF 1883080EM, Rotatable Fusion Blade, Quadcut 3.0 mm x 13 cm
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0147-2022·2021-10-27

    Central Line Dressing Kit Component Has Shorter Expiration Than Kit Label

    Medical Action Industries recalls Central Line Dressing Change kits because the ChloraPrep 3ml component expires sooner than the kit label indicates. Users should discontinue use and contact their supplier.

    Product
    Kit: Central Line Dres Change 20/Cs
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0148-2022·2021-10-27

    CADD Administration Sets with Incorrect Air Filter Assembly

    Smiths Medical is recalling CADD Administration Sets (Model #21-7346-24) distributed in Canada because the air filter may have been assembled incorrectly. This manufacturing defect could impact device function.

    Product
    CADD Administration Sets, Model #21-7346-24
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0067-2022·2021-10-20

    FDA Recalls Ultrasound Gels and Lotions for Potential Bacterial Contamination

    Ultrasound gels and lotions from National Distribution & Contracting Inc are being recalled due to potential bacterial contamination. About 25,103 units distributed nationwide and in Canada are affected.

    Product
    ECO 85CB Lotion Ultrasound 8.5 oz 24/cs ECO 40JBC Gel Ultrasound Transmission Clr 8.5 oz BTL 24/cs ECO 40JMC Gel Ultrasound Transmission Clr 5L 4/cs NDC P301008 ProAdvantage Gel Ultrasound Transmission 8.5 oz 12/cs NDC P301105 ProAdvantage Gel Ultrasound Transmission 5 Liter
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0068-2022·2021-10-20

    Ultrasound gel products recalled for potential bacterial contamination

    Ultrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. Products distributed nationwide in the U.S., Canada, and other locations have been recalled.

    Product
    ECO 50TS Gel Tens Ultrasound 1in X 5L cube 1 Dispenser Btl
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0004-2022·2021-10-20

    Abbott SARS-CoV-2 Diagnostic Kit Recalled for Risk of False Positive Results

    Abbott Molecular is recalling 60,563 units of the Alinity m SARS-CoV-2 Amp Kit due to potential false positive test results. The affected kits were distributed worldwide from September 2021 through January 2023.

    Product
    Alinity m SARS-CoV-2 Amp Kit (Emergency Use Authorization, US Assay) and Alinity m SARS-CoV-2 Application Specification File. in vitro diagnostic
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0089-2022·2021-10-20

    FDA recalls Eco-Med ultrasound gels due to bacterial contamination risk

    FDA recalls ultrasound gels manufactured by Eco-Med Pharmaceutical due to possible bacterial contamination. Healthcare providers and facilities should stop use immediately.

    Product
    EcoGel 200 - RED Medical Supplies Clear Ultrasound Gel, Model No. 250UG-CMU
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0006-2022·2021-10-20

    Abbott Alinity m SARS-CoV-2 Amp Kit Recalled for Potential False Positive Results

    Abbott Molecular has recalled 71,314 units of the Alinity m SARS-CoV-2 Amp Kit due to potential false positive results. The kit was distributed worldwide.

    Product
    Alinity m SARS-CoV-2 Amp Kit (Non-US Assay) and Alinity m SARS-CoV-2 Application Specification File. in vitro diagnostic
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0008-2022·2021-10-20

    Abbott Respiratory Diagnostic Kit Recalled for Potential False Positive Results

    Abbott Molecular is recalling the Alinity m Resp-4-Plex AMP Kit due to potential false positive results. The recall affects 5,855 units distributed worldwide.

    Product
    Alinity m Resp-4-Plex AMP Kit (Non-US Assay) and Alinity m Resp-4-Plex Application Specification File. in vitro diagnostic
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0088-2022·2021-10-20

    Eco-Med Ultrasound Gel Recalled for Bacterial Contamination Risk

    Eco-Med Pharmaceutical's ultrasound gel products are recalled due to potential bacterial contamination. The FDA advises healthcare providers to stop using the affected product.

    Product
    EcoGel 200 - RED Medical Supplies Blue Ultrasound Gel, Model No. 250UG-BMU
    Category
    Medical Device
    Distribution
    Distributed nationwide