The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

7951–7975 of 27393

  • HighFDA (Devices)·Z-1644-2023·2023-06-07

    EarlyVue VS30 Vital Signs Monitor lacks calibration alarm, risking inaccurate CO2 readings

    Philips EarlyVue VS30 vital signs monitors lack a calibration alarm, potentially causing inaccurate CO2 measurements. This could prevent clinicians from detecting changes in patient condition.

    Product
    EARLYVUE VS30 VITAL SIGNS MONITOR Software Version A.00.02 and A.00.01 Product Number: 863359
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1641-2023·2023-06-07

    QUADROX-i Pediatric Oxygenator Packaging Defects Compromise Sterile Barrier

    FDA Class II recall of QUADROX-i Pediatric oxygenators due to packaging defects—creases, holes, cracks—that compromise sterile barriers, risking infection and sepsis during cardiopulmonary bypass procedures.

    Product
    QUADROX-i Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 30000-USA 701070384 HMO 31000-USA 701070388
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1668-2023·2023-06-07

    Medical Device Recall: X-Ray Detector Drive Shaft May Break

    The FDA is recalling 32 X-ray detector units due to a potential drive shaft failure that could cause uncontrolled detector movement. Affected units are distributed in 10 US states.

    Product
    Intelli-C, Right side suspended Tabletop, REF: 03400000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1670-2023·2023-06-07

    Smith & Nephew ENGAGE Cementless Knee System Recalled Over Revision Rate Concerns

    Smith & Nephew has recalled 1,994 units of its ENGAGE Cementless Partial Knee System because recent data shows higher-than-expected revision rates compared to similar devices.

    Product
    Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial); c) REF 1-10003-200 (SIZE 2-Right Medial); d) REF 1-10003-210 (SIZE 2-Left Medial); e) REF 1-
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1669-2023·2023-06-07

    X-Ray Machine Detector May Move Unexpectedly Due to Drive Shaft Failure

    FDA recalls the Intelli-C X-Ray machine's left side suspended tabletop model due to potential drive shaft failure in the detector lift mechanism, which could cause uncontrolled detector movement.

    Product
    Intelli-C, Left side suspended Tabletop, REF: 03400010
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1654-2023·2023-06-07

    Knee Replacement Insert Mislabeled; Contents May Not Match Catalog Number

    Howmedica Osteonics Corp. is recalling TRIATHLON X3 knee replacement bearing inserts because products labeled with one catalog number may actually contain a different specification. This mislabeling could result in incorrect implant sizing during surgery.

    Product
    TRIATHLON X3 TIBIAL BEARING INSERT PS #4 13MM Intended for knee replacement Catalog Number: 5532-G-413-E
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1634-2023·2023-06-07

    Cardiopulmonary bypass oxygenator recalled due to sterile barrier packaging defects

    Maquet Medical Systems USA is recalling 8,670 HMOD 70000-USA/QUADROX-iD Adult oxygenators used in cardiopulmonary bypass due to packaging defects that may compromise sterile barriers and expose devices to contamination.

    Product
    HMOD 70000-USA/QUADROX-iD Adult, Oxygenator used during cardiopulmonary bypass Item no. 701067840
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1636-2023·2023-06-07

    FDA Recalls QUADROX-i Small Oxygenator Due to Packaging Integrity Issues

    Maquet Medical Systems USA is recalling 1,784 QUADROX-i Small Oxygenators due to packaging defects that may compromise the sterile barrier, risking infection and other serious complications.

    Product
    BEQ-HMO 50000-USA/ QUADROX-i Small, Oxygenator used during cardiopulmonary bypass Item no. 701067905
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1663-2023·2023-06-07

    Teleflex HORIZON Titanium ligating clips recalled for missing cartridge safety feature

    Teleflex LLC is recalling approximately 146,606 units of HORIZON Microclip Titanium Ligating Clips because a safety feature is missing from the clip cartridge. Without this step feature, clips may be positioned variably and applied inadequately to blood vessels during surgery.

    Product
    Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1656-2023·2023-06-07

    QUADROX-iD Adult Oxygenator recalled for packaging defect compromising sterile barrier

    Maquet Medical Systems USA recalls 24,250 QUADROX-iD Adult Oxygenators due to small pinholes in product packaging that may compromise the sterile barrier, potentially exposing patients to non-sterile devices during cardiopulmonary bypass surgery.

    Product
    QUADROX-iD Adult, Oxygenator used during cardiopulmonary bypass Model Item No. HMOD 70000-USA 701067840 BEQ-HMOD70000-USA 701067859
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1658-2023·2023-06-07

    QUADROX-iD Pediatric Oxygenator Recalled Due to Packaging Sterile Barrier Defect

    Maquet Medical Systems USA recalls 5,022 QUADROX-iD Pediatric Oxygenators distributed nationwide due to small pinholes in product packaging that may compromise the sterile barrier, potentially exposing patients to non-sterile devices.

    Product
    QUADROX-iD Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. BEQ-HMOD30000-USA 701070397 BEQ-HMOD30000-USA 701050330
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1643-2023·2023-06-07

    Maquet QUADROX-iR Cardiopulmonary Bypass Device Recalled for Packaging Damage

    Maquet Medical Systems USA is recalling QUADROX-iR cardiopulmonary bypass devices due to damaged packaging that may compromise sterility. This could result in infection, sepsis, or ischemia.

    Product
    QUADROX-iR, used during cardiopulmonary bypass Model Item No. Description BEQ-HMO 51100-USA 701067936 Small Adult with filter, with BIOLINE Coating BEQ-HMO 71100-USA 701067880 Adult with filter, with BIOLINE Coating HMO 5
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1674-2023·2023-06-07

    LINK SymphoKnee System Femoral Augment Recalled for Incorrect Implant in Packaging

    Linkbio Corp. is recalling 11 units of the LINK SymphoKnee System L-Shaped Femoral Augment because the packaging contains the opposite side of the implant. This error could result in implantation of the wrong component if not detected.

    Product
    LINK SymphoKnee System L-Shaped Femoral Augment, Medial-Right/Lateral-Left, Sz 5-6 Tilastan, cemented. Intended for primary and revision total knee replacement. Item Number: 880-325/13
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1660-2023·2023-06-07

    QUADROX-i Neonatal Oxygenator packaging may compromise sterile barrier

    QUADROX-i Neonatal Oxygenators may have small pinholes in packaging that compromise the sterile barrier, potentially exposing the device to non-sterile conditions and risk of infection, sepsis, or ischemia.

    Product
    QUADROX -i Neonatal, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 10000-USA 701070412 HMO 11000-USA 701070416
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1638-2023·2023-06-07

    Cardiac Oxygenator QUADROX-i Adult Recalled for Sterile Barrier Packaging Defects

    Maquet is recalling QUADROX-i Adult cardiopulmonary bypass oxygenators (Models HMO 70000-USA and HMO 71000-USA) due to packaging defects including holes, cracks, and improper placement of accessories that may compromise sterility.

    Product
    QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Model Item no. HMO 70000-USA 701067820 HMO 71000-USA 701067823
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1635-2023·2023-06-07

    Oxygenator devices recalled due to packaging damage compromising sterile barrier

    Maquet Medical Systems USA recalls 15,880 QUADROX-iD oxygenators nationwide due to packaging damage that may compromise sterile barriers, potentially exposing patients to infection.

    Product
    BEQ-HMOD70000-USA/QUADROX-iD, Oxygenator used during cardiopulmonary bypass Item no. 701067859
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1672-2023·2023-06-07

    Partial Knee Implant Recall Due to Higher-Than-Expected Revision Rates

    Smith & Nephew is recalling 2,146 cementless partial knee implants (ENGAGE system) due to complaint data showing higher-than-expected revision rates compared to similar devices.

    Product
    Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-10012-200, SIZE 2-LEFT MEDIAL; d) REF 1-10012-250, SIZE 2-RIGHT MEDIAL; e) REF 1-10012-300, SIZ
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • CriticalFDA (Devices)·Z-1618-2023·2023-05-31

    Pediatric Ventricular Assist Device Blood Pumps Recalled for Potential Membrane Disruption

    BERLIN HEART GMBH recalls EXCOR Pediatric Ventricular Assist Device Blood Pumps due to potential disruption in triple-layer membranes. The 335 affected units were distributed to multiple US states and Canada.

    Product
    EXCOR Pediatric Ventricular Assist Device Blood Pumps, Models: P10P-001 (10 ml blood pump); P15P-001 (10 ml blood pump); P25P-001x01 (25 ml blood pump); P30P-001x01 (30 ml blood pump), P50P-001 (50 ml blood pump); P60P-001 (60 ml blood pump). Recall occurred in 2018, and was r
    Category
    Medical Device
    Distribution
    16 states
  • SevereFDA (Devices)·Z-1594-2023·2023-05-31

    COVID-19 At-Home Test Kits Recalled for Bacterial Buffer Contamination

    SD Biosensor is recalling 2,712,767 Pilot COVID-19 At-Home Test kits nationwide due to bacterial contamination in the liquid buffer component. Users may notice an odor from affected kits.

    Product
    Pilot COVID-19 At-Home Test, Reference Number 9901-NCOV-10G, Catalog Number 99COV200L-EN02, UPC Code: 887473000207 (EAN-13: 08800111707728)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1623-2023·2023-05-31

    Philips Incisive CT System Recalled for Potential Image Orientation Error

    Philips recalls the Incisive CT whole-body X-Ray imaging system in 22 units due to potential incorrect image orientation that could result in flipped or reversed images, leading to misdiagnosis, incorrect treatment, or unnecessary radiation exposure.

    Product
    Incisive CT- Whole-body computed tomography (CT) X-Ray System Model Number: 728143 (OUS)
    Category
    Medical Device
    Distribution
    4 states
  • HighFDA (Devices)·Z-1617-2023·2023-05-31

    Omnicell IVX Station medical device recalled for dose accuracy failures

    OmniCell recalls six Omnicell IVX Station devices used for automated pharmaceutical preparation. The devices may not account for scale inaccuracy, potentially allowing incorrect doses to pass quality checks.

    Product
    Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-1621-2023·2023-05-31

    Aligned Medical Solutions Hand Pack covers may detach during surgical use

    Windstone Medical Packaging is recalling Aligned Medical Solutions Hand Pack kits because light handle covers may detach during surgical procedures, potentially compromising the sterile field.

    Product
    Aligned Medical Solutions Hand Pack, REF: AmS4601E, containing P 2 Cover Light Handle Blue STE, Sterile EO
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1626-2023·2023-05-31

    Diagnostic Cartridge Assay May Fail to Detect Chlamydia Trachomatis Correctly

    Binx Health is recalling a single lot of its CT/NG diagnostic cartridges distributed in Pennsylvania. The cartridges may not accurately detect Chlamydia trachomatis, potentially producing false negative or false positive results.

    Product
    binx io CT/NG Assay- Single-use Cartridge for the qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae Ref: 1.002.101
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1614-2023·2023-05-31

    Cranial Hand Drill recalled due to rust on chuck components

    Integra LifeSciences Corp. is recalling 337 units of the Codman Cranial Hand Drill due to rust contamination on the drill chuck. The rust may cause inflammation, infection, and complications requiring revision surgery.

    Product
    Codman Cranial Hand Drill- Indicated when a craniotomy is required for placement of an intracranial pressure (ICP) monitoring device and/or cerebrospinal fluid drainage device. Part Number: 82-6607
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1613-2023·2023-05-31

    Olympus Single Use Distal Cover for Endoscope Recalled Due to Detachment Risk

    Olympus is recalling about 309,278 single-use endoscope covers that may detach during use, risking aspiration, inhalation, obstruction, or burns.

    Product
    Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code: Catalogue #N5786100
    Category
    Medical Device
    Distribution
    Distributed nationwide