The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

16176–16200 of 27471

  • HighFDA (Devices)·Z-1040-2019·2019-03-27

    Fujifilm FDR Go PLUS Mobile X-Ray System Recalled for Braking Defect

    Fujifilm is recalling 56 mobile X-ray systems because the braking mechanism may fail to engage when the handle is released.

    Product
    Fujifilm FDR Go PLUS, Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency. Exposure can be captured on film, CR or Digital Radiography…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1036-2019·2019-03-27

    ConvaTec Recalls Feeding Tubes Due to Potential Sterile Barrier Breach

    ConvaTec is recalling specific lots of Feeding Tube Metric devices because an internal assessment found a potential breach in the sterile barrier.

    Product
    Feeding Tube Metric, disposable, sterile, 100 eaches per sales unit. Labeled as: xray CH05/40CM and xray CH08/40CM Product Usage: Duodenal and Stomach Gastroenterology Tubes are sterile disposable devices which may be inserted through the nose or mouth via the esophagus into…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1019-2019·2019-03-27

    AAP Implantate Lambotte Chisel Recalled for Potential Breakage at Welded Parts

    AAP Implantate AG is recalling 25 units of the 20 mm Lambotte chisel due to a potential risk of breaking at the welded piece parts of the chisel body and knocking plate.

    Product
    aap Implante AG, Lambotte chisel, 20 mm, REF IU 3000-20 Product Usage: Surgical instruments are intended to aid in the insertion of bone plates and screws as well as for reduction and support during orthopedic surgeries. If not otherwise specified, instruments are intended…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1037-2019·2019-03-27

    ConvaTec Recalls Nelaton Urinary Catheters Due to Sterile Barrier Breach Risk

    ConvaTec is recalling Nelaton Urinary Catheters because an internal assessment found a potential breach in the sterile barrier, which could compromise the device's sterility.

    Product
    Nelaton Urinary Catheter, disposable, sterile, 100 eaches per sales unit. Labeled as: CH16/40CM and CH14/40CM Product Usage: Urinary Catheters are sterile disposable devices intended to provide an intermittent pathway for drainage of fluids from the bladder. They are…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1027-2019·2019-03-27

    CP Medical Recalls Sterile Surgical Sutures Due to Potential Sterility Compromise

    CP Medical Inc is recalling Plain and Chromic Gut Absorbable Surgical Sutures because routine endotoxin testing showed an out-of-limit result, potentially compromising sterility.

    Product
    Plain and Chromic Gut Absorbable Surgical Sutures, sterile. 12 sutures per box. Product Usage: Plain Gut Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neural…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1020-2019·2019-03-27

    FDA Recalls 850 Units of Covidien Force TriVerse Electrosurgical Devices

    The FDA is recalling 850 units of the Covidien Force TriVerse electrosurgical device due to potential sterile packaging compromise.

    Product
    Covidien Force TriVerse electrosurgical device 10 cord-Intended as electrosurgical energy to cut and/or coagulate tissue. Product Usage: Single-use device intended for use in open procedures (such as general, urologic, thoracic, plastic and reconstructive, gynecologic) and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0992-2019·2019-03-20

    Ortho-Clinical Diagnostics Recalls VITROS Chemistry Products Na+ Slides

    Ortho-Clinical Diagnostics is recalling VITROS Chemistry Products Na+ Slides because imprecision in the affected lot may produce biased patient sample and quality control fluid results.

    Product
    VITROS Chemistry Products Na+ Slides, Catalog Number 8379034 For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600…
    Category
    Medical Device
    Distribution
    20 states
  • HighFDA (Devices)·Z-1006-2019·2019-03-20

    FDA Recalls NexGen Trabecular Metal Tibial Augments Due to Labeling Mismatch

    Zimmer Trabecular Metal Technology is recalling 26 units of NexGen Trabecular Metal Tibial Half Block Augments because the product in the box may not match the label.

    Product
    NexGenTrabecular Metal Tibial Half Block Augment, Left Lateral/Right Medial, Size 4 5MM Thickness W/Screws Item Number: 00544800436 Product Usage: The Trabecular Metal Tibial Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0994-2019·2019-03-20

    Durex Pleasure Pack Condoms Recalled for Burst Pressure Failure

    Reckitt Benckiser is recalling Durex Pleasure Pack condoms because five batches failed burst pressure specifications.

    Product
    Durex Pleasure Pack (60 ct), synthetic polisoprene male condom, SKU 02340-98729-00 02340-98729-00 02340-98729-00 02340-98729-00 02340-98729-00 02340-98729-00
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0993-2019·2019-03-20

    Durex Real Feel Condoms Recalled for Burst Pressure Failure

    Reckitt Benckiser LLC is recalling Durex Real Feel condoms because five batches failed burst pressure specifications.

    Product
    Durex Real Feel, synthetic polisoprene male condom, SKU 02340-89456-14 02340-89455-00 02340-89456-14 02340-89456-14 02340-89455-00 02340-95767-00 02340-89456-14 02340-89457-00 02340-89456-14 02340-89456-14 02340-89455-00 02340-89455-00 02340-95767-00 02340-95767-00
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0996-2019·2019-03-20

    Steris Recalls AMSCO 5000 Series Washer/Disinfector Units Due to Software Defect

    Steris Corporation is recalling 68 AMSCO 5000 Series Washer/Disinfector units because a software defect may prevent the intended cleaning cycle from processing, potentially leading to inadequate disinfection.

    Product
    AMSCO 5000 Series Washer/Disinfector Model # 5052
    Category
    Medical Device
    Distribution
    27 states
  • HighFDA (Devices)·Z-1011-2019·2019-03-20

    US Endoscopy Padlock Clip Recalled for Potential Esophageal Laceration Risk

    US Endoscopy Group Inc is recalling Padlock Clip defect closure systems due to a risk of esophageal laceration during patient procedures.

    Product
    US Endoscopy Padlock Clip defect closure system Ref: C910001 The Padlock Clip is indicated for use in flexible endoscopy and for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the wall of the gastrointestinal organs.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0995-2019·2019-03-20

    Steris Recalls AMSCO 3000 Series Washer/Disinfector Units Due to Software Defect

    Steris Corporation is recalling 11 AMSCO 3000 Series Washer/Disinfector units because a software defect may prevent intended processing cycles, risking inadequate cleaning or damage to medical instruments.

    Product
    AMSCO 3000 Series Washer/Disinfector Model # 3052
    Category
    Medical Device
    Distribution
    27 states
  • HighFDA (Devices)·Z-0991-2019·2019-03-20

    Linet AVE 2 Birthing Bed Recalled for Power Cable Safety Issue

    Linet Spol. S.r.o. is recalling 253 AVE 2 Birthing Beds because the power cable may break or fail, posing a risk of sparks or exposure to live conductors.

    Product
    AVE 2 Birthing Bed
    Category
    Medical Device
    Distribution
    21 states
  • SevereFDA (Devices)·Z-0987-2019·2019-03-13

    Maquet XS Flat Screen Monitor Holder Recalled for Grounding Defect

    Getinge is recalling six Maquet XS Flat Screen Monitor Holders because they may not meet grounding requirements, posing a risk of electric shock.

    Product
    Maquet XS Flat Screen Monitor Holder, XS32 SPE, Configuration Part Number ARD22201040A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
    Category
    Medical Device
    Distribution
    2 states
  • SevereFDA (Devices)·Z-0989-2019·2019-03-13

    Maquet Flat Screen Monitor Holder Recalled for Grounding Non-Compliance

    Getinge US Sales LLC is recalling six Maquet XS Flat Screen Monitor Holders because they may not meet grounding requirements, posing a risk of electric shock.

    Product
    Maquet XS Flat Screen Monitor Holder, PWD70+DF HD R XS32 K3 SPE, Configuration Part Number ARD268400150A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
    Category
    Medical Device
    Distribution
    2 states
  • SevereFDA (Devices)·Z-0990-2019·2019-03-13

    Maquet XS Flat Screen Monitor Holder Recalled for Grounding Defect

    Getinge is recalling four Maquet XS Flat Screen Monitor Holders in Arizona and Texas because they may not meet grounding requirements, posing a risk of electric shock.

    Product
    Maquet XS Flat Screen Monitor Holder, SAT XS32 12 SPE, Configuration Part Number ARD2SAT00130A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
    Category
    Medical Device
    Distribution
    2 states
  • SevereFDA (Devices)·Z-0988-2019·2019-03-13

    Maquet XS Flat Screen Monitor Holder Recalled for Grounding Defect

    Getinge is recalling six Maquet XS Flat Screen Monitor Holders in Arizona and Texas because they may not meet grounding requirements, posing a risk of electric shock.

    Product
    Maquet XS Flat Screen Monitor Holder, PWD50+DF XS32 K3 SPE, Configuration Part Number ARD268400140A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0975-2019·2019-03-13

    Draegar Medical Systems Recalls Omega Patient Monitor Software

    Draegar Medical Systems is recalling Omega patient monitor software due to cybersecurity vulnerabilities that may cause device reboots or loss of functionality.

    Product
    Omega Systems These devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0985-2019·2019-03-13

    Siemens Uroskop Omnia Max Fluoroscopic X-ray System Recall for Swivel Arm Defect

    Siemens is recalling 191 Uroskop Omnia Max fluoroscopic X-ray systems because the swivel arm holding the monitors may become dislodged from the carrying arm.

    Product
    Uroskop Omnia Max -fluoroscopic X-ray system Model # 10762473 Uroskop Omnia Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0984-2019·2019-03-13

    Siemens Uroskop Omnia Fluoroscopic X-ray System Recalled for Monitor Arm Defect

    Siemens is recalling six Uroskop Omnia fluoroscopic X-ray systems because the swivel arm holding the monitors may become dislodged from the carrying arm.

    Product
    Uroskop Omnia -fluoroscopic X-ray system Model # 10094910 The Uroskop Omnia is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnosis, endourology and minimal invasive urology/surgery).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0973-2019·2019-03-13

    Stryker Reprocessed Arthroscopic Shavers Recalled for Display and Function Errors

    Stryker is recalling 14,265 reprocessed arthroscopic shavers that may display incorrect settings or error messages, preventing proper use during surgery.

    Product
    Stryker Sustainability Solutions (SSS) reprocessed Arthroscopic Shavers Arthroscopic shavers can be used to abrade, cut and excise tissue and bone; remove loose fragments; and, shave away debris in arthroscopic surgeries, as well as surgeries of the jaw and sinuses. Intended…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0971-2019·2019-03-13

    Abbott Recalls Architect Magnesium Assay Due to Depressed Urine Results

    Abbott is recalling Architect Magnesium assays because urine tests may show falsely low results. The recall covers 24,995 units distributed in the US and four other countries.

    Product
    ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31 The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0974-2019·2019-03-13

    Draeger Infinity Delta Patient Monitors Recalled for Cybersecurity Vulnerabilities

    Draeger Medical Systems is recalling Infinity Delta patient monitors due to cybersecurity vulnerabilities that may cause device reboots or loss of functionality.

    Product
    Infinity Delta Family patient monitors The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0970-2019·2019-03-13

    Pentax C2 CryoBalloon Controller Recalled for Overpressure Detection Failure

    Pentax of America Inc is recalling 45 C2 CryoBalloon Controllers because the device fails to detect balloon overpressure during specific nitrous oxide applications, posing a risk of patient injury.

    Product
    C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s…
    Category
    Medical Device
    Distribution
    Distributed nationwide