The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

1326–1350 of 39948

  • HighFDA (Devices)·Z-1786-2026·2026-04-15

    Cook Medical multipurpose drainage catheters carry overstated expiration dates

    Cook Medical is recalling a Multipurpose Drainage Catheter because affected lots were labeled with expiration dates that exceed the product's true shelf life. 20 units are affected.

    Product
    COOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1777-2026·2026-04-15

    Cook Medical CTO-18 wire guides labeled with overstated expiration dates

    Cook Medical is recalling Approach CTO-18 Micro Wire Guides because affected lots were labeled with expiration dates that exceed the product's true shelf life. 75 units are affected.

    Product
    COOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1779-2026·2026-04-15

    Blue Rhino G2-Multi tracheostomy introducer sets carry overstated expiration dates

    Cook Medical is recalling Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Sets because affected lots were labeled with expiration dates that exceed the true shelf life. 916 units are affected.

    Product
    COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA, Order Number G57682; Reference Part Number C-PTIS-100-UNL-HC-G-EU, Order Number G57687; Refer
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1789-2026·2026-04-15

    Cook Medical Spectrum central venous trays carry overstated expiration dates

    Cook Medical is recalling Spectrum Central Venous Trays because affected lots were labeled with expiration dates that exceed the product's true shelf life. 2,348 units are affected.

    Product
    COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, Order Number G44809; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, Order Number G
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1788-2026·2026-04-15

    Cook Medical central venous catheter sets labeled with overstated expiration dates

    Cook Medical is recalling Spectrum Central Venous Catheter Sets because affected lots were labeled with expiration dates that exceed the product's true shelf life. 5,080 units are affected.

    Product
    COOK MEDICAL Spectrum¿ Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G43983; Reference Part Number C-UDLM-401J-ABRM-HC-CAH, Order Number G24754; Reference Part Number C-UDLM-401J-LSC-ABRM-HC, Order Number G43986; Reference Part Number
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1778-2026·2026-04-15

    Cook Medical arterial pressure monitoring trays carry overstated expiration dates

    Cook Medical is recalling Arterial Pressure Monitoring Trays because affected lots were labeled with expiration dates that exceed the product's true shelf life. 20 units are affected.

    Product
    COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1787-2026·2026-04-15

    Cook Medical nitinol stone extractors labeled with overstated expiration dates

    Cook Medical is recalling NCompass Nitinol Stone Extractors because affected lots were labeled with expiration dates that exceed the product's true shelf life. 70 units are affected.

    Product
    COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1785-2026·2026-04-15

    Cricothyrotomy catheter sets labeled with overstated expiration dates

    Cook Incorporated is recalling Melker Universal Cuffed Emergency Cricothyrotomy Catheter Sets because affected lots were labeled with expiration dates that exceed the true shelf life.

    Product
    Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1791-2026·2026-04-15

    Cook Medical Wayne pneumothorax sets carry overstated expiration dates

    Cook Medical is recalling Wayne Pneumothorax Sets because affected lots were labeled with expiration dates that exceed the product's true shelf life. 104 units are affected.

    Product
    COOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1790-2026·2026-04-15

    Cook Medical chest tube tray labeled with an overstated expiration date

    Cook Medical is recalling Thal-Quick Chest Tube Trays because affected lots were labeled with expiration dates that exceed the product's true shelf life. Two units are affected.

    Product
    COOK MEDICAL Thal-Quick Chest Tube Tray: Reference Part Number C-TQTSY-3200, Order Number G07242
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-1775-2026·2026-04-15

    Cook Medical Micro Wire Guides Recalled Due to Incorrect Expiration Dates

    Cook Incorporated is recalling specific lots of COOK MEDICAL Approach CTO-6 Micro Wire Guides because they were labeled with expiration dates that exceed their actual shelf life.

    Product
    COOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-1653-2026·2026-04-15

    Siemens ARTIS imaging systems may display an inflated patient dose reading

    Siemens ARTIS pheno and ARTIS Icono systems may sporadically display a major increase in the x-ray dose applied, although the actual dose applied remains within system settings and regulatory limits.

    Product
    ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1647-2026·2026-04-08

    Trilogy Evo O2 obstruction alarm may take up to four breaths

    Philips Respironics is recalling the Trilogy Evo O2 ventilator because in some situations the Obstruction Alarm does not trigger within the timeframe required by the relevant standards.

    Product
    Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1649-2026·2026-04-08

    Trilogy Evo Universal obstruction alarm may take up to four breaths

    Philips Respironics is recalling the Trilogy Evo Universal ventilator because in some situations the Obstruction Alarm does not trigger within the timeframe required by the relevant standards.

    Product
    Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1646-2026·2026-04-08

    Trilogy Evo obstruction alarm may take up to four breaths

    Philips Respironics is recalling the Trilogy Evo ventilator because in some situations the Obstruction Alarm does not trigger within the timeframe required by the relevant standards.

    Product
    Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1591-2026·2026-04-08

    Impella controller purge blocked alarm may be delayed on older pumps

    Abiomed is recalling Automated Impella Controllers because the Purge System Blocked alarm display could be delayed when using first generation Impella 5.5 pumps.

    Product
    Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Produc
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1637-2026·2026-04-08

    Volara ventilator adapter recalled as Class I over desaturation and barotrauma

    Baxter is issuing an urgent correction for the Blue Ventilator Adapter Module used with the Volara system because home users may be unaware of possible oxygen desaturation or lung injury from overexpansion.

    Product
    Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory Therapy system.
    Category
    Medical Device
    Distribution
    1 state
  • SevereFDA (Devices)·Z-1648-2026·2026-04-08

    Trilogy EV300 obstruction alarm may take up to four breaths

    Philips Respironics is recalling the Trilogy EV300 ventilator because in some situations the Obstruction Alarm does not trigger within the timeframe required by the relevant standards.

    Product
    Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1589-2026·2026-04-08

    Impella purge cassettes recalled as Class I over purge leak risk

    Purge Cassettes sold within Impella Pump Sets and individually packaged are being recalled because of an increased risk of purge leaks with Generation 1 purge cassettes.

    Product
    Purge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5 Pack; Product Code: 0043-0003. 2. Impella RP US Pump Set; Product Code: 004334. 3. Purge C
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1659-2026·2026-04-08

    Philips Allura Xper FD10C X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD10C because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD10C; System Code: 722001;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1749-2026·2026-04-08

    Paradigm REAL-Time insulin pump delivery varies with pump height

    Medtronic is recalling Paradigm REAL-Time insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1751-2026·2026-04-08

    MiniMed 530G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 530G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 530G Insulin Pump (MMT-551, MMT-751)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1743-2026·2026-04-08

    MiniMed 640G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 640G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1692-2026·2026-04-08

    Vue Motion V12 may display cine image frames out of sequence

    Philips is recalling Vue Motion V12 because mis-ordered frames during dynamic cine runs may cause image frames to display out of sequence.

    Product
    Vue Motion V12. Product Number: 1017979.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1740-2026·2026-04-08

    MiniMed 630G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 630G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
    Category
    Medical Device
    Distribution
    Distributed nationwide