Diagnostic MRI system recalled for pressure relief valve defect
A Philips Ingenia MRI system has been recalled due to installation of a 42 psi pressure relief valve instead of the correct 3.75 psi valve, which may result in helium gas release and injury.
- Product
- Ingenia 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number…
- Category
- Medical Device
- Distribution
- Distributed nationwide