The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

10626–10650 of 27441

  • HighFDA (Devices)·Z-0307-2022·2021-12-01

    Ocean Water Seal Chest Drain Recalled for Inadequate Setup Instructions

    Atrium Medical is recalling 87,972 Ocean Water Seal Chest Drains due to inadequate setup instructions that may result in treatment delays and risk of patient harm.

    Product
    The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device. It consists of a collection chamber that holds shed mediastinal blood, a water seal chamber using a water seal that prevents air from entering into the pleural space and a suction control…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0281-2022·2021-12-01

    Continuous CT Fluoroscopy Ceiling Mount Bracket Assembly Wear Recall

    Monitor bracket assemblies in Philips Continuous CT Fluoroscopy systems may experience significant wear and detach, potentially injuring patients, operators, or bystanders during surgical procedures.

    Product
    Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 64 System Number: 882471 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0291-2022·2021-12-01

    Medtronic Endurant II Stent Graft Systems Recalled Due to Potential Spindle Detachment

    Medtronic is recalling 511 Endurant II Stent Graft Systems because during deployment, the spindle may detach from the distal end of the spindle hypotube.

    Product
    Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0271-2022·2021-12-01

    Abbott ARCHITECT Analyzer Software Issues May Cause Inaccurate Test Results

    Abbott is recalling approximately 20,994 ARCHITECT i1000SR, i2000SR, and i2000 automated analyzers due to 12 software-related issues in version 9.41 and earlier that may produce inaccurate diagnostic test results.

    Product
    ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and 8C89
    Category
    Medical Device
    Distribution
    50 states
  • HighFDA (Devices)·Z-0278-2022·2021-12-01

    Imaging device ceiling mount assembly may detach, posing injury risk

    A ceiling-mounted component on Philips CT imaging equipment may detach due to wear. This poses a risk of injury to patients, operators, or staff in operating rooms.

    Product
    Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini GXL 16 Slice System Number: 882410 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0306-2022·2021-12-01

    BD Nexiva IV Catheters Recalled Due to Non-Sterile Packaging Breach

    Becton Dickinson is recalling BD Nexiva IV catheter systems because of a packaging breach that renders them non-sterile. The recall affects 155,840 devices distributed nationwide.

    Product
    BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL/min (3660 mL/hr)-0.5mL, REF 383536; BD Q-Style Luer Access Split Septum BD Vialon Material; BD Instaflash Needle Technology; Rx Only, Sterile EO. BD Nexiva closed IV…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0270-2022·2021-12-01

    Western Oxytote VIPR System Battery Rupture Risk with Fume Emission

    Western/Scott Fetzer Company is recalling 40 units of the Western Oxytote DTE Digital VIPR System due to battery rupture risk that can break the plastic shroud and release fumes and smoke.

    Product
    Western Oxytote DTE Digital VIPR System (with E-size cylinder) Model: MNDS-603
    Category
    Medical Device
    Distribution
    4 states
  • HighFDA (Devices)·Z-0277-2022·2021-12-01

    CT imaging system ceiling mount bracket shaft wear may cause equipment to fall

    A CT fluoroscopy imaging system ceiling mount's bracket shaft may wear excessively and go undetected, potentially causing the bracket to detach and fall, risking injury to patients, operators, or service personnel.

    Product
    Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - CT Vision ACQSIM CT System Number: 728209 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0280-2022·2021-12-01

    Philips CT Fluoroscopy Ceiling Mount Bracket Wear and Detachment Hazard

    Philips Continuous CT Fluoroscopy ceiling mounts may experience monitor bracket assembly wear that goes undetected, potentially causing the bracket to detach and fall on patients, operators, or bystanders during surgery.

    Product
    Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 16 System Number: 882470 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0238-2022·2021-12-01

    Philips Azurion System Recall for 3D Image Reconstruction Defect

    Philips Healthcare is recalling Azurion systems with software releases 2.1(L1) and 2.1(L2) due to a software defect that prevents proper reconstruction of 3D X-ray images when the detector is positioned outside required tolerance.

    Product
    Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0272-2022·2021-12-01

    Clinical Chemistry Analyzer Software Defects May Produce Inaccurate Results

    Abbott ARCHITECT clinical chemistry analyzers with software 9.41 and earlier contain defects affecting calibration, sample handling, and quality control that may produce inaccurate diagnostic results.

    Product
    ARCHITECT c4000 REF 02P24-01/2P24, 1P86, 1R24, and 1R25; ARCHITECT c8000 REF 01G06-11/1G06; ARCHITECT c16000 REF 03L77-01/3L77;
    Category
    Medical Device
    Distribution
    50 states
  • HighFDA (Devices)·Z-0301-2022·2021-12-01

    Siemens Artis Q angiography systems may fail to deliver x-ray during procedures

    Siemens Artis Q angiography systems may fail to release x-ray when source-to-image distance lift and x-ray release are activated simultaneously, causing procedure delays.

    Product
    Artis Q ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848281
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0274-2022·2021-12-01

    Freelite Human Lambda Free Kit Recalled for Calibration Curve Inaccuracy

    The Binding Site Group recalled 2,254 Freelite Human Lambda Free Kits due to calibration curve issues that may produce higher-than-expected activity readings and affect measurement accuracy.

    Product
    Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S
    Category
    Medical Device
    Distribution
    13 states
  • HighFDA (Devices)·Z-0295-2022·2021-12-01

    Medtronic Wireless Recharger Model WR9200 Becomes Unresponsive on Improper Use

    Medtronic is recalling the Wireless Recharger Model WR9200 used with RestoreUltra and RestoreSensor neurostimulator implants. The recharger becomes unresponsive if charging instructions from the user guide are not followed correctly.

    Product
    Medtronic Wireless Recharger Model WR9200 included in the RS7200 Recharger Kit used by the RestoreUltra (Model 37712), RestoreSensor (Model 37714), Restore Ultra SureScan MRI (Model 97712), and RestoreSensor SureScan MRI (Model 97714) implantable neurostimulators.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0283-2022·2021-12-01

    CT Scanner Software Issue May Interrupt Diagnostic Workflows and Delay Diagnoses

    Siemens SOMATOM Force CT scanners with software syngo.CT VB20 may experience workflow interruptions, causing diagnosis delays and requiring additional patient scans and contrast media. No illnesses or injuries have been reported.

    Product
    SOMATOM Force with software syngo.CT VB20 Model #10742326
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0234-2022·2021-11-24

    Arrow-Trerotola Percutaneous Thrombolytic Device: Inner Lumen Detachment Risk

    Arrow International is recalling 1918 units of the Arrow-Trerotola Percutaneous Thrombolytic Device due to inner lumen detachment that can cause embolization, requiring emergency retrieval and delaying treatment.

    Product
    Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-WC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0224-2022·2021-11-24

    FDA Recalls Scrip Ultrasound Gel for Potential Bacterial Contamination

    Scrip Inc is recalling Scrip Ultrasound Gel Clear (5L) nationwide due to potential bacterial contamination. Affected units may cause harm if contaminated product contacts patients during ultrasound procedures.

    Product
    Scrip Ultrasound Gel Clear, 5L Intended for use in acoustic coupling of ultrasound transducer to tissue
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0232-2022·2021-11-24

    Arrow-Trerotola Percutaneous Thrombolytic Device Catheter Lumen Detachment Recall

    Arrow International is recalling Arrow-Trerotola Percutaneous Thrombolytic Device catheters because the inner lumen can detach from the device basket, causing embolization requiring additional intervention and delaying dialysis therapy. Approximately 1,245 units are affected worldwide.

    Product
    Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-HFWC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0231-2022·2021-11-24

    Arrow-Trerotola Percutaneous Thrombolytic Device Inner Lumen Detachment Recall

    The inner lumen of Arrow-Trerotola Over-The-Wire PTD catheters (7FR) can detach from the device, causing an embolism. This requires additional medical intervention and delays treatment.

    Product
    Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-12709-WC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0225-2022·2021-11-24

    Scrip Creme ultrasound lotion recalled for bacterial contamination

    Scrip Creme All Purpose Lotion is being recalled due to bacterial contamination in this ultrasound coupling lotion. The product—1001 gallons in lots distributed between August 2018 and August 2021—was distributed nationwide.

    Product
    Scrip Creme All Purpose Lotion Intended for use in acoustic coupling of ultrasound transducer to tissue
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0233-2022·2021-11-24

    Arrow-Trerotola Percutaneous Thrombolytic Device inner lumen detachment and embolization risk

    Arrow-Trerotola PTD catheters are being recalled because the inner lumen can detach from the device basket, causing embolization that requires additional intervention. 1,167 units are affected.

    Product
    Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-W The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0236-2022·2021-11-24

    Syphilis Test Calibrator Recalled for False-Positive Reactivity Issues

    Bio-Rad BioPlex 2200 Syphilis calibrator set (791 units, Lot 52953) is recalled due to elevated test reactivity that does not confirm with treponemal tests and potential early shelf-life failure causing measurement bias.

    Product
    BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended for the calibration of the BioPlex 2200 Syphilis Total & RPR Reagent Pack.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0226-2022·2021-11-24

    Sensis Diagnostic System UI Freeze Risk Without Regular System Reboots

    Siemens' Sensis diagnostic systems may experience partial UI freezing and fail to update vital signs display if not rebooted at least every 7 days per manufacturer instructions.

    Product
    Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641 (3)Sensis Vibe Combo Model 11007642 Intended Use: A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0258-2022·2021-11-24

    LEEP PRECISION Integrated System may fail to coagulate during surgery

    CooperSurgical is recalling 75 units of the LEEP PRECISION Integrated System due to a defect where the hand switch may not activate the coagulation function, potentially causing prolonged surgery and bleeding complications.

    Product
    LEEP PRECISION Integrated System, 220V CooperSurgical part numbers: LP-10-220
    Category
    Medical Device
    Distribution
    Distributed nationwide