The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

10076–10100 of 27441

  • HighFDA (Devices)·Z-0967-2022·2022-04-27

    Artis zee floor X-ray systems recalled for cooling system failure

    Siemens is recalling 305 Artis zee floor X-ray imaging systems because the cooling circuit may fail to maintain adequate coolant levels, potentially causing the X-ray tube to overheat. The system includes automatic safeguards that shut down X-ray operation if overheating is detected.

    Product
    Artis zee floor, Model Number 10094135
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0976-2022·2022-04-27

    Artis Q biplane X-ray systems cooling failure may prevent adequate tube cooling

    Siemens is recalling 269 Artis Q biplane X-ray systems due to potential coolant level drops that may prevent the X-ray tube from cooling adequately, triggering automatic system shutdown.

    Product
    Artis Q biplane, Model Number 10848282
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0960-2022·2022-04-27

    HUNTLEIGH Disposable Intraoperative Probe recalled for detachable faceplate

    Huntleigh Healthcare Ltd. is recalling 150 HUNTLEIGH Disposable Intraoperative Probes (8MHz) because the faceplate may become detached from the probe body during use.

    Product
    HUNTLEIGH Disposable intraoperative Probe, 8Mhz (box), REF DIPP10, containing 10 individual Disposable Intraoperative Probes, REF DIOP8
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0961-2022·2022-04-27

    Medical device software defect causes incorrect rotation count in surgical handpiece

    Medtronic Xomed integrated power console software version 2.7.3.0 has a defect that causes the set number of rotations to mismatch the actual rotations in specific modes when used with M5 handpieces. No injuries have been reported.

    Product
    Software 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND EHS, Product Number EC300
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0979-2022·2022-04-27

    Siemens Artis Q.zen ceiling system recall: inadequate X-ray tube cooling risk

    Siemens recalls 81 Artis Q.zen ceiling imaging systems distributed nationwide due to risk of inadequate X-ray tube cooling if coolant levels fall below a critical threshold, potentially triggering system shutdown.

    Product
    Artis Q.zen ceiling, Model Number 10848354
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0964-2022·2022-04-27

    Merlin PCS 3650 Programmer May Fail to Terminate Pacing Test Properly

    A software anomaly in the Merlin PCS 3650 programmer may cause an in-clinic pacing test to fail terminating and prevent restoration of the implanted device's normal settings. Approximately 30,069 affected devices have been distributed worldwide.

    Product
    Merlin PCS 3650 programmer Model 3330 software
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0970-2022·2022-04-27

    Siemens Artis zee biplane X-ray system cooling defect recall

    Siemens is recalling 373 Artis zee biplane X-ray diagnostic systems due to a cooling circuit defect that could prevent adequate tube cooling. When the coolant level drops, the system may overheat and shut down imaging capabilities.

    Product
    Artis zee biplane, Model Number 10094141
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0983-2022·2022-04-27

    GE Healthcare MRI Systems May Fall During De-Installation Without Proper Rigging

    GE Healthcare MRI systems may fall during magnet de-installation if rigging and hardware are not properly installed and secured, creating a potential injury hazard.

    Product
    GE Healthcare MR Systems, Magnetic Resonance imaging (MRI). SIGNA Premier, SIGNA Architect, SIGNA Pioneer, Discovery MR750w 3.0T, Discovery MR750 3.0T, SIGNA PET/MR, 1.5T SIGNA Creator, 1.5T SIGNA Explorer, 1.5T SIGNA MR355, 1.5T SIGNA MR360, 1.49T SIGNA MR380, SIGNA Voyager…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0918-2022·2022-04-20

    Cub 2 Enclosed Bed Canopy System Recalled Due to Potential Misuse

    Sensory Medical Inc recalled 225 units of the Cub 2 Enclosed Bed Canopy System for potential misuse risk. The firm is updating product warnings and precautions for devices distributed before December 16, 2021.

    Product
    Cub 2 Enclosed Bed Canopy System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0916-2022·2022-04-20

    ROTAWIRE guidewires recalled due to potentially compromised sterile pouch seals

    Boston Scientific is recalling ROTAWIRE Drive guidewires due to potentially open pouch seals that could compromise product sterility. Affected units were distributed in the US, territories, Japan, and Canada.

    Product
    ROTAWIRE Drive and wireClip Torquer, ROTAWIRE Drive Extra Support
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0922-2022·2022-04-20

    FDA Recalls Abbott Vascular Priority Pack Accessory Kits for Leak and Connection Issues

    Abbott Vascular is recalling 23 lots of Priority Pack accessory kits and packs due to reports of leaks and loose connections in the devices.

    Product
    20/30 Priority Pack Accessory Kit/.096 RHV - REF 1000186 WPL2122270-01 (2019-03-01) 20/30 Priority Pack w/.115 RHV - REF 1000186-115 WPL2122270-02 (2019-03-01)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0872-2022·2022-04-20

    Conductivity Sensors in Dialog Hemodialysis Systems May Crack and Leak

    B Braun Medical Inc is recalling conductivity sensors used in Dialog and Dialog+ hemodialysis systems due to potential cracks that could cause dialysate circuit leaks, affecting ultrafiltration during treatment.

    Product
    Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0940-2022·2022-04-20

    CLARAVUE ECG Electrode May Fail to Deliver Effective Defibrillation

    CLARAVUE Pre-wired ECG electrodes may not meet electrical safety standards for defibrillation, potentially delivering an ineffective electrical charge during emergency use. This is a Class II recall affecting devices nationwide.

    Product
    CLARAVUE Pre-wired Electrode - IP SET 5V RT ADULT Model: 32028781 50604-US
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0873-2022·2022-04-20

    Hemodialysis Conductivity Sensors recalled for potential cracks and dialysate leaks

    B Braun Medical Inc is recalling conductivity sensors used in Dialog hemodialysis systems due to potential cracking during use, which could cause leaks in the dialysate circuit and improper ultrafiltration. Affected patients should contact their healthcare provider.

    Product
    Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0931-2022·2022-04-20

    Siemens Atellica CH Lithium reagent recalled for potential reagent carryover

    Siemens Healthcare Diagnostics recalls approximately 2,158 units of Atellica CH Lithium_2 reagent (Lot Numbers 110108, 110185, 110238, 110302) due to potential for reagent carryover affecting quality control results, patient samples, and calibrator results.

    Product
    Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11097637
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0928-2022·2022-04-20

    Siemens Atellica CH Calcium_2 Reagent Carryover Recall

    Siemens is recalling Atellica CH Calcium_2 diagnostic reagent due to potential reagent carryover that could impact quality control, patient samples, and calibrator results. The recall affects approximately 2,395 units distributed globally and in the US.

    Product
    Atellica CH Calcium_2 (CA_2), Siemens Material Number 11097644
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0903-2022·2022-04-20

    Access-9 Hemostasis Valves Recalled for Guidewire Entrapment Risk

    Merit Medical Systems is recalling 36,445 Access-9 Large Bore Hemostasis Valves due to an internal gap that can trap the guidewire, potentially delaying medical procedures.

    Product
    Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-00137K, K05-00410G, K05-00438G, K05-00502F, K05-00587D, K05-01052, K05-01706C, K05-01880K, K05-02130, K05-02595, K09-12599A, K10-05487P, K12-00264, MAP111/F, MAP112/B, MAP112/F…
    Category
    Medical Device
    Distribution
    49 states
  • HighFDA (Devices)·Z-0934-2022·2022-04-20

    Surgical Head Fixation Device Mount Clamp Assembly Deflection Recall

    Deerfield Imaging recalls the Trumpf Mount Clamp Assembly component used in surgical head fixation devices. The clamp may deflect during use and become difficult to loosen after procedures, potentially requiring surgical intervention.

    Product
    Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used with ORT400 operating room tables.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0910-2022·2022-04-20

    Orthopedic Surgical Instrument Kit Recalled for Improper Driver Fit

    The Nextra Hammertoe Correction System surgical kit is being recalled due to an oversized driver that cannot properly connect to and remove from the implant. This defect can delay surgery.

    Product
    Nextra Hammertoe Correction System 4.5 Middle 3.2 Proximal Kit SKU: NX-4532K Orthopedic surgical instrument
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0915-2022·2022-04-20

    Boston Scientific ROTAWIRE Drive guide wire devices recalled for open pouch seals

    Boston Scientific is recalling ROTAWIRE Drive and ROTAWIRE Drive Floppy guide wire devices because pouch seals may be open, compromising sterility. The recall affects 433 5-packs distributed in the US, territories, Japan, and Canada.

    Product
    ROTAWIRE Drive and wireClip Torquer, ROTAWIRE Drive Floppy
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0935-2022·2022-04-20

    ECG Electrode Set Fails Electrical Safety, May Prevent Effective Defibrillation

    Graphic Controls' CLARAVUE pediatric electrode set fails electrical safety standards. The device may not deliver adequate charge during defibrillation, potentially rendering emergency treatment ineffective.

    Product
    CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0919-2022·2022-04-20

    MUSE Cardiology System edits and diagnosis data lost after test signed

    GE Healthcare's MUSE Cardiology Information System may lose test measurement edits and diagnosis statements after signing. The Class II recall affects 147 units in the US and internationally.

    Product
    MUSE Cardiology Information System, Model Numbers: a) 3101011-040, b) 3101010-401, c) 3101010-009, d) 3101010-008
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0943-2022·2022-04-20

    Ventilator trolley mounting brackets can detach and cause device to fall

    The mounting brackets on Breas Ventilator Trolleys can become detached, allowing ventilators to fall. This poses an injury risk to people near the equipment.

    Product
    Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0921-2022·2022-04-20

    Abbott Vascular 20/30 Indeflator Recalled for Leaks and Connection Failures

    Abbott Vascular recalls 20/30 Indeflator medical devices due to increased complaints of leaks and loose connections. The recall affects 17 product lots distributed nationwide and internationally.

    Product
    20/30 INDEFLATOR -REF 1000184 WPL2122268-01 (2019-03-01)
    Category
    Medical Device
    Distribution
    Distributed nationwide