Gastrointestinal videoscopes recalled due to unvalidated channel drying process
Aizu Olympus is recalling 3,627 gastrointestinal videoscopes because the channel air drying process was not validated. A small percentage of devices returned after repair had wet channels, creating potential for contamination and risk of infection.
- Product
- Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q16
- Category
- Medical Device
- Distribution
- Distributed nationwide