The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

6451–6475 of 39844

  • HighFDA (Devices)·Z-2466-2024·2024-08-14

    SIGNA PET/MR diagnostic imaging system poses implant overheating risk

    GE HealthCare recalled SIGNA PET/MR systems with specific software versions due to a configuration issue that can cause RF energy to exceed safety limits, potentially overheating MR-conditional implants.

    Product
    SIGNA PET/MR, with affected software versions: MP24.0, MP26.0, MP26.1, MR30.0, MR30.1; Tomographic imager combining emission computed tomography with nuclear magnetic resonance
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2509-2024·2024-08-14

    Philips IntelliVue Patient Monitors with disabled monitoring software options

    Philips IntelliVue Patient Monitors (MX400/450/500/550) with software versions L.x or M.x lack required monitoring capabilities due to removed software entitlements. The missing features could lead to incorrect or delayed patient treatment.

    Product
    IntelliVue Patient Monitor MX450 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866062. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, an
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2515-2024·2024-08-14

    Centurion Procedure Kits with Defective Syringes Recalled Worldwide

    Medline Industries is recalling Centurion procedure kits due to defective plastic syringes affected by an FDA safety alert. The syringes may leak or break and pose a potential risk to patient health.

    Product
    Centurion procedure kits labeled as NEURO ANGIO PACK, Pack Number DYNDA1431A
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2511-2024·2024-08-14

    Philips IntelliVue Patient Monitors Shipped with Missing Monitoring Capabilities

    Philips IntelliVue patient monitors shipped or updated with software that removes monitoring capabilities, potentially delaying or resulting in incorrect treatment.

    Product
    IntelliVue Patient Monitor MX550 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866066. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2488-2024·2024-08-14

    BD PYXIS Automated Medication Cabinet Software Label Error Affects 11 Units

    CareFusion's BD PYXIS medication dispensing cabinet with software version 3.9.1.9 may print incorrect bin labels during restocking, potentially leading to wrong medication dispensing. Eleven units distributed nationwide are affected.

    Product
    BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P, REF: 169-82 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2523-2024·2024-08-14

    Biopsy Drape Pack sterilization failure recall from American Contract Systems

    American Contract Systems is recalling 224 Biopsy Drape Pack convenience kits due to insufficient aeration time during sterilization, which may have compromised the sterility of affected units. Lot 981241 distributed in CA, KS, ME, MN, MO, NE, and SD is affected.

    Product
    American Contract Systems Biopsy Drape Pack convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2471-2024·2024-08-14

    GE SIGNA 1.5T MR Systems May Overheat Implants in Low SAR Mode

    GE Medical Systems recalled 2,146 SIGNA 1.5T MR systems that may overheat implants during certain imaging sequences in Low SAR mode due to a software defect in radiofrequency safety calculations.

    Product
    1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2533-2024·2024-08-14

    Verigene Gram Positive Blood Culture Test Recalled for Defective Cartridges

    Luminex recalls approximately 1,100 diagnostic kits due to defective test cartridges that may produce false positive results for Staphylococcus lugdunensis and Enterococcus faecium bacteria.

    Product
    Verigene Gram Positive Blood Culture (BC-GP) Nucleic Acid Test
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2498-2024·2024-08-14

    Automated Dispensing Cabinet Software Error May Print Incorrect Medication Bin Labels

    CareFusion's BD PYXIS MEDBANK automated dispensing cabinet contains a software flaw that may print incorrect medication bin labels, potentially leading to dispensing of wrong medications.

    Product
    BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P, REF: 169-93 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2507-2024·2024-08-14

    Integra Recalls TruDi NAV Suction Instruments for Calibration Error

    Integra LifeSciences has recalled 265 TruDi NAV Suction Instruments due to incorrect calibration. The miscalibrated units may display inaccurate position information during surgical navigation procedures.

    Product
    TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery
    Category
    Medical Device
    Distribution
    35 states
  • HighFDA (Devices)·Z-2520-2024·2024-08-14

    Medical Procedure Kit Recalled Over Potential Sterilization Compromise

    American Contract Systems is recalling Special Procedure Pack convenience kits due to insufficient aeration time during sterilization, which may have compromised sterility. Distributed in California, Kansas, Maine, Minnesota, Missouri, Nebraska, and South Dakota.

    Product
    American Contract Systems Special Procedure Pack convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2496-2024·2024-08-14

    Automated Medication Dispensing Cabinet Software Issue with Incorrect Bin Labeling

    BD PYXIS MEDBANK automated medication dispensing cabinets with software version 3.9.1.9 may generate incorrect bin location labels during restocking. This could cause medications to be placed in wrong storage locations.

    Product
    BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P, REF: 169-91 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2501-2024·2024-08-14

    Medication Dispensing Cabinet Software Error Can Print Incorrect Bin Labels

    A software error in CareFusion's BD PYXIS automated medication dispensing cabinet can cause incorrect medication bin numbers to be printed on restock labels, risking medication dispensing errors.

    Product
    BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P, REF: 169-97 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2483-2024·2024-08-14

    MRI System Software Error Can Overheat Patient Implants

    GE Medical Systems recalled three SIGNA 7.0T MRI systems due to a software error that can exceed preset safety limits during imaging, potentially overheating implants in patients under certain conditions.

    Product
    SIGNA 7.0T, with affected software versions: 7T29.1, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2532-2024·2024-08-14

    FDA Recalls CryoPatch Cardiac Tissue Grafts Due to Invalid Test Results

    Artivion has recalled CryoPatch SG cardiac tissue patches used in pulmonary surgeries due to approval and release with invalid test results. Patients who received implants should consult their healthcare provider.

    Product
    (1) CryoPatch SG Pulmonary Human Cardiac Hemi-Artery - Artery Patch, SGPH00 Cryopreserved Decellularized Cardiac Tissue Allograft Prepared with SynerGraft (SG) Technology; and (2) CryoPatch SG Pulmonary Human Cardiac Branch Patch, SGP020, Cryopreserved Decellularized Cardiac Tis
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2477-2024·2024-08-14

    GE Brivo MR355 MRI System Software Issue Causing Implant Overheating Risk

    GE HealthCare has recalled certain Brivo MR355 MRI systems due to a software issue that can cause overheating of implanted medical devices in specific imaging scenarios. Approximately 375 units have been affected worldwide.

    Product
    Brivo MR355, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2467-2024·2024-08-14

    Discovery MR450 MRI System Can Overheat Implants Under Specific Conditions

    GE Discovery MR450 MRI systems with certain software versions can cause implant overheating under specific scanning conditions. 23 units worldwide require firmware updates.

    Product
    Discovery MR450 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2538-2024·2024-08-14

    Steris Endoscope Reprocessor Hookups: Incorrect Instructions Risk Improper Disinfection

    Steris recalls Hookup Model 2-8-540CAS endoscope reprocessor units due to incorrect instructions that could prevent proper disinfection. Using incorrect hookup connections may lead to patient contamination or infection risk.

    Product
    Hookup Model/Catalog Number: 2-8-540CAS Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispens
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2474-2024·2024-08-14

    SIGNA MR380 MRI system software may exceed implant heating limits

    GE SIGNA MR380 MRI systems with software versions SV25.5 or SV25.6 may exceed radiofrequency heating limits during certain imaging sequences, potentially causing MR-conditional implants to overheat. The issue occurs in Low SAR Mode when specific T2 FLAIR sequences are used with the Optimized T2 FLAIR option disabled.

    Product
    SIGNA MR380, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2461-2024·2024-08-14

    MRI System Software Miscalculation May Overheat Patient Implants

    A software error in GE SIGNA Hero MRI systems can cause overheating of MR conditional implants under specific conditions. The recall affects 95 units distributed worldwide.

    Product
    SIGNA Hero, with affected software versions: PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2518-2024·2024-08-14

    Gyn Laparoscopy Kits Recalled Due to Inadequate Sterilization Process

    American Contract Systems is recalling 29 gynecological laparoscopy surgical kits due to insufficient aeration during sterilization, which may compromise sterility and increase infection risk for patients.

    Product
    American Contract Systems Gyn Laparoscopy convenience kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2476-2024·2024-08-14

    GE SIGNA MR360 MRI System Software Defect May Overheat Implants

    GE SIGNA MR360 MRI systems with software versions SV25.5 or SV25.6 may exceed prescribed RF exposure limits in certain scan configurations, potentially overheating MR-conditional implants. The recall affects 81 units worldwide.

    Product
    SIGNA MR360, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2542-2024·2024-08-14

    Olympus Everest Bipolar Cutting Forceps Recalled for Packaging Defects

    Olympus is recalling Everest Bipolar Cutting Forceps due to fractures and breakages in packaging trays and Tyvek covers that may compromise sterility. The recall affects 4,364 units distributed worldwide.

    Product
    The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer diameter size of the forceps. Coagulation is achieved using electrosurgical energy under visualization. The device is to be used with bipolar outputs of c
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2484-2024·2024-08-14

    SIGNA MAGNUS MRI Software Can Overheat Implanted Devices

    GE Medical Systems recalled SIGNA MAGNUS MRI systems with software versions MR29.1 or RX29.1. Under specific imaging conditions in Low SAR Mode, the system may exceed safe radiofrequency power limits and cause implanted medical devices to overheat.

    Product
    SIGNA MAGNUS, with affected software versions: MR29.1, RX29.1; Nuclear Magnetic Resonance Imaging System - research use only
    Category
    Medical Device
    Distribution
    0 states