The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-2485-2017·Announced 2017-06-21

Biomerieux Recalls VITEK 2 Gram Negative Susceptibility Cards Due to Moisture Risk

Biomerieux Inc is recalling VITEK 2 Gram Negative Susceptibility cards because compromised container integrity may expose the product to moisture, potentially causing antibiotic degradation.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves potential loss of potency due to moisture exposure without reports of injury or illness, fitting the criteria for a moderate severity voluntary precautionary recall.

Plain-English summary

Biomerieux Inc is recalling 2,541 cartons of VITEK 2 Gram Negative Susceptibility cards (AST-N325), REF 418513. The recall is due to a potential compromise in the integrity of the product container, which may expose the product to moisture. This exposure could lead to antibiotic degradation and a loss of potency.

The affected product was distributed nationwide in the U.S., including to military and government sites. It was also distributed to numerous international locations, including Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Chile, China, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Gambia, Georgia, Germany, Greece, Grenada, Guatemala, Guinea, Haiti, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Uganda, Kuwait, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Malta, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Zimbabwe.

The source text does not specify specific actions for consumers or healthcare providers beyond the recall notice. The agency classification for this recall is Class II.

The recalled product

Product
VITEK¿ 2 Gram Negative Susceptibility card (AST-N325), IVD, REF 418513, 20 cards per carton.
Manufacturer
Biomerieux Inc
Affected units
2,541

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • 7650146203 21/Apr/18 7650240103 24/Jul/18

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Part of a larger recall action

This product is one of 187 recalled by Biomerieux Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 187 products in this recall →

Similar recalls