Integra LifeSciences Recalls C7000 CUSA Clarity Ultrasonic Tissue Ablation System
Integra LifeSciences is recalling the C7000 CUSA Clarity Ultrasonic Tissue Ablation System because the 36 kHz Sterilization Tray does not meet the required dry time criteria.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a procedural discrepancy in drying time rather than a direct, reported injury or high-risk pathogen, fitting the criteria for a moderate severity voluntary precautionary recall.
Plain-English summary
Integra LifeSciences Corp. is recalling the C7000 - CUSA Clarity Ultrasonic Tissue Ablation System. The recall involves the 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system.
The recall was initiated because the 36 kHz Sterilization Tray system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components, but 40 minutes is required for complete drying.
The affected products were distributed in the United States and internationally. Users should contact the manufacturer for further instructions regarding the recall.
The recalled product
- Product
- C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
- Manufacturer
- Integra LifeSciences Corp.
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Finnish OPS Manual P/N 60905782
Distribution
Part of a larger recall action
This product is one of 42 recalled by Integra LifeSciences Corp. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 42 products in this recall →Related recalls
Same manufacturer · Integra LifeSciences Corp.
See all →- HighAURORA Surgiscope System Obturator May Break or Separate
FDA (Devices) · 2025-03-12
- HighMediHoney Gel Wound Dressing Recall: Potential Induction Seal Defect
FDA (Devices) · 2025-02-19
- ModerateIntegra implantable hydrocephalus valve recalled for incorrect product description on labels
FDA (Devices) · 2025-01-29
- ModerateCODMAN CERTAS Plus programmable valve recalled for incorrect label descriptions
FDA (Devices) · 2025-01-29
- ModerateProgrammable Valve Device Recalled Due to Incorrect Product Description Labels
FDA (Devices) · 2025-01-29
Same hazard · inadequate sterilization
See all →- HighBD Medical Extension Sets Recalled for Potential Sterilization Defects
FDA (Devices) · 2026-08-19
- HighOlympus Toomey Evacuator 30cc medical device sterilization parameters recalled
FDA (Devices) · 2026-07-22
- HighKarl Storz Ureteroscope Recall Due to Unapproved Reprocessing Instructions
FDA (Devices) · 2025-02-05
- HighMedline procedure kits recalled for surgical masks failing sterilization tests
FDA (Devices) · 2025-02-05
- ModerateKarl Storz Hopkins Telescope 6 endoscope recalled for unapproved reprocessing instructions
FDA (Devices) · 2024-10-02