Integra CUSA Clarity System Recalled for Inadequate Sterilization Tray Drying Time
Integra LifeSciences is recalling 820 C7000 CUSA Clarity systems because the 36 kHz Sterilization Tray requires 40 minutes to dry, not the 30 minutes stated in the manual.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a discrepancy in the stated drying time for sterilization trays, which is a procedural or labeling error rather than a direct structural defect or reported injury, fitting the criteria for a moderate severity score.
Plain-English summary
Integra LifeSciences Corp. is recalling 820 units of the C7000 - CUSA Clarity Ultrasonic Tissue Ablation System. The recall involves the 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) and the associated Italian OPS Manual (P/N 60905201).
The recall was initiated because the 36 kHz Sterilization Tray system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes as the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components, but 40 minutes is required for complete drying.
The affected units were distributed in the United States and internationally. Users should refer to the recall notice for specific instructions on how to proceed with the correction.
The recalled product
- Product
- C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
- Manufacturer
- Integra LifeSciences Corp.
- Affected units
- 820
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Italian OPS Manual P/N 60905201
Distribution
Part of a larger recall action
This product is one of 42 recalled by Integra LifeSciences Corp. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 42 products in this recall →Related recalls
Same manufacturer · Integra LifeSciences Corp.
See all →- HighAURORA Surgiscope System Obturator May Break or Separate
FDA (Devices) · 2025-03-12
- HighMediHoney Gel Wound Dressing Recall: Potential Induction Seal Defect
FDA (Devices) · 2025-02-19
- ModerateIntegra implantable hydrocephalus valve recalled for incorrect product description on labels
FDA (Devices) · 2025-01-29
- ModerateCODMAN CERTAS Plus programmable valve recalled for incorrect label descriptions
FDA (Devices) · 2025-01-29
- ModerateProgrammable Valve Device Recalled Due to Incorrect Product Description Labels
FDA (Devices) · 2025-01-29
Same hazard · inadequate drying
See all →- HighUretero-reno fiberscopes recalled due to inadequate channel drying
FDA (Devices) · 2023-11-29
- HighTransnasal Esophagovideoscope Recalled Due to Unvalidated Drying Process
FDA (Devices) · 2023-11-29
- HighSmall intestinal videoscope recalled due to inadequate channel drying process
FDA (Devices) · 2023-11-29
- ModerateIntegra CUSA Clarity System Recalled for Sterilization Tray Drying Defect
FDA (Devices) · 2019-12-18
- ModerateIntegra C7000 CUSA Clarity System Recalled for Inadequate Drying Time
FDA (Devices) · 2019-12-18
Similar recalls
- ModerateCUSA Clarity C7000 Ultrasonic Surgical Aspirator Touch Screen Unresponsiveness
FDA (Devices) · 2026-04-29
- ModerateOmniTract Surgical Retractor System sterilization instructions require extended dry time
FDA (Devices) · 2026-07-15
- HighIntegra CUSA Clarity Footswitch Recalled for Unintended Activation
FDA (Devices) · 2018-03-07
- ModerateMedical device recall: sterilization validator with incorrect incubation time
FDA (Devices) · 2025-03-05
- HighFDA Recalls CUSA Excel Surgical Handpiece for Housing Cracks
FDA (Devices) · 2024-07-03
- HighRecall of Sterile Custom Procedure Trays with Defective Light Gloves
FDA (Devices) · 2017-04-05