Integra C7000 CUSA Clarity System Recalled for Sterilization Tray Drying Defect
Integra LifeSciences is recalling 820 C7000 CUSA Clarity Ultrasonic Tissue Ablation Systems because the 36 kHz Sterilization Tray does not meet the required 30-minute drying time criteria.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: This is a Class II recall involving a labeling/documentation discrepancy regarding drying times. No illnesses or injuries are reported, and the hazard is theoretical, fitting the criteria for a moderate severity score.
Plain-English summary
Integra LifeSciences Corp. is recalling 820 units of the C7000 - CUSA Clarity Ultrasonic Tissue Ablation System. The recall is due to a defect in the 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system, which does not meet the Dry Time Criteria for a 30-minute drying cycle.
The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz components, but 40 minutes is required for complete drying. The affected units were distributed in the United States and internationally to Singapore, Mexico, Canada, Japan, Australia, Colombia, Thailand, and Hong Kong.
Healthcare facilities and users should stop using the affected sterilization trays and contact Integra LifeSciences for further instructions on the recall.
The recalled product
- Product
- C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
- Manufacturer
- Integra LifeSciences Corp.
- Affected units
- 820
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- OUS Ops Manual P/N 60905789
Distribution
Part of a larger recall action
This product is one of 42 recalled by Integra LifeSciences Corp. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 42 products in this recall →Related recalls
Same manufacturer · Integra LifeSciences Corp.
See all →- HighAURORA Surgiscope System Obturator May Break or Separate
FDA (Devices) · 2025-03-12
- HighMediHoney Gel Wound Dressing Recall: Potential Induction Seal Defect
FDA (Devices) · 2025-02-19
- ModerateIntegra implantable hydrocephalus valve recalled for incorrect product description on labels
FDA (Devices) · 2025-01-29
- ModerateCODMAN CERTAS Plus programmable valve recalled for incorrect label descriptions
FDA (Devices) · 2025-01-29
- ModerateProgrammable Valve Device Recalled Due to Incorrect Product Description Labels
FDA (Devices) · 2025-01-29
Same hazard · mis labeling
See all →- ModerateMedline Trach Tube Kit Recalled for Component Labeling Discrepancy
FDA (Devices) · 2026-09-02
- ModerateAVID Medical Recalls Halyard Total Joint Kits for Latex Labeling Deficiency
FDA (Devices) · 2026-09-02
- ModerateOlympus Recalls PowerSeal Sealer Due to Incorrect Outer Box Labeling
FDA (Devices) · 2026-09-02
- HighSura Tanmen Ramen Noodles Recalled for Undeclared Fish Allergens
FDA (Food) · 2026-09-02
- SevereWhite Vein Advanced Alkaloids Chewable Tablets Recall for Excess Active Ingredient
FDA (Food) · 2026-09-02