The Recall Desk

Manufacturer

Philips North America, LLC

423 recalls in our database name Philips North America, LLC as the manufacturing or recalling firm.

What the numbers show

Total
423
Critical
2
Severe
27
Most recent
2026-10-07

Of the 423 recalls from Philips North America, LLC we have scored against our rubric, 2 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 27 (6%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

151–175 of 423

  • HighFDA (Devices)·Z-1841-2024·2024-05-29

    Philips Achieva 3.0T PET imaging device poses potential fire hazard

    A loose terminal connection in the power distribution unit of Philips Achieva 3.0T TX PET imaging systems may create an electrical hotspot, potentially causing smoke or fire. The device is distributed worldwide.

    Product
    Achieva 3.0T TX for PET Model Number (REF): 781479
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1862-2024·2024-05-29

    Philips MR Systems Recalled for Loose Electrical Connection Fire Hazard

    Philips has recalled 200 SmartPath to dStream MR imaging systems worldwide due to a loose electrical connection in the g-MDU unit that may create a fire hazard in hospital technical rooms.

    Product
    SmartPath to dStream for XR and 3.0T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782129; (2) 782113; (3) 781270
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1872-2024·2024-05-29

    Philips Incisive CT Scanner Metal Mounting Box May Expel Fragments

    A metal mounting box on the rotating scanner in the Philips Incisive CT system may become damaged and contact other components, potentially expelling fragments at low velocity. This recall affects 155 units distributed nationwide.

    Product
    Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728144
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1845-2024·2024-05-29

    Philips Ingenia 1.5T MRI machines recalled for loose power connection fire risk

    Philips is recalling 2,262 Ingenia 1.5T MRI systems worldwide because a loose terminal connection in the power distribution unit can overheat and cause fire or smoke.

    Product
    Ingenia 1.5T Model Number (REF): (1) 782140; (2) 782115; (3) 782101; (4) 781396; (5) 781341; (6) 781315
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1853-2024·2024-05-29

    Philips Intera 1.5T MRI System Mains Connection Defect Poses Fire Risk

    Philips is recalling 200 Intera 1.5T MRI systems worldwide due to loose mains distribution unit connections that may cause electrical hotspots, creating fire and smoke hazards in hospital technical rooms.

    Product
    Intera 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781295
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1852-2024·2024-05-29

    MRI System Recalled Due to Loose Terminal Connection Fire Hazard

    Philips Ingenia Elition X- MRI systems have a loose terminal connection in the power distribution unit that may create a hotspot and cause smoke or fire in hospital settings. No injuries have been reported.

    Product
    Ingenia Elition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781358
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1863-2024·2024-05-29

    Philips MRI System Recall: Loose Electrical Terminal May Cause Fire

    Philips MRI diagnostic systems are recalled because a loose electrical connection in the g-MDU unit may create a hotspot, potentially causing fire and smoke in hospital technical areas.

    Product
    SmartPath to Ingenia Elition X - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 782118
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1861-2024·2024-05-29

    Philips Medical Device Recalled for Fire Hazard in Electrical Connection

    Philips is recalling the SmartPath to dStream for 3.0T medical device due to a loose electrical connection in the mains distribution unit that may cause smoke or fire in hospital technical rooms. Three units are affected.

    Product
    SmartPath to dStream for 3.0T Model Number (REF): 782145
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1844-2024·2024-05-29

    Medical Device Mains Distribution Unit Recall Due to Fire Risk

    The Evolution upgrade 3.0T mains distribution unit is being recalled because its L3 terminal connection may become loose, creating a hotspot that could cause fire or smoke in hospital technical rooms.

    Product
    Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1850-2024·2024-05-29

    Philips Ingenia Ambition S MRI systems recalled for potential fire hazard

    Philips is recalling Ingenia Ambition S MRI systems due to a potential fire hazard. A loose terminal connection in the electrical distribution unit may create a hotspot and smoke or fire in hospital technical rooms.

    Product
    Ingenia Ambition S Model Number (REF): (1) 782139; (2) 782133; (3) 782108; (4) 781359
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1851-2024·2024-05-29

    Philips Ingenia Ambition X MR systems recalled for loose terminal connection fire hazard

    Philips is recalling 509 Ingenia Ambition X MR systems due to loose terminal connections in the g-MDU L3 unit that may create hotspots and cause smoke or fire in hospital technical rooms.

    Product
    Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782138; (2) 782109; (3) 781356
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1855-2024·2024-05-29

    Philips Panorama HFO MR Systems Loose Terminal Connection May Create Fire Hazard

    Philips Panorama HFO MR diagnostic imaging systems (Model 781350) contain a loose terminal connection in the mains distribution unit that may create a hotspot and cause smoke or fire. Approximately 118 units are affected worldwide.

    Product
    Panorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781350
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1848-2024·2024-05-29

    Philips Recalls Ingenia 3.0T MR Systems for Loose Mains Terminal Fire Hazard

    Philips is recalling 1,517 Ingenia 3.0T magnetic resonance (MR) systems due to a potentially loose terminal connection in the mains distribution unit that may create a hotspot and cause fire or smoke.

    Product
    Ingenia 3.0T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782103 (2) 781377 (3) 781342
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1838-2024·2024-05-29

    Philips MRI System Recalled for Potential Fire in Power Distribution Unit

    Philips is recalling 2 Achieva 1.5T MRI systems worldwide due to a potential loose connection in the power distribution unit that could cause smoke or fire in hospital technical rooms.

    Product
    Achieva 1.5T Initial system Model Number (REF): 781178
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1836-2024·2024-05-29

    Philips Achieva 1.5T MR Systems Electrical Connection Fire Hazard

    Philips is recalling 1,585 Achieva 1.5T MR systems worldwide due to a loose g-MDU L3 terminal connection that may create an electrical hotspot causing fire or smoke. No injuries have been reported.

    Product
    Achieva 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 781296 (2) 781343
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1839-2024·2024-05-29

    Philips Achieva 3.0T MRI System Recall: Fire Hazard Risk

    Philips recalls 386 Achieva 3.0T MRI systems due to a loose terminal connection that may create a hotspot and cause smoke or fire in the equipment room.

    Product
    Achieva 3.0T Model Number (REF): (1) 781345; (2) 781344; (3) 781278; (4) 781277; (5) 781177
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1827-2024·2024-05-22

    Philips Patient Information Center iX Software: Event Catalog Data Loss Recall

    Philips Patient Information Center iX software version 4.x fails to properly save or transfer event catalog information, risking data loss in healthcare facilities. The FDA issued a Class II recall affecting 358 units globally.

    Product
    Patient Information Center iX and Patient Information Center iX Expand, Software Version 4.x.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1830-2024·2024-05-22

    Philips Ingenia 3.0T MR Systems May Not Warn of Low Ventilation Settings

    Philips Ingenia 3.0T MR systems with software R5.7.1 may fail to warn about low ventilation, potentially causing patient overheating or burning sensations. 19 units affected worldwide.

    Product
    Ingenia 3.0T with Evolution Upgrade 3.0T - - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782117
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1829-2024·2024-05-22

    MRI System Software May Fail to Warn of Inadequate Ventilation

    Philips Ingenia 1.5T MRI systems with software R5.7.1 may not provide adequate warning for low ventilation conditions, potentially causing increased patient body temperature or burning sensation.

    Product
    Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782116
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1756-2024·2024-05-08

    Philips Spectral CT Oncology Imaging System Recall: Laser Reflection Hazard

    Philips recalls 12 units of its Spectral CT Oncology Essentials system because a marking laser could reflect off the gantry cover, potentially causing misdiagnosis and incorrect patient treatment.

    Product
    The Oncology Essentials Package of the Spectral CT imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1690-2024·2024-05-01

    Philips Patient Information Center iX UPS devices recalled for potential power failure

    Philips is recalling certain Patient Information Center iX Uninterruptable Power Supply devices due to potential power supply failures that could cause the monitoring system to shut down, risking delays in detecting patient condition changes.

    Product
    Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterruptable Power Supply Model: SCL500RM1U 2. APC (Schneider Electric IT Corporation) 230V Uninterruptable Power Supply Model…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1581-2024·2024-04-24

    Philips Intera 1.5T MRI System Patient Support Table Installation Defect

    Philips is recalling certain Intera 1.5T MRI systems because the patient support table floor plate may be incorrectly installed, which could affect patient safety.

    Product
    Intera 1.5T Omni/Stellar, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1584-2024·2024-04-24

    Philips MRI System Patient Support Table Floor Plate Installation Issue

    Philips is recalling 9 Intera 3.0T Quasar Dual MRI systems (Model 781150) distributed worldwide. The patient support table floor plate may be incorrectly installed on these units.

    Product
    Intera 3.0T Quasar Dual, Magnetic Resonance System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1580-2024·2024-04-24

    Philips MRI patient support table floor plate installation defect recall

    Philips is recalling 107 Intera 1.5T MRI systems worldwide due to potential incorrect installation of the patient support table floor plate, which may create safety risks during patient use.

    Product
    Intera 1.5T Master/Nova, Magnetic Resonance System.
    Category
    Medical Device
    Distribution