Hemodynamic recording systems recalled over pressure sensor error during blood pressure measurement
Merge Hemo systems are recalled because certain Schiller ARGUS PB-3000 patient data modules trigger a Pressure Sensor Defective error during non-invasive blood pressure measurement above 240 mmHg. 84 units are affected.
- Product
- Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood fl
- Category
- Medical Device
- Distribution
- Distributed nationwide