The Recall Desk

Hazard

Occupant Impact Risk recalls

1 recall in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2023-05-04

Of the 1 occupant impact risk recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all occupant impact risk recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–1 of 1

  • HighNHTSA·23V324000·2023-05-04

    2022-2023 Land Rover Range Rover Second-Row Armrest Latch Failure Recall

    Jaguar Land Rover is recalling 2022-2023 Range Rover vehicles because the second-row armrest latch may fail, allowing unsecured objects to strike occupants during a crash. Dealers will replace the locking arm free of charge.

    Product
    LAND ROVER — LAND ROVER RANGE ROVER SPORT, RANGE ROVER
    Category
    Vehicle
    Distribution
    Distributed nationwide