The Recall Desk

FDA (Devices) recall action 93643

Philips North America recalled 3 products on 2024-01-17

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-01-17
Critical
0
Units
—

Why these products were recalled

While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.

The agency’s own wording, quoted from the enforcement report.

1–3 of 3

  • HighFDA (Devices)··2024-01-17

    Philips BrightView Gamma Camera System Recalled for Extremity Entrapment Hazard

    Philips is recalling BrightView Gamma Camera Systems because a gap can form between patient support and detector during quality assurance scans, creating an extremity entrapment hazard that may cause fractures or crush injuries.

    Product
    BrightView, Gamma Camera System, Product Code 882480.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-01-17

    Gamma Camera Recalls Due to Potential Patient Extremity Entrapment Hazard

    Philips BrightView XCT gamma cameras may pose an extremity entrapment hazard during certain scanning operations. A gap can form between the patient support and detector, risking fractures, lacerations, and crush injuries.

    Product
    BrightView XCT, Gamma Camera, Product Code 882482
    Category
    Distribution
    0 states