The Recall Desk

FDA (Devices) recall action 92635

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalled 2 products on 2023-08-09

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-08-09
Critical
0
Units
1,296

Why these products were recalled

The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2023-08-09

    X-ray therapy system carriage bolts may loosen and suspension rail crack

    Philips Azurion 7 M20 X-ray imaging systems may have loose or broken carriage support bolts and cracks in the suspension rail. The Class II recall affects 648 units distributed worldwide.

    Product
    Azurion 7 M20. X-ray Image guided Therapy System.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-08-09

    X-ray Imaging System Carriage Bolts May Loosen or Break

    Philips Allura Xper FD10 and FD20 X-ray systems may have loose or broken bolts supporting the FlexMove Carriage and cracks in the suspension rail. Affected units distributed worldwide; users should inspect equipment and contact manufacturer.

    Product
    Allura Xper FD10 and FD20. X-ray Imaging System.
    Category
    Distribution
    0 states