The Recall Desk

FDA (Devices) recall action 87313

Cardinal Health 200, LLC recalled 3 products on 2021-03-24

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2021-03-24
Critical
0
Units
358,050

Why these products were recalled

Safety shield not locking into the cannula. If the safety shield activation fails to lock a contaminated needle, a needlestick may occur and there is the potential for exposure to bloodborne pathogens

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2021-03-24

    Cardinal Health Recalls Magellan Syringes Due to Safety Shield Defect

    Cardinal Health is recalling Magellan 3 mL syringes with hypodermic safety needles because the safety shield may fail to lock, posing a needlestick risk.

    Product
    Magellan 3 mL Syringe with Hypodermic Safety Needle, 25 G x 5/8" (0.508 mm x 1.6 cm). A single lumen needle intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. Product Code: 8881833558
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-03-24

    Cardinal Health Recalls McKesson Prevent M Needles Due to Safety Shield Defect

    Cardinal Health is recalling 108,000 McKesson Prevent M needles because the safety shield may fail to lock, posing a risk of needlestick injuries and exposure to bloodborne pathogens.

    Product
    McKesson PREVENT M NEEDLE (Magellan Technology) 25G X 5/8". A single lumen needle intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. Product Code: 26558
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-03-24

    Cardinal Health Recalls Magellan Hypodermic Safety Needles Due to Locking Failure

    Cardinal Health is recalling Magellan Hypodermic Safety Needles because the safety shield may fail to lock, creating a risk of needlestick injuries and exposure to bloodborne pathogens.

    Product
    Magellan Hypodermic Safety Needle, 25 G x 5/8" (0.508 mm x 1.6 cm). A single lumen needle intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. Product Code: 8881850558
    Category
    Distribution
    0 states