FDA (Devices) recall action 85297

WELCH ALLYN, INC/MORTARA recalled 10 products on 2020-05-06

The FDA (Devices) publishes a recall like this one as 10 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 10 pages.

Products
10
Announced
2020-05-06
Critical
0
Units
785,210

Why these products were recalled

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–10 of 10

  • HighFDA (Devices)··2020-05-06

    Welch Allyn X12+ Telemeter and Patient Cable Recall for Defibrillation Safety

    Welch Allyn is recalling X12+ Telemeters and patient cables because they may not meet defibrillation withstand requirements.

    Product
    X12+ and patient cable for the X12+ Telemeter with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP, PAT CBL 4WIRE AHA SNAP HC.
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn ELI PC Service Recalled for Defibrillation Safety Non-Compliance

    Welch Allyn is recalling ELI PC Service software because it may not meet defibrillation withstand safety requirements for medical electrical equipment.

    Product
    ELI PC. The ELI PC Service automatically converts resting ECG recordings collected with WAM PC patient cables and WebUpload into DICOM 12-lead ECG Waveform files.
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn Surveyor S12 and S19 Patient Monitors Recalled for Defibrillation Safety

    Welch Allyn is recalling 78,521 Surveyor S12 and S19 bedside monitors and patient cables because they may not meet defibrillation withstand safety requirements.

    Product
    Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA SNAP JSCREW, PAT CBL 10WIRE IEC SNAP JSCREW. The Surveyor S12 patient monitor is designed for continuous monitoring in either stationary or portable…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn Q-Tel RMS System Recalled for Defibrillation Safety Non-Compliance

    Welch Allyn is recalling 78,521 Q-Tel RMS systems because they may not meet defibrillation withstand safety requirements.

    Product
    Q-Tel. The Q-Tel RMS system is a computer-based cardiac and pulmonary rehabilitation data acquisition and editing system.
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn ECG Lead Sets Recalled for Defibrillation Safety Non-Compliance

    Welch Allyn is recalling 78,521 ECG lead sets because they may not meet defibrillation withstand safety requirements.

    Product
    Replaceable lead set for the Wireless Acquisition Module (WAM), which is the ECG front-end for various Electrocardiographs. Sold under the following device names: LEAD SET WAM 10 WIRE BANANA AHA GRAY, LEAD SET WAM 10 WIRE BANANA IEC GRAY, RPLCE LD SET WAM/AM12 LIMBS BANA AHA…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn T12 Wireless ECG Devices Recalled for Defibrillation Safety

    Welch Allyn is recalling T12 wireless ECG devices because they may not meet defibrillation withstand safety requirements.

    Product
    T12. Incorporates wireless electrocardiographic technology to achieve the real-time acquisition and RF transmission of simultaneous 12-lead ECG data with diagnostic quality while allowing the patient to be ambulatory.
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn S4 Wireless Telemeter Recall for Defibrillation Safety

    Welch Allyn is recalling S4 Wireless Telemeters and patient cables because they may not meet defibrillation withstand safety standards.

    Product
    S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE LF ECG SNAP ENDS IEC GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS AHA GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS IEC GRAY. The S4 mobile monitor…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn Q-Stress and XScribe Diagnostic Devices Recalled for Safety Standard Deficiency

    Welch Allyn is recalling Q-Stress and XScribe diagnostic devices because they may not meet specific defibrillation withstand safety requirements.

    Product
    Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wired or wireless acquisition modules.
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn Electrocardiographs Recalled for Defibrillation Safety Non-Compliance

    Welch Allyn is recalling 78,521 lead diagnostic electrocardiographs because they may not meet defibrillation withstand requirements.

    Product
    Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and…
    Category
    Distribution
    51 states
  • HighFDA (Devices)··2020-05-06

    Welch Allyn H12+ Holter Recorders Recalled for Defibrillation Safety Non-Compliance

    Welch Allyn is recalling H12+ Holter Recorders and patient cables because they may not meet defibrillation withstand requirements.

    Product
    H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP. The 24-hour H12+ recorder uses a single AA…
    Category
    Distribution
    51 states