The Recall Desk

FDA (Devices) recall action 83583

Beckman Coulter Inc. recalled 2 products on 2019-10-30

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2019-10-30
Critical
0
Units
40,760

Why these products were recalled

NAPQI, a metabolite of Acetaminophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.

NAPQI, a metabolite of Acetominophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2019-10-30

    Beckman Coulter Uric Acid Assay Recalled Due to Interference

    Beckman Coulter is recalling Uric Acid assays because acetaminophen metabolites cause negative interference in overdose patients.

    Product
    Beckman Coulter Uric Acid, REF OSR6098, OSR6198, and OSR6698.
    Category
    Distribution
    0 states