Sentinel CH SpA recalled 2 products on 2019-10-02

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2019-10-02
Critical
0
Units
17,136

Why these products were recalled

Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2019-10-02

    Sentinel CH SpA Recalls Alinity c Lithium Reagent Kits

    Sentinel CH SpA is recalling Alinity c Lithium Reagent Kits due to a potential carryover issue that may result in falsely elevated lithium patient results.

    Product
    Alinity c Lithium Reagent Kit- intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT c Systems and the Alinity c analyzer List Number: 08P5320
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2019-10-02

    Sentinel CH SpA Recalls MULTIGENT Lithium Reagent Kits

    Sentinel CH SpA is recalling 15,919 MULTIGENT Lithium reagent kits due to a potential carryover issue that may cause falsely elevated patient results.

    Product
    MULTIGENT Lithium-intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT c Systems and the Alinity c analyzer List Number: 8L25-30
    Category
    Distribution
    0 states