The Recall Desk

FDA (Devices) recall action 83425

GE Healthcare, LLC recalled 4 products on 2019-12-25

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2019-12-25
Critical
0
Units
27,494

Why these products were recalled

Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2019-12-25

    GE Healthcare Recalls Aestiva 7900 Anesthesia Systems for Cybersecurity Vulnerability

    GE Healthcare is recalling 7,168 Aestiva 7900 anesthesia systems worldwide due to a cybersecurity vulnerability that could allow malicious actors to send fraudulent flow sensor correction parameters.

    Product
    Aestiva 7900, Model Numbers: a) 1006-9300-000 b) 1006-9305-000 c) 1006-9118-000 d) 1006-9112-000 e) 1006-9113-000 f) 1006-9020-000 g) 1006-9033-000 h) 1006-9117-000 i) 1006-9119-000 j) 1006-9030-000 k) 1006-9032-000 l) 1006-9022-000 m) To be provided n)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-12-25

    GE Healthcare Recalls Aestiva 7100 Anesthesia Systems Due to Cybersecurity Vulnerability

    GE Healthcare is recalling 473 Aestiva 7100 anesthesia systems worldwide due to a cybersecurity vulnerability that could allow malicious actors to send fraudulent flow sensor correction parameters.

    Product
    Aestiva 7100, Model Numbers: 1. 1006-9320-000 2. 1006-9321-000 3. 1006-9320-000-001101 4. 1006-9021-000 5. 1006-9012-000 6. 1006-9111-000 7. 1006-9321-000-007639 8. 1006-9321-000-000023 9. 1006-9321-000-015180 10. 1006-9320-000-010369 11. 1006-9321-000-019486 12. 1
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-12-25

    GE Healthcare Recalls Aespire and Aestiva Anesthesia Systems for Cybersecurity Vulnerability

    GE Healthcare is recalling Aespire and Aestiva anesthesia systems due to a cybersecurity vulnerability that could allow malicious actors to send fraudulent flow sensor correction parameters.

    Product
    Aespire 7100, Model Numbers: 1. 1009-9000-000-009469 2. 1009-9000-000 3. 1009-9000-000-013913 4. 1009-9000-000-001366 5. 1009-9000-000-009857 6. 1009-9003-000 7. 1009-9000-000-008120 8. 1009-9000-000-009600 9. 1009-9000-000-014290 10. 1009-9000-000-010585 11. 1009-90
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-12-25

    GE Healthcare Recalls Aestiva MRI Systems Due to Cybersecurity Vulnerability

    GE Healthcare is recalling 502 Aestiva MRI systems worldwide due to a cybersecurity vulnerability that could allow malicious actors to send fraudulent flow sensor correction parameters.

    Product
    Aestiva MRI, Model Numbers: a) 1006-9310-000 b) 1006-9110-000 c) 1006-9023-000 d) 1006-9028-000 e) 1006-9310-000-305077 f) 1006-9310-000-015243 g) 1006-9310-000-017602 h) 1006-9310-000-103785 I) 1006-9310-000-025109 j) 1006-9310-000-009650 k) 1006-9310-000-015224 l) 1
    Category
    Distribution
    0 states