FDA (Devices) recall action 80271

Philips Medical Systems Nederlands recalled 2 products on 2018-07-04

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2018-07-04
Critical
0
Units
985

Why these products were recalled

The fixation of the upper and lower tilt actuator, of the Omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. This rotating movement can not be stopped by the user. This could potentially lead to the patient falling off the table. There is a risk of injury for the patient, user, or bystander.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • SevereFDA (Devices)··2018-07-04

    Philips OmniDiagnost X-ray System Recalled for Tilt Actuator Failure Risk

    Philips is recalling 477 OmniDiagnost Classic X-ray systems because a broken tilt actuator could cause the table to rotate uncontrollably, posing a risk of injury.

    Product
    OmniDiagnost Classic X-ray system Angiographic and Diagnostic, Product Nos. 70859, 708023, 708024, 708025
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2018-07-04

    Philips OmniDiagnost-Eleva X-ray System Recalled Due to Table Rotation Hazard

    Philips is recalling 508 OmniDiagnost-Eleva X-ray systems because a broken actuator fixation can cause the table to rotate uncontrollably, posing a risk of injury.

    Product
    OmniDiagnost-Eleva X-ray system Angiographic and Diagnostic, Product Nos. 708026, 708027, 708028
    Category
    Distribution
    0 states