Zimmer Biomet, Inc. recalled 2 products on 2018-04-04

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2018-04-04
Critical
0
Units
—

Why these products were recalled

The AIM Tibial Nails failed a clinical analysis where the surgical revision rate exceeded the acceptance criteria in the Device Performance Review protocol for the implant system.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls AIM Tibial Nails Due to High Revision Rates

    Zimmer Biomet is recalling AIM Tibial Nails because they failed a clinical analysis showing surgical revision rates exceeded acceptance criteria.

    Product
    AIM Tibial Nails intramedullary fixation rod - NON-STERILE; Item numbers: 810508255 NAIL TIB DYN 08 X 25.5 810508270 NAIL TIB DYN 08 X 27.0 810508285 NAIL TIB DYN 08 X 28.5 810508300 NAIL TIB DYN 08 X 30.0 810508315 NAIL TIB DYN 08 X 31.5 810508330 NAIL TIB DYN 08 X 33.0…
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2018-04-04

    Zimmer Biomet Recalls AIM Tibial Nails Due to High Revision Rates

    Zimmer Biomet is recalling AIM Tibial Nails because the surgical revision rate exceeded acceptance criteria in a clinical analysis.

    Product
    AIM Tibial Nails intramedullary fixation rod - STERILE; Item numbers: 800508255 NAIL TIB DYN 08X25.5 - STERILE 800508270 NAIL TIB DYN 08X27.0 - STERILE 800508285 NAIL TIB DYN 08X28.5 - STERILE 800508300 NAIL TIB DYN 08X30.0 - STERILE 800508315 NAIL TIB DYN 08X31.5 - STERILE…
    Category
    Distribution
    1 state