The Recall Desk

FDA (Devices) recall action 77145

Beckman Coulter Inc. recalled 2 products on 2017-05-31

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2017-05-31
Critical
0
Units
—

Why these products were recalled

There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2017-05-31

    Beckman Coulter UniCel DxH 800 Recalled for Potential Data Errors

    Beckman Coulter is recalling UniCel DxH 800 systems because a software issue may cause erroneous white blood cell differential results.

    Product
    UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322.
    Category
    Distribution
    1 state